We report a new methodology for the electromechanical characterization of organic monolayers based on the implementation of dual AC resonance tracking piezo force microscopy (DART-PFM) combined with a sweep of an applied DC field under a fixed AC field. This experimental design allows calibration of the electrostatic component of the tip response and enables the use of low spring constant levers in the measurement. Moreover, the technique is shown to determine both positive and negative piezo response. The successful decoupling of the electrostatic component from the mechanical response will enable more quantitative electromechanical characterization of molecular and biomaterials and should generate new design principles for soft bio-compatible piezoactive materials. To highlight the applicability, our new methodology was used to successfully characterize the piezoelectric coefficient (d33) of a variety of piezoactive materials, including self-assembled monolayers made of small molecules (dodecane thiol, mercaptoundecanoic acid) or macromolecules (peptides, peptoids), as well as a variety of inorganic materials, including lead zirconate titanate [PZT], quartz, and periodically poled lithium niobate [PPLN]. Due to high differential capacitance, the soft organic monolayers demonstrated exceedingly large electromechanical response (as high as 250 pm V−1) but smaller d33 piezocoefficients. Finally, we find that the capacitive electrostatic response of the organic monolayers studied are significantly larger than conventional inorganic piezoelectric materials (e.g., PZT, PPLN, quartz), suggesting organic electromechanical materials applications can successfully draw from both piezo and electrostatic responses.
Accurate and reliable methods to measure piezoelectric outputs from a given material are vital to investigating these phenomena and realizing their potential range of applications. Atomic force microscopy (AFM) was initially developed to map the morphological variations in materials at the nanoscale.6,17 Beyond simple topology and morphology, functional AFM methods have been developed to map properties including surface potential, charge transport, magnetic response, and piezoresponse.17–21 The latter, piezo force microscopy (PFM), determines the mechanical response of materials to an applied electrical field by measuring the converse piezoelectric effect. However, classical single frequency PFM suffers from low sensitivity and poor frequency tracking due to crosstalk in the phase feedback loop between material topology and electromechanical response. To increase sensitivity and avoid dielectric breakdown of materials, dual AC resonance tracking (DART) was developed by Kalinin to allow the use of small bias voltages while maintaining good frequency tracking despite varying topological features.4,6,18,20–24 Building on the principles of PFM, DART drastically improved the sensitivity of PFM measurements and helped move the field towards more quantitative piezoelectric measurements. Beyond DART, the band excitation (BE) method was intended to overcome distortions associated with tip–sample interactions experienced in DART, in which the lever is excited at multiple frequencies around the fundamental frequency to alleviate shifts in the fundamental, due to topography.25 More recently, several groups have tried to reduce/eliminate these distortions by using high spring-constant (kl) levers, with or without a fixed external DC field, or by creating new lever technologies, such as “inner paddled levers”.2,3,26–28 These techniques reduce the electrostatic component of the measurement for specific cases; however, this may not be true for systems, such as organic polymers and biomaterials, in which the electrostatic component is quite large or where the Young's modulus of the material is small in comparison to the lever.
The above methods, particularly the use of high spring-constant levers, perform best with materials in which the elastic modulus is significantly higher than that of the lever. Unfortunately, when the modulus of the material under study is small in comparison to modulus of the lever, such as organic and biomaterials, the lever may deform the target surface, reducing or eliminating the sensitivity enhancements garnered by DART or band excitation techniques.
In this work, we describe a method for improved accuracy in measurements of the piezo-response (d33, the response of a material in the z-axis to a field applied in the same axis) of soft organic monolayers. The method entails the use of a soft (low-kl) lever, coupled with the quantification of the electrostatic component of tip response by completing a DC field sweep in addition to the AC field sweep already employed to measure the independent lever electrostatics. By compensating for the electrostatic component, the true d33 of the material can be established.
The tip response can be determined as in eqn (1)17,24
(1) |
In the case of soft materials, such as organic and biomaterials, using stiff, high kl levers will likely cause significant deformation of the target material. Since DART-PFM uses contact resonance for signal enhancement, the mismatch between the soft surface and stiff AFM lever leads to small tune amplitudes even under large applied fields and thus poor signal to noise. An apt analogy to this situation would be measuring the response of grass with a hammer – compressing the plant and limiting the observable response. Consequently, as proposed in the introduction, softer, low kl levers should minimize surface deformation in soft organic and biomaterials; however, they bring additional complications in the form of significant electrostatic contributions to the observed deff. Unlike in traditional AC sweep methods here the electrostatic component is expected to be non-zero at zero applied field highlighting the effects of electrostatics on the measurement system. To account for this electrostatic effect, we envisioned sweeping the DC field to accurately determine the electrostatic component of the observed response, as well as the VDC point at which the electrostatic response is minimized (Fig. 1B). If successfully realized, we hypothesized this new technique would allow for increased quantitative accuracy in determining the deff piezo response even in soft materials.
To test the proposed DC field sweeping DART-PFM technique, five different levers were chosen with spring constants (kl) varying from 0.02 to 2.8 N m−1 and used to determine the electromechanical response of four organic self-assembled Au–S monolayers (Scheme 1). These organic SAM systems were chosen due to their innate polar alignment; thus reducing or eliminating any electrostriction or flexoelectric response of the films in conjunction with being non ferroelectric. The SAMs tested included small molecule ligands (DDT, MUA) as well as bio-inspired peptide and peptoid oligomers (A and B) examined in our prior work.16 The response of each target film was measured at varying piezo stack voltages, generating varying effective kl values. Fig. 2 illustrates the resulting experiment, in which the recorded response for a given target material increases exponentially as the effective kl value of the lever used in the measurement decreases. The results confirm that for soft materials like SAMs, using levers with spring constants comparable to the modulus of the material's leads to increased response. In some cases, experimental tip responses reach 250 pm V−1, far exceeding previously reported electromechanical response in these soft materials. Though the overall electromechanical response is high, as discussed below, these responses are influenced more by electrostatics than the intrinsic piezo response of the materials (d33). While the spring constant of the lever (kl) is shown to influence the response of the films, it is merely contributing to changes in the contact stiffness (kc).17,20,24
Scheme 1 Compounds under study.16 |
Fig. 2 Relationship between tip response (kl and kc as calculated using eqn (2)) for various SAMs using AFM levers with spring constant from 0.02–2.8 N m−1, for (a) DDT, (b) MUA, (c) peptide A and (d) peptoid B respectively. The best-fit line is to y = a + bxc. |
As eqn (1) illustrates, while stiffer levers affect the response, it is the contact stiffness (kc) that directly influences the measurement.24 While the distinction may seem subtle, kl is merely a single component of the contact. Thus, the spring constant of the contact derives from the lever, the mechanical response of the material in the x, y, and z-axis, influence of surface electrostatics, and any tip–sample meniscus that may be present (e.g., in ambient conditions). Fortunately, these factors can be estimated by applying eqn (2) to the already measured kl values (as part of tip–sample tuning in DART-PFM).17
(2) |
Eqn (2) approximates the spring constant of contact (kc) from the spring constant of the lever (kl) by taking the ratio of the resonance frequency of the free lever (ω0) to the lever in contact with the sample (ω1) used in the DART experiment.17
While the use of stiffer levers is correlated to an increased contact stiffness, using the calculated kc values to model tip response, yield better fits (Table 1), reflecting the correct physics due to the higher spring constant of the contact stiffness dominating. The comparable fits of tip response to kl values, found in Table S1,† qualitatively maintain the same trend – decreased spring constant yields increased electromechanical tip response, albeit with lower quality of fit (R2).
Material | Constant (a) | Coefficient (b) | Power (c) | R 2 |
---|---|---|---|---|
DDT | 2.53 | 293 | −1.30 | 0.948 |
MUA | 2.56 | 206 | −1.44 | 0.959 |
Peptide A | 2.32 | 187 | −1.47 | 0.985 |
Peptoid B | −2.42 | 327 | −1.07 | 0.993 |
However, as indicated in eqn (1), an ideal dependence would yield an exponent of kc−1 (example plot in ESI†), but the values determined by fits in Fig. 2 and Table 1 deviate significantly. In all the organic monolayers, the tip response falls off faster than predicted (i.e., exponents closer to ca. −1.3) with increasing contact stiffness. We speculate that this derives from the stiffer levers distorting the monolayers instead of remaining at the interface. The only exception is for the peptoid B SAMs, in which the tip-response curve yields an exponent close to −1.0, suggesting that the peptoid film is significantly stiffer than the other films, as confirmed by AMFM measurements discussed below, and consistent with expectations of a peptoid PPI helix.27
While eqn (2) allows an approximate conversion of kl to kc values, assuming a uniform shift from the fundamental frequency of the lever to the measured frequency of the lever while interacting with surface, kc was also measured directly using amplitude modulated force microscopy (AMFM).10,30–32 Due to the trends observed in the original kl measurements, the kc was not directly measured by AMFM for all levers. Only the ASYELEC.01 R2 and the TR400PB (S) levers, 2.8 and 0.09 N m−1 respectively, were chosen as the relative extremes of contact stiffnesses observed in the initial study, (Table in ESI†). We note that the measured kc values deviate substantially from eqn (2) for stiffer levers, again suggesting that the stiffer levers are distorting the monolayers, effectively limiting the ability of the soft materials to mechanically respond to the applied electric fields.
As mentioned above, while soft levers give higher tip response, they also suffer from greater levels of electrostatic interference than stiffer levers. One way to account for this effect would be to apply a VDC to the tip that is equal to VCPD, thus eliminating the electrostatic term in eqn (1). Intuitively one simple solution would be to measure the VCPD by SKPFM, and then apply that VDC, as has been previously implemented.27 The problem arises from the nature of the DART measurement where a VAC is applied on top of the VDC, altering the electrostatic environment around the contact, modulating the intrinsic VCPD of the sample. Instead, we swept the DC field to find the point of minimal tip response at which the contact potential equals the applied DC field under a constant VAC (Fig. 1B).
The tip response (R) is the measured output of the DART experiment after the simple harmonic oscillator (SHO) calculation corrects for the tip–sample resonance enhancement. This tip response can be separated, using eqn (1), into the intrinsic piezoresponse of the material and the electrostatic response. When VDC is equal to VCPD, the electrostatic component of the measurement will go to zero leaving only the mechanical response of the material under the applied field. The organic SAM films are intrinsically polar, permanent piezoelectric materials, since one end is attached via an Au–S bond. Consequently, one expects no ferroelectric hysteresis from sweeping VDC, only two intersecting lines of equal slope proportional to kc−1 dC/dz (Fig. 1B). The intersection point will represent the piezoelectric response deff × VAC. The results are highlighted in Fig. 3 and summarized in Table 2, in which three different AFM levers are used on two different organic SAMs.
Material | k l (N m−1) | V CPD (V) | Slope (A) | R 2 | d eff (pm V−1) |
---|---|---|---|---|---|
DDT | 2.8 | −0.369 | 0.827 | 0.818 | 0.12 |
DDT | 0.09 | 0.001 | 75.1 | 0.998 | −0.28 |
DDT | 0.02 | −0.419 | 360 | 0.999 | −0.32 |
Peptide A | 2.8 | −0.245 | 0.979 | 0.921 | 0.12 |
Peptide A | 0.09 | −0.111 | 56.3 | 0.995 | −0.16 |
Peptide A | 0.02 | −0.153 | 373 | 0.987 | 3.2 |
Fig. 3 establishes that the proposed new method works for fixed polar molecular monolayer films. The technique is demonstrated on two SAMs: one piezo active peptide16 and a control of DDT, used to highlight the natural polarization of organic SAMs when adsorbed to a metallic surface. The high electrostatic component of the low kl levers is easily compensated through the new method. The results point to a piezoresponse range of −0.33 to 0.11 pm V−1 for DDT and −0.16 to 3.2 pm V−1 for peptide A. The measured deff of peptide A using the 0.02 N m−1 lever is significantly larger than the values determined with the stiffer levers.
Further, by highlighting three different spring constant levers ranging from 0.02 to 2.8 N m−1 the results from Fig. 2 can be reaffirmed. Here, film response increases with decreasing kl due to electrostatic effects, reducing the maximal response at 2.0 VDC and 4.0 VAC from near 1000 pm to ∼3 pm. These results represent a greater than 300-fold decrease in measured response; moreover the inset charts in Fig. 3 demonstrate that at high kc, relative to the sample material, the instrument sensitivity bottoms out, effectively identifying the noise floor of the measurement technique. The inset charts emphasize the trend towards higher R2 values where at high kc and kl, response is sporadic and hard to model in contrast to the low kl levers. The increase in sensitivity is further confirmed by the changes in the tune amplitude, at the described set points, from <2 V to >50 V. These, results reflect the benefits of the new method by demonstrating increased precision in the determination of the deff for soft monolayers through enhanced signal to noise ratios.
Based on the evidence in Fig. 2 and 3, the use of soft, low-spring-constant the TRS levers (0.09 N m−1) are less likely to perturb organic monolayers, and the DC-sweep DART-PFM technique enables separation of inherent piezoelectric response of a material from the electrostatic components to tip response. Consequently, TRS levers were used with DC-sweep DART-PFM across four organic SAMs and a quartz crystal microbalance (QCM). The latter serves as a non-ferroelectric control with known piezoresponse (d33), while DDT and MUA SAMs were used as control organic monolayers with low expected piezoresponse, but varying hydrophobicity. If the contact stiffness depends on the effects of a meniscus at the tip sample interface under ambient conditions, modulating from a hydrophobic DDT monolayer to a hydrophilic MUA monolayer should reveal such effects on measured electromechanical response. Peptide A and peptoid B represent helical piezoactive materials with different backbone motifs that give rise to differences in helical propensity.16
The DC-sweep DART-PFM response of these films under a constant 3.0 VAC field is illustrated in Fig. 4 and compiled in Table 3. The resulting field plots yield deff of the varying materials. QCM stands out with a deff value consistent with literature (i.e. 1.68 pm V−1vs. 2.3 pm V−1),19 but the observed tip response (e.g., Fig. 4a) is much smaller, compared to the other monolayer samples. The low slope indicates that the magnitude of the deff in quartz is not significantly different from that of the monolayers, but its electrostatic component is minimal compared to the monolayer films. This likely indicates that the ability of quartz to build a large differential capacitance in the z-axis is significantly smaller in comparison to the SAMs. Further, these results reconfirm previously reported conclusions that the helix forming peptide and peptoid have higher piezo electric coefficients than DDT and MUA.16
Material | V CPD (V) | Slope (A) | R 2 | d eff (pm V−1) |
---|---|---|---|---|
QCM | −1.06 | 6.40 | 0.993 | 1.68 |
DDT | −0.022 | 119 | 0.999 | 0.100 |
MUA | −0.249 | 63.9 | 0.991 | −0.560 |
Peptide A | 0.157 | 94.1 | 0.975 | −6.42 |
Peptoid B | 0.165 | 134 | 0.999 | −1.35 |
More significant than the magnitude of the tabulated piezo coefficients in Table 3 is the sign. Noticeably three out of four SAMs have a negative deff, indicating that they compress under an applied field. Only DDT produced a positive deff, albeit close to zero. This negative piezo response differs from conventional piezo ceramics such as ZnO or PZT, but is similar to that observed in PVDF and a variety of piezoelectric materials.33–35 Thus, the new method not only determines positive, but also negative piezoresponse, even at low applied voltages.
We note both the deff and VCPD from the peptide A monolayer shifts by applying different VAC between Fig. 3 and 4 (4.0 VAC and 3.0 VAC respectively). To test if the VCPD and deff is subject to shifting under various experimental conditions a film of peptoid B was tested against four different AC voltages sweeping through six DC voltages at each AC voltage. Extracting the surface potential under experimental conditions from Fig. 5A and comparing it to the applied AC field a linear trend emerges. As the applied electric field increases under the experimental conditions so does the VCPD. This indicates that a static VDC determined by sKPFM cannot be used directly to eliminate the electrostatic component of the measured response, as has been previously suggested.27 Further Fig. 5A represents the equivalent of eight experimental runs on one sample using the more traditional AC sweep method, thus confirming the repeatability of the new measurement system and the lack of dielectric brake down of the films due to the applied fields.
Fig. 5A highlights that the maximal response of the film increases with increasing VAC as suggested in conventional piezoelectric materials and measurement techniques. To confirm this, a map was extracted from Fig. 5A to generate 5C where the response of the film is plotted against the applied VAC at each DC voltage, recreating the conventional approach to the determination of deff by DART-PFM. This exercise emphasizes the ability of the new DC sweep method to remove the effects of electrostatic response and thereby reduce the variation in the measured deff. The new method gives the deff of peptoid B to be −0.24 ± 1.36 pm V−1 in comparison to the traditional method with the electrostatics unaccounted for at 188.50 ± 39.25 pm V−1. To supplement these conclusions the points at each VDC in Fig. 5A were averaged with VAC removed to generate Fig. 5B. In Fig. 5B the variation at each point is so insignificant that the error does not show on the plot. As expected, the calculated deff from Fig. 5B agrees with Fig. 5A at −0.223 pm V−1. Though, these figures are not identical to the deff values determined in Fig. 4, they are within error of each other (Table 4).
V AC (V) | V CPD (V) | Slope (A) | R 2 | d eff (pm V−1) |
---|---|---|---|---|
3.0 | 0.403 | 277.6 | 0.999 | 0.360 |
2.0 | 0.225 | 184.9 | 0.999 | −2.18 |
1.0 | 0.192 | 85.99 | 0.996 | 3.37 |
0.5 | 0.079 | 43.86 | 0.999 | −2.51 |
NA | 0.227 | 89.67 | 0.999 | −0.223 |
The comparison between the molecular monolayers and quartz highlights a significant shift of material response to an applied field. The slopes of the plots in Fig. 3 represent the electrostatic component of the material response. When comparing the materials there is a significant shift in the slope of the fits indicating a variation in the effect of electrostatics on the reported response. Quartz has a fundamentally shallower slope than any of the molecular films. Likely the applied AC field or the differential capacitance in the z-axis are the influencing factors. The AC field however is uniformly applied at 3.0 V across all samples and accounted for when the final response is computed. In addition, a humidity-controlled chamber held at approximately 20% provides no likely outside source for field augmentation, ensuring little to no variation in the meniscus formed at the tip–sample interface. Hence, the contribution from the differential capacitance in the z-axis is likely the source of the discrepancy in the overall measured response. This difference in capacitance is likely due to a difference between the relatively high dielectric constant of quartz (ε ∼ 4) compared to the lower dielectric constant of the SAMs (MUAε ∼ 2).36,37
To test this hypothesis, several conventional hard-ceramic piezoelectric materials were tested, in addition to the non-ferroelectric quartz material sampled above, including: ferroelectric PZT (∼1 cm thick), PPLN (∼1 mm thick), and a second non-ferroelectric material ZnO (∼1 mm thick) (results in ESI†). In all three cases, the technique also worked for minimizing the electrostatic effect on tip response. Only ZnO gave responses indicating a large electrostatic effect from the measured response (see ESI†). The testing of PZT and PPLN mirrored the results of quartz, where the slopes of the fits are shallow, but present higher baseline piezo response. These results confirm that soft-molecule based piezoelectric materials are fundamentally different from classical ceramic based materials and must be analyzed with new methods that allow for operation at higher signal to noise ratios while simultaneously removing the electrostatic component of the response. This has been demonstrated to be achievable by alternatively sweeping the VDC instead of the VAC and finding the point of inflection where the VDC is equal to the VCPD and extracting the deff from that point instead of the slope of the fit.
We find incredibly large electromechanical tip responses, nearing 250 pm V−1, which derive from large differential capacitance of the films rather than the innate piezoresponse. This large electrostatic component from organic monolayers is in stark contrast to a range of inorganic materials studied, which may show greater intrinsic piezoresponse, but much lower electrostatic components. We speculate that while the organic monolayers have lower dielectric constants than piezo ceramics such as PZT, the differential capacitance is high due to their lower elastic modulus and thin layer thickness (e.g., ∼2 nm).
The new method of DC-sweep DART-PFM was used to determine the deff piezoresponse and electrostatic components of four organic monolayers and four conventional inorganic piezo materials. The method finds peptide and peptoid SAMs with both positive and negative piezo response and, coefficients in agreement with previously reported values.16 Control molecular SAMs composed of DDT and MUA show close to zero piezoresponse. While scans across multiple films and different AC voltages do affect the measurement somewhat, the DC-sweep DART-PFM technique shows much improved reproducibility relative to previous efforts using varied AC voltages with DART-PFM.
We believe this new technique will improve accurate measurements of electromechanical response in organic and biomaterials. Moreover, the large electrostatic component of electromechanical response found in organic materials can likely be utilized for sensing or other applications.
Footnote |
† Electronic supplementary information (ESI) available: Effect of kl on material response plots, example histogram of new method on non-fixed polar materials, AMFM results, figures and tabulated results for PZT, PPLN, and ZnO recorded piezoelectric response with new DC field sweeping method, and Kelvin probe results. See DOI: 10.1039/c9na00638a |