aUnconventional Computing Laboratory, University of the West of England, Bristol, BS161QY, UK. E-mail: Tom.Draper@uwe.ac.uk
bDepartment of Engineering Design and Mathematics, Faculty of the Environment and Technology, University of the West of England, Bristol, BS161QY, UK
cDepartment of Applied Sciences, Faculty of Health and Applied Sciences, University of the West of England, Bristol, BS161QY, UK
dInstitute of Biosensing Technology, Centre for Research in Biosciences, University of the West of England, Bristol, BS161QY, UK
First published on 21st November 2019
Liquid marbles (LMs) are of growing interest in many fields, including microfluidics, microreactors, sensors, and signal carriers. The generation of LMs is generally performed manually, although there has recently been a burst of publications involving ‘automatic marble makers’. The characteristics of a LM is dependent on many things, including how it is generated, it is therefore important to be able to characterise LMs once made. Here is presented a novel contactless LM sensor, constructed on a PCB board with a comb-like structure of 36 interlacing electrical traces, 100 μm wide and 100 μm apart. This cheap, scalable, and easy to use sensor exploits the inherent impedance (comprised of the electrical resistance, capacitive reactance and inductive reactance) of different LMs. With it, parameters of a LM can be easily determined, without interfering with the LM. These parameters are (1) particle size of the LM coating, (2) the concentration of a NaCl solution used as the LM core, and (3) the volume of the LM. Additionally, due to the comb-like nature of the sensor, the accurate positioning (down to the inter-trace spacing) of the LM can be ascertained. The new sensor has been shown to work under both static and dynamic (mobile) conditions. The capacitance of a LM was recorded to be 0.10 pF, which compares well with the calculated value of 0.12 pF.
Liquid marbles (LMs) are an increasingly popular1–5 modification to the droplet. Initially reported by Aussillous and Quéré in 2001,6 they are comprised of a droplet of liquid, surrounded by hydrophobic particles adhered to the surface of the droplet. By having such an ‘armoured droplet’, the effects of wetting are nullified. A LM can freely move around on a typically wetting surface, with minimal hindrance or hysteresis, and no loss of volume due to adhesion. This provides many advantages to microfluidics,7 and creates a great many other applications for LMs. These include rapid blood-typing assays,8 unconventional computation signals,9 and control of micro-encapsulated reactions.10,11
LMs provide additional advantages in microfluidics. Due to the powder coating on the surface of the droplet, the evaporation rate of the liquid core can be reduced,12–15 providing ease and flexibility to experimental setups. Resultantly the oil phase, traditionally present in microfluidic systems, can also be removed as it is no longer required for heat dissipation during reactions. This helps to remove any cross-contamination that may have occurred through the oil phase, as well as reducing the energy required for heating.
The development of microfluidics has enabled massive reductions in the quantity of both reagents and waste, helping to reduce costs and keep chemistry greener. Movement of droplets through digital microfluidic devices16 is normally achieved through electrowetting on dielectric (EWOD),17 surface acoustic waves (SAW),18 or thermocapillary actuation.19 Once the droplet has been relocated, sensing the new position of the droplet is required. Capacitance measurement provides for a means of achieving this in a non-invasive manner20–23 (though other techniques, such as labelling,24 are also used).
Microfluidic systems using LMs can also use EWOD for actuation,25,26 though due to the low hysteresis and lack of wetting the marbles can also roll under gravity.27 The coalescence of two LMs can be controlled through many techniques, such as collisions,28–30 acoustic waves,31 or electrical potential.32–34 This enables the controlled initiation of reactions. Continued efforts to enable complete automation of such devices necessitates the ability to accurately determine the positions of LMs.
We report on the first use of impedance and capacitance to accurately and reliably determine the presence and properties of a LM. Using a custom-made printed circuit board (PCB) and a high frequency AC supply, the voltage drop across an in-series shunt resister provides the necessary information. The purpose of the device is to detect the location of a LM, whilst simultaneously providing quantifiable information on some of the LM's properties. Such detection is increasingly required as the complexity of the paths and/or interactions between LMs increase. Whilst it is sometimes possible to use high-speed cameras for this purpose, optical tracking can be very expensive and doesn't scale easily, as the LM paths can become increasingly obscured in large 3D structures. We propose that this system will be useful in a range of fields, from digital microfluidics (e.g. for the real-time monitoring of LM location and properties) to unconventional computing (e.g. for determining if a billiard ball-style collision has resulted in a coalesced LM of a larger volume).
The main experimental system was designed such that it would be small, cheap and scalable. The sensor itself is an array of two sets of 18 traces (small copper tracks) on a PCB, shown in Fig. 1(a), with each set of 18 traces connected to each other to maximise sensitivity. A range of geometries and forms were considered for the electrode arrays. The area of direct contact between the marble's powder coating and surface of electrode array depends on several factors in particular the volume of the marble. For 20 μl volume the area of direct contact is approximately 3.5 mm2, therefore the resolution of the electrode array needed to be fine, as the influence of changing relative permittivity varies with the inverse square of distance. Consequently the electrode conductors, and gaps between them, were made as narrow as practical. Using ‘standard’ PCB manufacturing facilities provided 100 μm traces with 100 μm gaps (on a two layer board). Alternative methods of forming the conductors (such as printing conductive inks) were investigated but they didn't offer substantial benefits over PCBs. To minimise electrical noise (from external sources) a ground plane was located on the bottom layer of the PCB (which connected to ground on instrumentation) and the fibreglass (FR4) middle portion of the board was made as thin as practical (0.4 mm). The width of the traces from the electrode array were tapered to reduce electrical resistance and tight bends were avoided to reduce impedance at higher frequencies.
The thinness of the board (0.4 mm) meant that it had a degree of flexibility. We exploited this by curving the PCB, so that the sensor grid was at the bottom and the electrical connectors were at the top (i.e. the electrical traces ran down either side of the slope). This shape was maintained through use of a custom made 3D-printed support, visible in Fig. 1(b). The new curvature of the PCB means that the LM is always centred to the sensor, along one axis.
The small insulated gaps between the traces mean that there is a large resistance across the circuit under DC, however under AC the system acts like a capacitor. This can be easily confirmed by application of a square wave to the system and monitoring via an oscilloscope, resulting in the characteristic curved profile.
The sensing grid was positioned in series with a high-accuracy resistor (Vishay Z foil, 10 kΩ, tolerance 0.01%, temperature coefficient 2 mΩ °C−1, inductance <0.08 μH). The smaller the value of the resistor, the smaller the potential drop across it, which in turn means a greater potential across the sensor, increasing sensitivity. Conversely, a larger valued resistor gives a greater potential drop, which increases the accuracy of the voltage measuring device (oscilloscope in our case). This value of the resistor was a good compromise between both extremes.
When a high frequency AC potential was applied (20 V, 10 MHz), a current flowed through the system and the applied voltage was split across the two components. The presence or absence of a LM on the sensing grid resulted in a change in the capacitance of the sensor, and therefore a change in impedance. This change in impedance caused a change in the voltage divide, and therefore the potential across the resistor. The potential across the resistor could be accurately measured using a digital oscilloscope. Sinusoidal and square waveforms were generated using a BK Precision 4053B waveform generator. Static measurements were taken using an Iso-Tech IDS 6072A-U digital oscilloscope linked using a ‘serial over USB’ connection, and interfaced using a custom script in MatLab R2018a (MathWorks). The script is available in the ESI.†
When taking static measurements, the system measures the amplitude voltage (Vamp) across Rext 20 times in quick succession (0.22 s per individual measurement), which is then averaged to give the desired value with a typical standard deviation of just 0.2%. Total time is 4.4 s for full 20 measurements and averaging. The outcome of this value is dependent on the presence (and properties) or absence of a LM on the sensor.
Dynamic measurements were conducted using the same set-up as the static measurements, but with a Rigol MSO5074 oscilloscope (8 GSa s−1, 200 MPt memory). Measurements were performed in realtime with a sample rate of 100 MSa s−1, which provides 10 data points per waveform with the applied 10 MHz signal, and so is greater than the Nyquist rate. Recorded waveforms were processed using the MatLab script provided in the ESI.† The signal was averaged over 250 μs, for smoothing.
Location data was obtained by energising individual pins on one side of the sensor, whilst connecting all 18 on the other side to a common ground. The outermost pins were not used, to prevent data skewing due to their asymmetric nature. The pins were individually subjected to 20 Vp–p at 10 MHz, both in the presence and absence of a 30 μL PTFE LM made with a 25 wt% NaCl aqueous solution (a typical of example of which can be seen in Fig. 1(c)). A total of 10 measurements were made for each pin under each condition (each measurement is made of 20 averaged measurements).
The parasitic capacitance of the system, and that with a LM present, was measured through application of a square wave. The current was monitored by measuring the potential drop across the series resistor Rext, using it as a shunt resistor. The potential across the sensor was used to measure the charging potential. A 30 μL PTFE LM, containing a 25 wt% NaCl aqueous solution was used.
The contact area of DIW PTFE LMs of varying volumes was determined using light microscopy. The LMs were made by hand, before transferring to a hemocytometer (a slide with graduated marks of known dimensions, improved Neubauer pattern). The LM was observed via a Zeiss Axiovert 200M inverted light microscope through a 10× objective lens, providing a depth of field of approximately 8.5 μm. The region of contact of the LM was identified by varying the focal length of the image. Multiple images were recorded using a ThorLabs 340M monochromatic CCD camera and ThorCam 3.3.1 software, and combined using the image-stitching plugin in ImageJ in order to gain a wider horizontal field of view than achievable with a single image.35–37 The diameter of the contact zone was measured, and compared to the known graduations of the hemocytometer. Each LM volume was measured in triplicate. The contact area could then be calculated using circle theory.
(1) |
As explanation, if the LM sensor is considered to be acting purely as an impedance Zm, the circuit can be considered to be acting as a DC potential divider, and therefore would follow (2); where Vs is the applied voltage, Vext is the voltage across the external resistor, Zm is the impedance of the LM and PCB wires, and Rext is the resistor.
(2) |
(3) |
The value of Zm calculated using eqn (3) — rearranged from eqn (2) — includes both the LM and parasitic PCB impedance. In order to remove the PCB impedance, separate measurements were taken: (a) with a short-circuit sensor, and (b) with open-circuit sensor without a LM on the sensor. The impedance of the LM can therefore be given by:
(4) |
It is possible to calculate capacitance using the plots in Fig. 3 by measuring the time taken to reach one time constant: the time taken to reach 63.2% of the capacitors full charge. This can then be used to calculate the capacitance of the system using C = t/R, where t is the time constant. The capacitance of the empty system is the parasitic capacitance, Cpara, which when subtracted from the capacitance of the measured LM system Cm gives the capacitance of the LM in the system, CLM: such that Cm = Cpara + CLM. Note that Cpara = Cww + Cwg, and CLM = Ccoat/2. The measured value of capacitance was CLM = 0.10 pF. This value was compared to a simulated system, where twin capacitors were considered in series. By knowing the electrode contact area, it is possible to calculate the expected capacitance using eqn (5), where εr is the relative permittivity (dielectric constant) of the powder coating, ε0 is the permittivity of free space, Ae is the electrode plate area, and dc is the thickness of the LM's powder coating. The integer divisor is required as the system acts as two capacitors in series: first going across the powder coating into the LM, and second going back across the coating out of the LM.
(5) |
The calculated value for the system is 0.12 pF, which is in good agreement with the measured value. It is worth noting that an equivalent simplified circuit model can be envisaged, by replacing the twin capacitors with a single capacitor that has double the coating particle diameter (spacing between electrodes).
Importantly, the sensor functions as desired for all tested particle sizes and volumes. There is minimal variance between repeated tests of the same particle sizes, though there is variation across different particle sizes. Despite the slight polydisperse nature of the powder, a clear trend can be seen: as the particle size of the LMs powder coating increases, the signal decreases. This is due to the affected impedance of the capacitive sensor. In the absence of a LM, the applied AC current passes through the parasitic capacitance of the electrodes which results in a larger voltage drop across the sensor. When a LM is placed on top of the sensor, a new route is available. The AC current can pass up into the LM, through the core, and back down to the grounded side of the circuit, thus completing it. This new pathway is lower in impedance than the previous path, and so the voltage drop across the capacitor decreases. This causes a consequent increase of potential across the resistor, which is measured using an oscilloscope. The larger LMs cause a greater signal amplitude, which is to be expected, as the larger LMs cover more of the sensor grid. This is discussed fully in section 4.4.
The trend in signal amplitude is due to the different LM coating thickness, which relates to particle size. As described in section 3, the coating around LM can be modelled as a capacitance (Ccoat) and it is inversely proportional to the coating thickness. As the coating thickness increases, Ccoat decreases (see eqn (5)). Therefore, according to eqn (1), ZLM increases. Due to eqn (2), the increasing ZLM reduces the signal amplitude (Vext) which is measured across Rext.
It is interesting to note that the PTFE LMs (hollow markers in Fig. 4(a)) are in line with the PE LMs. The larger signal amplitude is due to the smaller particle size of the PTFE powder, and so the reduced coating. This suggests that the powder coating is not greatly interacting with the sensor. There are two reasons for this: firstly, relative permittivities of PE and PTFE are very similar;38 and secondly, the powder coating of the LM occupies a small amount of the current pathway, approximately 100 μm twice, compared to hundreds of microns through the core of the LM, as the current has many different grounding points.
For the 30 μL LMs (red squares), there is a clear strong relationship between the NaCl concentration inside the LM, and the amplitude of the signal. A linear relationship can be extrapolated, yielding the equation V = 4.97 × 10−4·[NaCl] + 0.322. This function has been plotted on the graph as a solid red line, and has an R-squared of 0.996. Such a relationship is expected, as the conductivity of NaCl in water also increases proportionally with concentration, due to its complete dissociation when solvated.38 It therefore follows the proportionality of resistance, R ∞ l/σA, where l is the length of the ‘resistor’, σ is its conductivity, and A is its cross sectional area. Due to these effects, ZNaCl reduces as the NaCl concentration increases the reduction in ZNaCl reduces ZLM and so increases Vext, the signal amplitude.
The larger LMs (blue triangles) also follow a similar linear trend, with the equation V = 5.55 × 10−4·[NaCl] + 0.331. This function has also been plotted on the graph, as a blue dashed line, and has an R-squared of 0.977. The larger LMs result in a greater signal amplitude than the smaller LMs, this is because the larger LMs cover more of the sensor grid. This is discussed fully in section 4.4.
It should be noted that changes in the salt concentration would have an effect on the surface tension of the aqueous core,39 which would in turn affected the shape of the LM. However, any such change in the LM shape (and ensuing contact area with the sensor) are accounted for. As the relationship shown in Fig. 4(b) uses empirical data, any changes in LM shape are incorporated by design.
This relationship makes it possible to not only use the sensor to detect the presence or absence of a LM, but also to gain useful characteristic data on it. This has many applications, including potential determination of the concentration of the cargo inside a LM, without interfering with said cargo, which is useful in microfluidics. For example, when the progress of a reaction needs to be monitored, but there is a desire to not directly interact with the reaction. Contactless determination of the LM's cargo concentration also allows for fast and clean measurements: there are no probes to place inside the LM, and so no time-consuming cleaning after the measurement, and therefore no risk of contamination with the next LM to be tested.
It can be seen that there is a strong positive relationship between the volume of the LM and the signal amplitude: the signal amplitude increases with increasing volume of the liquid marble. This is true for both the DIW (blue triangles) and saturated NaCl (red squares) LMs. This relationship is due to two main factors: the increase in amount of the LM in contact with the sensor and the reduction of ZNaCl. The LM, being soft, does not rest on a spherical bottom point, but instead sits on a flattened out base, visible in Fig. 1(c). It is this base which increases with size as the LM volume is increased. As the base increases in size, additional electrical traces in the sensor are in contact with the LM, allowing more current to pass into and exit the LM. Further, since there is more liquid volume, the cross-sectional area for the current path is significantly higher compared to smaller LMs. These two factors lower the impedance of the sensor and thereby its potential drop, again due to the proportionality of resistance.
As expected, and as previously explained in section 4.3, the saturated NaCl(aq) LMs have a greater signal amplitude at a fixed volume. The gradient is slightly different between the two cores however: 5.01 × 10−4 for DIW and 5.58 × 10−4 for NaCl(aq). This is explained by the greater conductivity of the NaCl(aq) LMs permitting more effective use of the entire of the cross-section of the LM, as opposed to DIW LMs whose lower core conductivity restricts effective signal conduction to regions closer to the bottom of the LM (and thereby closer to the sensor traces).
The phenomenon of maximal LM height plays a role in the linear nature of the plot in Fig. 4(c). As the volume of the LM increases, it reaches a maximal height in a logarithmic manner,40 resultantly the breadth of the LM increases disproportionally with its volume (a cartoon demonstration is shown in Fig. 5(a)). The contact radius of variously sized LMs was measured using light microscopy by observing where the powder particles moved out of the focal plane (as explained in the Experimental section), an example image of which is shown in Fig. 5(b). The contact area was then calculated using circle theory and plotted against LM volume, shown in Fig. 5(c).
Fig. 5 (a) A sketch of a LM resting on the sensor, demonstrating how the height of an expanding LM plateaus as the volume is increased, causing the breadth of a LM to increase disproportionately. Not to scale. (b) A collection of 9 light microscope images stitched together, providing a ‘bottom-up’ view of a LM. This facilitated measurement of the contact diameter of the LM (a 10 μL PTFE DIW LM in the example image). The contact diameter can be determined by observing where the powder particles move out of the focal plane. The scale bar is 500 μm. (c) The contact area of a LM against its volume. Error bars indicate the mean absolute deviation. LMs had a PTFE powder coating and a deionised water core. The dashed red line indicates the positive growth phase where the contact area is proportional to V4/3, and the dotted blue line indicates the region where the LMs maximal height results in the contact area becoming linearly proportional to V. Both fittings were first introduced for LMs by Aussillous and Quéré.1,6 |
The contact area of differently sized LMs was first discussed by Aussillous and Quéré in 2001, and then again in 2006.1,6 They noted that for smaller LMs the contact radius, l, could be defined as . Where R is the radius of the idealised spherical droplet; and the capillary length, λ, is defined as , where γ is the surface tension, ρ is the density, and g is gravitational acceleration. When related to contact area and LM volume, V, a relationship is revealed where the contact area is proportional to V4/3. This has been added to Fig. 5(c) as a red dashed line.
Once the volume of a LM is above a certain threshold, the mass of the liquid causes the height to plateau, typically at twice the capillary length. At this stage, as also reported by Aussillous and Quéré,1,6 the contact area increase is no longer polynomial. The contact radius of a larger LM can therefore be defined by . When this is related to contact area and LM volume, it reveals that the contact area is linearly proportional to volume. This relationship is displayed in Fig. 5(c) as a blue dotted line. It is this linear relationship that dominates the volume-sensor response plot in Fig. 4(c).
Measurement resolution of the device can be further improved by applying higher voltages, higher frequencies or a smoother sinusoidal waveform, as the capacitive reactance of the components is inversely proportional to the frequency of the applied signal, and therefore decreases. The frequency can be increased significantly, for example CPUs operating at 100 MHz have been available for decades. Testing has also revealed that the voltage can be greatly increased using the current sensor setup. On connecting two adjacent pins (100 μm apart) and ramping the DC potential, no breakdown (indicated by a drop in resistance) of the PCB was detected, within the limits of the SourceMeter (210 V, >200 MΩ). Therefore, it is possible to increase the voltage to 200 Vp–p and the frequency to 100 MHz, resulting in an increase of two orders of magnitude on excitation.
As long as the sensor grid is wider than the contact length of the LM, then it is possible to calculate the speed of the LM, from the breadth of the signal peak. The leading edge of the LM entering the sensor causes the signal to start to increase, and the leading edge of the LM leaving the sensor causes the signal to begin to decrease. For peaks of this style, the start of the peak rise to the start of the peak decrease is equivalent to the full-width at half-maximum (FWHM), which provides a time. By knowing the width of the sensor (7.1 mm) it is possible to calculate the speed of the LMs in the above plots. The LM in Fig. 6(a) had a speed of 2.0 m s−1, the first LM in Fig. 6(b) had a speed of 1.0 m s−1, whilst the second LM in that figure had a speed of 1.1 m s−1. This assumes a steady speed across the 7.1 mm of the sensor.
In the billiard-ball style of computing44–46 it is possible to fuse signals (LMs) in a collision, and thereby implement logic gate 〈x, y〉 → 〈y, xy, xȳ〉. The resulting larger volume of the new signal would be confirmation that such an operation had been performed. Conservative logic requires that the output signal be identical to the input, and so the design would need to be supplemented with a LM divider (as previously mentioned) to return the signal to its original volume. It would also be possible to implement either a more traditional billiard-ball gate or a Margolus logic gate — 〈x, y〉 → 〈y, xy, xy, xȳ〉 — where the signals collide and rebound, rather than fuse. In such a system, our sensor could not only be used to measure the size and speed of the colliding LMs, but an appropriately positioned sensor grid could also determine directional information of the rebounding marbles.
Computation can also be achieved by coalescing LMs with cores of different concentrations, to form a new combined core which has a concentration determined by the values of the original inputs. The properties of the new core (including concentration and size) can be measured using the techniques covered in this paper. This enables a new direction in computation to be explored. We performed a preliminary billiard-ball collision experiment by colliding two 20 μL PTFE LMs to yield a single 40 μL PTFE LM. The reported sensor successfully recorded the volumes of both the starting LMs, and the resulting larger LM — indicating a successful coalescence.
Capacitance of LMs could be applied in design of flexible and stretchable logical circuits47–49 and soft robotics,50 especially where disposable devices are used. Capacitive threshold logic51 is an extension of standard threshold logic, wherein logical values 1 and 0 are interpreted from the voltage being above or below a certain value. It can also be used to design capacitive synapse architecture.52,53 The easily variable capacitance of the LM sensor system (through the addition or removal of LM volume), permits straightforward tuning of the set-up.
It is worth noting that the evaporation of LMs would need to be controlled in a computing environment. Several previous studies have investigated the evaporation of LMs,12–15,54,55 and the rate has been shown to be related to several factors, including the powder coating, initial volume, and ambient humidity. Limiting the evaporation of LMs by maintaining a high-humidity environment has been previously demonstrated with success.56
The ability to automatically detect the volume of a LM in a non-destructive and contactless manner has many potential uses, especially in pharmaceutical manufacturing processes and biomedical diagnostic testing due to benefits associated with self-contained small sample volume bioreactors.57 It should be noted that this technique distinguishes between LMs and a naked droplet of equivalent mass (unlike a balance). There are now several papers that address the facile and automatic formation of LMs,9,30,58 and so having a method for easy confirmation that each LM formed is within required tolerance limits is certainly advantageous.