Guo-Cheng Han*a,
Huifang Lia,
Annaleizle Ferrancob,
Tao Zhana,
Yunyun Chenga,
Zhencheng Chena,
Mingyue Xuea,
Xiao-Zhen Feng*a and
Heinz-Bernhard Kraatz*b
aSchool of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin 541004, P. R. China. E-mail: hangc81@guet.edu.cn; fxz97118@guet.edu.cn
bDepartment of Physical and Environmental Sciences, University of Toronto Scarborough Campus, Toronto, Ontario M1C 1A4, Canada. E-mail: bernie.kraatz@utoronto.ca
First published on 22nd September 2020
Several simple sensors were fabricated through a one-step method. By depositing electro-active compounds, such as β-cyclodextrins (β-CD), heme, dopamine (DA), or Fc-ECG, onto a screen-printed electrode (SPE), the successful simultaneous detection of nitrite (NO2−) and thiosulfate (S2O32−) ions was observed. Under optimal operating conditions, the notable electrocatalytic abilities of a Heme/SPE sensor were detected for the oxidation of NO2− and S2O32−, with remarkable peak potential differences, after characterization via SEM, CV, and DPV. Linear relationships were obtained in the ranges of 5.0–200.0 μmol L−1 and 1.0–100.0 μmol L−1 for the current response versus concentration of NO2− and S2O32−, respectively. The limits of detection were determined to be 1.67 and 0.33 μmol L−1 while the sensitivities of detection were noted to be 0.43 and 1.43 μA μM−1 cm−2, respectively. During the detection of NO2− and S2O32−, no interfering common ions were observed. Furthermore, average recoveries from 96.0 to 104.3% and a total R.S.D. of less than 3.1% were found for the detection of NO2− and S2O32− in pickled juice and tap water using the simple sensor. These results showed that rapid and precise measurements for actual application in NO2− and S2O32− detection could be conducted in food samples, indicating a potential use in food safety.
Sodium thiosulfate (Na2S2O3) can be produced by thiosulfate (S2O32−) and sodium metal. S2O32− can also be used for applications in medical treatment, such as the detoxification of cyanide. Thiocyanase, which is found in the body, can combine S2O32− with the toxic cyanide ion to form the nontoxic thiocyanate which can subsequently be discharged through urine, preventing cyanide from harming the human body.7 There are other uses for S2O32− other than metal detoxification and gold and/or silver leaching, including the treatment of some skin diseases. It is also widely used in biochemistry, environmental treatment, and other fields.8,9
NO2− and S2O32− can affect the human body: excessive intake of NO2− can lead to death, but S2O32− can also be used in the detoxification of cyanide. So how to detect the mixture of NO2− and S2O32− accurately and sensitively has a certain significance for food safety and human health.10,11
Numerous analytical techniques are used for the detection of NO2− and S2O32− with good sensitivity. Some of the common hyphenated techniques include gas chromatography-mass spectrometry (GC-MS),12,13 liquid chromatography-mass spectrometry (LC-MS),14,15 HPLC,16 fluorescence detection,17 and electrochemical methods.18
Electrochemical techniques are ideal analytical and detection methods because of their sensitive response, rapid detection, simple operation, low cost, and portability.19,20 The main mode of detection is based on the oxidation of both NO2− and S2O32− on the surface of the electrode or the reaction with a probe that is capable of enhancing the electrochemical signal. For electrochemical detection, something to consider when creating a modified electrode is that it should be able to possess a charge transport capability in order to acquire a strong response. Obtaining a suitable catalyst-modified electrode should lead to enhanced catalytic activity for NO2− and S2O32− detection. Shabani-Nooshabadi et al. reported a sensor for 4-chlorophenol that used a graphene oxide NiO/NPs-ionic liquid (n-hexyl-3-methylimidazolium hexafluoro phosphate) carbon paste electrode. The proposed sensor was used for the determination of 4-chlorophenol in the presence of nitrite. It exhibited a strong enhancement effect on 4-chlorophenol and showed superior catalytic activity towards the electro-oxidation of 4-chlorophenol.21 Shabani-Nooshabadi et al. were able to analyze thiosulfate, 4-chlorophenol and nitrite in mixed samples using a voltammetric sensor. Through a combination of the highly conductive properties of NiO nanoparticles and a novel catechol derivative (2,4-dimethyl-N′-[1-(2,3-dihydroxyphenyl)methylidene]aniline) (2,4-DDMA) that has excellent mediator properties, the simultaneous analysis of thiosulfate, 4-chlorophenol and nitrite was achieved with a limit of detection of 0.01 μM for thiosulfate, of 0.7 μM for 4-chlorophenol, and of 5.0 μM for nitrite.18 Jahan Bakhsh Raoof et al. fabricated a 2,7-bis(ferrocenylethyl)-fluoren-9-one-modified carbon paste electrode (2,7-BFEFMCPE) for the electrocatalytic oxidation of thiosulfate in which a detection limit of 0.15 μM was obtained.22 All the reported compounds were used as catalysis materials, which possess an electrocatalytic property, but they were very complex compounds that cannot be obtained easily.
In this article, we report a facile process for the construction of a modified screen-printed electrode (SPE) by the direct deposition of easily obtained electro-active materials, including β-cyclodextrins (β-CD), heme, dopamine (DA), or Fc-ECG, onto SPE for the simultaneous detection of NO2− and S2O32−. Easily fabricated sensors, β-CD/SPE, Heme/SPE, DA/SPE and Fc-ECG/SPE, were acquired. Nonetheless, the results indicated that the Heme/SPE sensor displayed the highest oxidation peak current with remarkable peak potential differences for the simultaneous detection of NO2− and S2O32− when compared to the other sensors. The presence of heme on SPE can improve the electrocatalytic activity of the bare electrode. Investigations with Heme/SPE were conducted to determine the most effective concentrations of NO2− and S2O32−. Further investigations using the proposed method were also undertaken to detect NO2− and S2O32− in pickled juice and tap water, which led to satisfactory results.
A three-electrode system was used to conduct electrochemical analyses, where the counter, reference and working electrodes were a carbon electrode, an Ag wire, and the modified carbon electrode (φ = 3 mm), respectively. The obtained data were acquired using a CHI660E electrochemical workstation (Shanghai Chenhua Instrument Co., Ltd., China), which was attached to the three-electrode system for testing. The CV experiments were conducted in 0.1 mol L−1 PBS as the supporting electrolyte with a 100 mV s−1 scan rate and the parameters were set to scan from 0 to 1200 mV. The DPV experiments were conducted and the parameters were set to scan in the range from 0 to 1000 mV and 50 mV as the pulse amplitude. In addition, EIS measurements were conducted using a frequency range of 100000 to 0.1 Hz, an amplitude of 5 mV, an initial voltage of 300 mV, and with open circuit potentials. An initial voltage of 100 mV was recorded for the electrochemical i–t curve.
Scheme 1 The fabrication of the Heme/SPE sensor by the DPV technique for the detection of NO2− and S2O32−. |
Fig. 1 SEM images of different modified electrodes ((A): naked SPE; (B): β-CD/SPE; (C): DA/SPE; (D): Fc-ECG/SPE; (E): Heme/SPE). |
A significant number of rough amorphous carbon particles were observed on the surface morphology of naked SPE, as shown in Fig. 1A, and all the pure carbon particles look very smooth, like a platform. After separately depositing four different types of electro-active materials, β-CD, DA, Fc-ECG, and heme, onto SPE, each modified electrode appeared to be dissimilar. As demonstrated in Fig. 1B, multiple flower buds could be observed on the electrode surface without uniformity, and it looked very different from the carbon particles. In contrast, bamboo clusters, sparse dendritic rods, and strip stones coated the surfaces of the electrodes shown in Fig. 1C, D and E when the electro-active materials DA, Fc-ECG and heme were deposited, respectively. Through one-step deposition, the modified electrode surfaces differ from each other. In addition, successful coating of the electro-active materials was observed for each SPE surface.
After surface morphology characterization of the as-prepared sensors, the report subsequently directed its focus towards the electro-active response property for potential applications. The DPV results shown in Fig. 2 were measured in 0.1 mol L−1 PBS (pH 7.0) and DPV curves are displayed for SPE, β-CD/SPE, DA/SPE, Fc-ECG/SPE and Heme/SPE (A). Additionally, Fig. 2 also shows EIS curves of the naked and modified electrodes in 5.0 mmol L−1 potassium ferricyanide solution present in 0.1 mol L−1 KCl as the electrolyte (B).
In Fig. 2A, curves (a) and (b) display an almost straight line without an electrocatalytic response. It is apparent in Fig. 2A that the other curves were all observed to possess electrocatalytic responses from DA/SPE, Fc-ECG/SPE and Heme/SPE in PBS. PBS was used as the blank when DA, Fc-ECG and heme modified electrodes were examined and exhibited positive oxidation peaks at 0.15, 0.30 and 0.70 V, with 0.98, 1.72 and 3.25 μA weak peak current responses, respectively, which corresponded to their internal electrocatalytic performance. Fig. 2B displays common EIS curves for each individual sensor examined in a solution of potassium ferricyanide. Sensors DA/SPE and Fc-ECG/SPE showed relatively large impedances (Fig. 2B(c and d)), indicating that the electron transfer was hampered in the electrochemical process for each modified electrode, suggesting that the medium in which DA and Fc-ECG were present shows low electro-activity in ferricyanide solution. On the other hand, after depositing electro-active materials β-CD and heme on the naked electrode, the impedances remain almost the same, indicating that electron transfer was not significantly hindered compared to sensors DA/SPE and Fc-ECG/SPE. By comparing the electron transfer ability of the modified electrodes, it was observed that both β-CD/SPE and Heme/SPE sensors showed ideal electronic transmitting abilities.
In Fig. 3A and B, an electrochemical signal for the oxidation of two separate solutions containing NO2− and S2O32− was observed on the surfaces of five sensors in the range of 0.6–0.8 and 0.8–1.0 V, respectively, but not one sensor showed remarkable current responses toward the two analytes. In Fig. 3C, similar values for NO2− and S2O32− were observed since only one oxidation peak appeared when investigating the detection of NO2− and S2O32− mixed solutions using each of the five sensors: naked SPE, β-CD/SPE, DA/SPE, Fc-ECG/SPE, and Heme/SPE. Therefore, none of the modified sensors was used to measure the detection of NO2− and S2O32− by the CV technique with lower sensitivity since two distinguishable signals were not observed. Consequently, DPV was employed to detect NO2− and S2O32− in a solution mixture. Fig. 3D and E indicate that no remarkable current response appears for two separate solutions containing either NO2− or S2O32−. In a mixed solution containing NO2− and S2O32−, the following sensors, naked SPE, β-CD/SPE, DA/SPE, Fc-ECG/SPE, still showed one oxidation of two ions to confirm no catalysis effect from the medium towards oxidation of NO2− and S2O32−. However, it was exciting that two separated peaks were observed on the Heme/SPE sensor (curve (e) in Fig. 3F). The DPV results showed separate electrochemical signals for the oxidation of each of NO2− and S2O32− in a solution mixture. An enhancement in the current response was also observed for the Heme/SPE sensor compared to the other sensors. Therefore, the Heme/SPE sensor was revealed to be the most suitable sensor for NO2− and S2O32− catalysis due to heme with its compact rod shapes possessing a larger surface area and more active sites which can promote the charge transfer on the electrode. Our group previously reported that the Heme/SPE sensor enhanced current responses for UA, AA, DA and BPA catalysis due to the electrocatalytic ability of heme.23,24 As a result, Heme/SPE exhibits advantageous catalytic capabilities for oxidation of NO2− and S2O32−. Additional analyses examining the ability for detection of NO2− and S2O32− were consequently undertaken.
The effects of deposition concentrations from 0.25 to 2.0 mM of the electro-active material heme for catalysis of NO2− and S2O32− are demonstrated in Fig. 4A. The observed results showed notable values that 1.0 mmol L−1 of heme led to the highest current response while the responses for the other concentrations that were investigated showed lower responses. As a result, 1.0 mmol L−1 concentration was selected for subsequent investigations. Based on these results, lower concentrations of heme showed that the electrode was incapable of suitably catalyzing NO2− and S2O32−. However, increased concentrations greater than 1.0 mmol L−1 heme led to ineffective electron transfer as the electrode became hampered. As shown in Fig. 4B, increased the deposition time of heme onto the electrode surface led to the current value increasing. The largest current response was observed at 200 s, indicating that this time was the value at which the most effective enrichment of heme would take place on the electrode surface. For that reason, the preferred preparation time for heme deposition was 200 s. An additional condition to investigate when establishing the ideal conditions for detection of NO2− and S2O32− is determining the ideal pH value. As demonstrated in Fig. 4C, at pH = 7.0, the Heme/SPE sensor exhibited the highest current responses for oxidation peaks of NO2− and S2O32−. At higher pH values, the shortage of protons caused decreased oxidation peak currents and the electrocatalytic oxidation of NO2− and S2O32− became affected. However, a noticeably decreased peak current was observed under lower pH conditions, indicating that more acidic environments caused the formation of nitrous acid and thiosulfuric acid, reducing the amount of NO2− and S2O32− available for oxidation. Lastly, two distinct oxidation peaks and maximum oxidation peak currents were observed at pH 7. Consequently, further electrochemical analyses were carried out at pH 7 as the ideal condition. As depicted in Fig. 4D, the temperature did not appear to remarkably affect the current response. Therefore, subsequent studies were investigated at 25 °C (room temperature) for oxidation of NO2− and S2O32−.
Fig. 5A shows the amperometric responses against different concentrations of NO2− and S2O32− mixed solutions with the Heme/SPE sensor and includes calibration plots of current response and concentrations. It can be observed that the Heme/SPE sensor gave two distinct oxidation peaks at 0.60 V and 0.80 V under optimal conditions with peak potential differences of 200 mV from NO2− and S2O32−. In addition, the current responses of the sensor increased with linearly when the NO2− and S2O32− concentration increased. Hence, the Heme/SPE sensor was entirely capable of the simultaneous detection of NO2− and S2O32−. Electrochemical oxidation peaks for NO2− and S2O32− were further enhanced as the concentration increased over a broad range and an established linear relationship was observed, as shown in Fig. 5B and C. A strong linear relationship was obtained between the NO2− concentration and the current: the linear equation was I = 4.05227 + 0.02909c (I is the current, and c is concentration) with the correlation coefficient determined to be 0.99714; the linear range was between 5.0 and 200.0 μmol L−1; the limit of detection was 1.67 nmol L−1, and the sensitivity of detection was 0.43 μA μM−1 cm−2. For analysis of S2O32− using the same method, the linear equation was I = 1.31585 + 0.10262c, with correlation coefficient 0.99783, the linear range was between 1.0 and 100.0 μmol L−1 with 0.33 nmol L−1 limit of detection and 1.43 μA μM−1 cm−2 sensitivity of detection. Table 1 displays a comparison for the detection of NO2− and S2O32− by other modified electrodes. The method investigated in this report was made simpler based on the construction of the one-step fabricated electrochemical sensor: by depositing heme, an easily obtained compound, onto the SPE for detection of NO2− and S2O32−.
Modified electrode | Analytical method | Detection limit (μM) | Linear range (μM) | Ref. | ||
---|---|---|---|---|---|---|
NO2− | S2O32− | NO2− | S2O32− | |||
2,4-DDMA-NiO/NP/CPE | SWV | 5.0 | 0.01 | 10.0–700.0 | 0.05–400.0 | 18 |
2,7-BFEFMCPE | DPV | 0.15 | 0.6–90.0 | 22 | ||
CdO/NPs/CNDPMA/CPE | SWV | 0.04 | 0.09–350 | 25 | ||
CdO/SWCNTs/MBIB/CPE | SWV | 0.07 | 0.1–900.0 | 26 | ||
CoPc/MWCNTs | DPV | 2.11 | 10.0–1.05 × 106 | 27 | ||
Heme/SPE | DPV | 1.67 | 0.33 | 5.0–200.0 | 1.0–100.0 | This work |
Analyte | Detected (μmol L−1) | Added (μmol L−1) | Average Found (μmol L−1) | Average Recovery (%) | RSD (%) | |||
---|---|---|---|---|---|---|---|---|
Average | Intra-assay | Inter-assay | ||||||
a — not detected. | ||||||||
Pickled juice | NO2 | 16.8 ± 0.3 | 20.0, 40.0, 60.0 | 35.7 ± 0.4, 57.2 ± 0.5, 77.8 ± 0.4 | 98.0, 100.7, 101.3 | 1.6, 3.2, 2.8 | 2.5 | 3.1 |
S2O32− | — | 20.0, 40.0, 60.0 | 19.2 ± 0.5, 38.7 ± 0.8, 58.9 ± 0.3 | 96.0, 96.7, 98.2 | 2.5, 2.9, 4.3 | 3.2 | ||
Tap water | NO2− | — | 20.0, 40.0, 60.0 | 19.5 ± 0.5, 39.9 ± 0.8, 61.9 ± 0.3 | 97.5, 99.8, 103.2 | 3.2, 4.1, 2.3 | 3.2 | |
S2O32− | — | 20.0, 40.0, 60.0 | 19.8 ± 0.4, 41.7 ± 0.6, 62.2 ± 0.4 | 99.0, 104.3, 103.7 | 2.7, 3.8, 4.1 | 3.5 |