Pouline M. P.
van Oort‡
a,
Iain R.
White‡
*bc,
Waqar
Ahmed
bd,
Craig
Johnson
b,
Jonathan
Bannard-Smith
e,
Timothy
Felton
ed,
Lieuwe D.
Bos
a,
Royston
Goodacre
f,
Paul
Dark
ed and
Stephen J.
Fowler
ed
aDepartment of Intensive Care, Amsterdam UMC – location Academic Medical Centre (AMC), Amsterdam, the Netherlands
bManchester Institute of Biotechnology, University of Manchester, Manchester, UK
cLaboratory for Environmental and Life Sciences, University of Nova Gorica, Nova Gorica, Slovenia
dDivision of Infection, Immunity and Respiratory Medicine, School of Biological Sciences, Faculty of Biology, Medicine and Health, and NIHR Manchester Biomedical Research Centre, University of Manchester, Manchester, UK
eManchester Academic Health Science Centre, Manchester University Hospitals NHS Foundation Trust, Manchester, UK. E-mail: stephen.fowler@manchester.ac.uk
fDepartment of Biochemistry, Institute of Integrative Biology, University of Liverpool, UK
First published on 21st October 2020
Exhaled breath analysis is a promising new diagnostic tool, but currently no standardised method for sampling is available in mechanically ventilated patients. We compared two breath sampling methods, first using an artificial ventilator circuit, then in “real life” in mechanically ventilated patients on the intensive care unit. In the laboratory circuit, a 24-component synthetic-breath volatile organic compound (VOC) mixture was injected into the system as air was sampled: (A) through a port on the exhalation limb of the circuit and (B) through a closed endo-bronchial suction catheter. Sorbent tubes were used to collect samples for analysis by thermal desorption-gas chromatography-mass spectrometry. Realistic mechanical ventilation rates and breath pressure–volume loops were established and method detection limits (MDLs) were calculated for all VOCs. Higher yields of VOCs were retrieved using the closed suction catheter; however, for several VOCs MDLs were compromised due to the background signal associated with plastic and rubber components in the catheters. Different brands of suction catheter were compared. Exhaled VOC data from 40 patient samples collected at two sites were then used to calculate the proportion of data analysed above the MDL. The relative performance of the two methods differed depending on the VOC under study and both methods showed sensitivity towards different exhaled VOCs. Furthermore, method performance differed depending on recruitment site, as the centres were equipped with different brands of respiratory equipment, an important consideration for the design of multicentre studies investigating exhaled VOCs in mechanically ventilated patients.
Gas chromatography-mass spectrometry (GC-MS) is the current gold standard for detection and identification of exhaled VOCs.3 GC-MS is not suitable for the bedside and so several breath sampling methods have been investigated to enable capture and transportation of samples off-site. Previous studies reported the use of glass syringes and gas sampling bags.14–16 Although they are low cost and have been used to capture breath gas in many studies, they are associated with loss of metabolites and sample contamination for example dimethylacetamide, phenol, 2-methyl-1,3-dioxalane.17–19 Stainless steel tubes filled with sorbent material pre-concentrate the breath sample and are suitable for transportation and storage with minimal loss of VOCs.20 Needle trap devices have also been used to trap VOCs.21,22 Currently there is no consensus regarding the ideal method for bedside breath sampling and transport although the need to develop a strategy to achieve this is widely accepted.23
Standardization of VOC sampling methods in mechanically ventilated patients remains challenging and is contingent on fully defining sampling systems in terms of their analytical performance, including dynamic range, detection limit, noise and the sensitivity of matrix effects towards ambient changes. Previously, two methods of VOC capture using different sampling sites within the ventilator circuit have shown to be safe and feasible: a semi-invasive method using a closed endo-bronchial suction catheter,11 and a completely non-invasive technique whereby samples are diverted through a port on the exhalation limb of the ventilator circuit.24 The aim of this study was to compare these two different breath sampling techniques on a circuit optimised for breath VOC collection, utilising a test-lung device as a model for patient settings in the ICU. We hypothesized that by sampling at a position further upstream (nearer to the patient) within the ventilator circuit, larger abundances of VOCs would be detectable as they are measured closer to their source of origin. In order to evaluate the results from the test circuit, detection limits are compared to VOC levels obtained from patients enrolled in the National Institute for Health Research (NIHR) BRAVo (BReath Analysis in Intensive Care: Proof of Concept for Non-Invasive Diagnosis of Ventilator Associated Pneumonia) project.
A schematic representing the ventilator circuit is displayed in Fig. 1. Two breath sampling methods were investigated: (A) a non-invasive technique collecting air from the external ventilator connection circuit based on ref. 20 and (B) a semi-invasive technique using a suction catheter inserted via the endotracheal tube based on ref. 11. Using a precision air-sampling pump (Escort ELF pump, Supelco, Dorset, UK), air samples of 1.2 L were collected at a flow of 0.5 L min−1 using 100% O2 in order to mimic ICU setting. Air was drawn through stainless steel tubes (0.25 inch outer diameter, 3.5 inch length) packed with 200 mg ± 1.7% of TenaxGR adsorbent (35/60, Markes International, Llantrisant, UK). TenaxGR was chosen for its hydrophobic properties.
For the semi-invasive sampling method B (Fig. 1, red circled B) “inspiratory” airflow followed the route (in consecutive order): air/O2 cylinder – air/O2 mechanical ventilator hoses (3 M air hose and 3 M O2 hose; MEC Medical Ltd, Hitchin, UK) – mechanical ventilator – corrugated tubing (Intersurgical) – T-piece connector (GE Healthcare Finland, Helsinki, Finland) – HME – closed suction catheter – endotracheal tube – small piece of corrugated tubing – test lung device. During the simulation of expiration and subsequent air sampling, the route was as follows: test lung device – small piece of corrugated tubing – endotracheal tube – closed suction catheter – PFA tubing (approximately 50 cm PFA-T2-030-100:1/8′′ OD, both ends connected to approximately 8 cm of PFA-T4-062-100:1/4′′ OD; Swagelok, Warrington, UK) – steel adsorbent tube. The inspiratory airflow route for the non-invasive method (Fig. 1, red circled A) was: air/O2 cylinder – air/O2 mechanical ventilator hoses – mechanical ventilator – corrugated tubing (Intersurgical) – T-piece connector (GE Healthcare Finland) – HME – endotracheal tube – small piece of corrugated tubing – test lung device. The expiratory route was: test lung device – small piece of corrugated tubing – endotracheal tube – HME – T-piece connector – PTFE tubing – steel adsorbent tube. Before attaching the steel adsorbent tubes and collecting air samples the apparatus was purged for 1 min at 0.5 L min−1. Experiments were performed at Salford Royal Hospital NHS Foundation Trust, UK.
In order to investigate the extent to which different materials give rise to analytical contamination, three brands of closed suction catheter were tested within the circuit as above. Filtered room air was purged through each system and collected onto sorbent tubes. Two of these catheters (catheter 2 and 3) were used to acquire breath from ICU patients at the two Manchester sites; these measurements therefore provided the background signal from which detection limits were ascertained.
Compound | Conc. in 1° stock/(mmol) | RMM/(g mol−1) | Quant ion | RT (min) | log(KOW) | k H/(atm m3 mol−1) | VP/(mmHg) |
---|---|---|---|---|---|---|---|
RMM = relative molecular mass; KOW = octanol/water partition coefficient; kH = Henry's law constant; VP = vapour pressure. KOW, KH and VP are experimentally derived values reported in Episuite v4.1.a KOWWIN v1.68 estimated value.b Estimate from bond method in HENRYWIN v3.2.c Mean of Antoine and Grain methods.d Modified Grain method. | |||||||
Acetone | 100 | 58 | 58 | 1.72 | −0.24 | 3.50 × 10−5 | 2.32 × 102 |
Isoprene | 100 | 68 | 67 | 1.73 | 2.42 | 7.67 × 10−2 | 5.50 × 102 |
Benzene | 25 | 78 | 78 | 2.4 | 2.13 | 5.55 × 10−3 | 9.48 × 101 |
3-Pentanone | 25 | 86 | 57 | 2.66 | 0.91 | 8.36 × 10−5 | 3.77 × 101 |
1,4-Dioxane | 55 | 88 | 88 | 2.77 | −0.27 | 4.80 × 10−6 | 3.81 × 101 |
Pyridine | 25 | 79 | 79 | 3.18 | 0.65 | 1.10 × 10−5 | 2.08 × 101 |
Toluene | 10 | 92 | 91 | 3.49 | 2.73 | 6.64 × 10−3 | 2.84 × 101 |
Octane | 25 | 114 | 85 | 3.94 | 5.18 | 3.21 × 100 | 1.41 × 101 |
Butyl acetate | 25 | 116 | 56 | 4.2 | 1.78 | 2.81 × 10−4 | 1.15 × 101 |
p-Xylene | 25 | 106 | 91 | 5.27 | 3.15 | 6.90 × 10−3 | 8.84 × 100 |
Nonane | 25 | 128 | 57 | 5.87 | 5.65 | 3.40 × 100 | 4.45 × 100 |
Benzaldehyde | 25 | 106 | 77 | 7.31 | 1.48 | 2.67 × 10−5 | 1.27 × 100 |
1-Heptanol | 25 | 116 | 70 | 7.51 | 2.62 | 1.88 × 10−5 | 2.34 × 10−1 |
Decane | 25 | 142 | 57 | 8.27 | 5.01 | 5.15 × 100 | 1.43 × 100 |
3-Carene | 10 | 136 | 93 | 8.56 | 4.38 | 1.07 × 10−1b | 2.09 × 100c |
Limonene | 25 | 136 | 68 | 9.03 | 4.38 | 3.19 × 10−2 | 1.44 × 100 |
Undecane | 10 | 156 | 57 | 10.83 | 5.74a | 1.93 × 100 | 4.12 × 10−1 |
Nonanal | 50 | 142 | 57 | 10.94 | 3.27b | 7.37 × 10−4 | 3.70 × 10−1 |
Tetralin | 10 | 132 | 104 | 12.46 | 3.49 | 1.36 × 10−3 | 3.68 × 10−1 |
Dodecane | 10 | 170 | 57 | 13.37 | 6.10 | 8.18 × 100 | 1.35 × 10−1 |
1-Methylindole | 10 | 131 | 130 | 15.17 | 2.72 | 1.89 × 105b | 4.71 × 10−2d |
Tridecane | 10 | 184 | 57 | 15.81 | 6.73 | 2.88 × 100 | 5.58 × 10−2 |
Tetradecane | 10 | 198 | 57 | 18.13 | 7.20 | 9.20 × 100 | 1.16 × 10−2 |
Pentadecane | 10 | 212 | 57 | 20.33 | 7.71a | 1.26 × 101 | 3.43 × 10−3 |
Both methods were used to sample breath at the two sampling sites: 1.2 L of breath was collected at a flow of 0.5 L min−1 onto TenaxGR adsorbent tubes (Markes International, Llantrisant, UK). Samples were collected in duplicate (i.e., 40 samples were submitted from each centre for each sampling method). After sampling, the tubes were returned to the Manchester Institute of Biotechnology, University of Manchester, UK, for analysis by TD-GC-MS.
Thermal desorption unit | Model | Markes International TD100 |
Cold trap model | U-T11GPC-2S general purpose carbon (Markes International) | |
Cold trap temp | 0 °C | |
Primary desorb | 280 °C for 5 min (splitless) | |
Secondary desorb | 280 °C for 3 min (splitless) | |
Chromatography instrument | Model | Agilent 7890B GC |
Column model | DB-5 ms GC (0.25 μm, 0.25 mm × 30 m; Agilent Technologies) | |
Carrier gas | Constant pressure (10 kPa) He | |
Oven program | Initial temperature of 40 °C; 170 °C at 6 °C min−1; 190 °C at 15 °C min−1 | |
Postrun | 250 °C (2 min) | |
Mass spectrometer | Model | Agilent 7010 series triple quadrupole mass spectrometer |
Source | EI+ at 70 eV (200 °C) | |
Scan m/z range | 40–500 (5 Hz) | |
Reconditioner | Model | Markes TC20 |
Carrier gas | 50 mL min−1 N2 | |
Oven program | 330 °C for 1 h | |
Dry purge (when required) | 50 mL min−1 N2 for 4 min |
A targeted deconvolution was performed on 0.1 min time windows surrounding the retention times presented in Table 1 and target quantifier ions were integrated using MassHunter's Agile2 integrator (Agilent Technologies, Stockport, UK).
VOCs associated with catheter 1 (% TIC) | RT/(min) | RIobs | RILit | Δ RT/(s) | |
---|---|---|---|---|---|
1 | Cyclohexanone (10%) | 5.8 | 895 | 891 (ref. 27) | 4 |
2 | 2-Ethyl-1-hexanol (7%) | 9.0 | 1029 | 1035 (ref. 28) | 9 |
3 | 1,3-Di-tert-butylbenzene (5%) | 14.8 | 1258 | 1249 (ref. 29) | 13 |
4 | Dodecamethylcyclopentasiloxane (4%) | 16.5 | 1329 | 1341 (ref. 30) | 17 |
5 | 3-Carene (3%) | 8.6 | 1011 | 1010 (ref. 31) | 2 |
6 | Decamethylcyclopentasiloxane (3%) | 12.2 | 1156 | 1165 (ref. 30) | 13 |
7 | p-Xylene (3%) | 5.3 | 868 | 865 (ref. 31) | 3 |
8 | α-Pinene (3%) | 6.7 | 934 | 939 (ref. 32) | 8 |
9 | 1,2-Dichloroethane (3%) | 2.3 | 648 | 649 (ref. 33) | 0 |
10 | 3,7-Dimethylnonane (2%) | 9.2 | 1035 | 1036 (ref. 34) | 2 |
VOCs associated with catheter 2 (% TIC) | RT/(min) | RIobs | RILit | Δ RT/(s) | |
---|---|---|---|---|---|
1 | Cyclohexanone (21%) | 5.8 | 894 | 891 (ref. 27) | 3 |
2 | Decamethylcyclopentasiloxane (9%) | 12.3 | 1158 | 1165 (ref. 30) | 11 |
3 | Dodecamethylcyclopentasiloxane (6%) | 16.5 | 1328 | 1341 (ref. 30) | 17 |
4 | 2-Ethyl-1-hexanol (6%) | 9.0 | 1030 | 1035 (ref. 28) | 8 |
5 | Tetradecamethylcycloheptasiloxane (5%) | 20.3 | 1500 | 1516 (ref. 30) | 21 |
6 | α-Pinene (4%) | 6.7 | 934 | 939 (ref. 32) | 7 |
7 | β-Pinene (4%) | 7.7 | 978 | 980 (ref. 35) | 3 |
8 | Ethyl acetate (3%) | 2.1 | 614 | 612 (ref. 31) | 1 |
9 | Isophorone (2%) | 11.4 | 1122 | 1124 (ref. 28) | 2 |
10 | p-Xylene (2%) | 5.3 | 868 | 865 (ref. 31) | 4 |
VOCs associated with catheter 3 (% TIC) | RT/(min) | RIobs | RILit | Δ RT/(s) | |
---|---|---|---|---|---|
The VOCs printed in italics are shared between the three suction catheters.a Estimated n alkane retention index reported in NIST v14. RIobs is the retention index calculated from observed retention times; RIlit is the retention index reported in literature; ΔRT is the difference between the observed retention time and the retention time calculated from RIlit. | |||||
1 | Dodecamethylcyclopentasiloxane (11%) | 16.5 | 1328 | 1341 (ref. 30) | 17 |
2 | Decamethylcyclopentasiloxane (11%) | 12.3 | 1158 | 1165 (ref. 30) | 11 |
3 | Tetradecamethylcycloheptasiloxane (10%) | 20.3 | 1500 | 1516 (ref. 30) | 21 |
4 | 2-Ethyl-1-hexanol (5%) | 9.0 | 1029 | 1035 (ref. 28) | 9 |
5 | 2,5-Dimethylhexane-2,5-dihydroperoxide (5%) | 17.3 | 1365 | 1367a | 3 |
6 | Cyclohexanone (5%) | 5.8 | 894 | 891 (ref. 27) | 6 |
7 | 2-Butanone (3%) | 2.0 | 601 | 602 (ref. 32) | 0 |
8 | p-Xylene (3%) | 5.3 | 868 | 865 (ref. 32) | 3 |
9 | Acetone (3%) | 1.7 | 500 | 500 (ref. 32) | 0 |
10 | Dodecane (3%) | 13.4 | 1200 | 1200 | 1 |
Compound | RIObs | RILit | Δ RT /(s) | S RI | S RI | Method detection limit/ppbV | ||
---|---|---|---|---|---|---|---|---|
Method B | Method A | Method B (site 1) | Method B (site 2) | Method A (both sites) | ||||
RIobs = retention indices calculated from breath samples; RIlit = literature retention index; ΔRT is the difference between the observed retention time and the retention time calculated from RIlit; SRI is the standard deviation in retention indices. | ||||||||
Acetone | 504 | 500 (ref. 31) | 1 | 8 | 6 | 6.28 | 5.60 | 1.25 |
Isoprene | 515 | 520 (ref. 36) | 1 | 9 | 7 | 0.10 | 0.22 | 0.21 |
Benzene | 660 | 657 (ref. 31) | 1 | 2 | 2 | 0.09 | 0.25 | 0.13 |
3-Pentanone | 700 | 689 (ref. 31) | 4 | 7 | 7 | 2.13 | 1.11 | 1.39 |
1,4-Dioxane | 707 | 705 (ref. 37) | 2 | 1 | 1 | 0.02 | 0.13 | 0.12 |
Pyridine | 740 | 736 (ref. 31) | 3 | 3 | 3 | 0.12 | 0.14 | 0.06 |
Toluene | 764 | 760 (ref. 31) | 3 | 0 | 1 | 0.17 | 0.19 | 0.14 |
Octane | 800 | 800 | 0.14 | 0.10 | 0.12 | |||
Butyl acetate | 813 | 811 (ref. 32) | 2 | 6 | 1 | 0.03 | 0.05 | 0.06 |
p-Xylene | 868 | 865 (ref. 31) | 3 | 29 | 0 | 2.80 | 4.20 | 0.48 |
Nonane | 900 | 900 | 0.09 | 0.03 | 0.13 | |||
Benzaldehyde | 961 | 960 (ref. 32) | 2 | 8 | 1 | 1.00 | 0.39 | 1.33 |
1-Heptanol | 969 | 969 (ref. 31) | 1 | 6 | 1 | 0.56 | 0.21 | 0.55 |
Decane | 1000 | 1000 | 0.35 | 2.36 | 0.32 | |||
3-Carene | 1011 | 1010 (ref. 31) | 1 | 1 | 1 | 3.46 | 1.29 | 0.13 |
Limonene | 1030 | 1031 (ref. 38) | 2 | 1 | 0 | 0.77 | 1.18 | 0.45 |
Undecane | 1100 | 1100 | 0.83 | 0.15 | 0.28 | |||
Nonanal | 1105 | 1105 (ref. 38) | 1 | 1 | 0 | 2.67 | 0.60 | 1.96 |
Tetralin | 1165 | 1163 (ref. 27) | 3 | 1 | 1 | 0.10 | 0.06 | 0.02 |
Dodecane | 1200 | 1200 | 0.11 | 0.80 | 0.30 | |||
1-Methyl indole | 1276 | 1273 (ref. 27) | 4 | 5 | 5 | 0.02 | 0.01 | 0.01 |
Tridecane | 1300 | 1300 | 0.11 | 0.03 | 0.49 | |||
Tetradecane | 1400 | 1400 | 0.08 | 0.13 | 0.32 | |||
Pentadecane | 1500 | 1500 | 0.01 | 0.01 | 0.09 |
Fig. 4 The proportion of results above the MDL for both sampling methods used to obtain breath samples from two ICUs (N = 40, 2 replicates at each site). |
Fig. 5 displays the proportion of the breath dataset above the MDL for all 24 compounds under study. At site 1, using method B to sample breath resulted in a significant increase in the number of compounds measured at detectable levels. Method A and method B are better suited to sampling different compounds however at site 2, the proportion of the data set above the MDL is approximately even. In this regard, the two sampling techniques complemented one another. A full evaluation of the suitability of these sampling methods towards sampling breath in mechanically ventilated patients may only be possible when target compounds have been identified and biomarkers indicative of infection are known.
Fig. 5 The proportion of results above the MDL for both sampling methods used to obtain breath samples from two ICUs based on all 24 compounds under study (N = 40, 2 replicates at each site). |
The dynamic range of the system appeared to cover several orders of magnitude, appropriate for the wide concentration distributions frequently observed in breath research. The observed larger absolute sensitivities for method B (the semi-invasive sampling method) are likely explained by the proximity to the sampling position closer to the source of origin of the VOCs. Also the position of the HME filter (between the source of VOCs and the sampling tube) may contribute to a lower absolute sensitivity for method A. Several less volatile, more polar species were recovered in significantly lower proportions using the non-invasive method compared to the semi-invasive method, likely reflecting interaction between breath VOCs and surfaces downstream of the ventilator circuit. However, this enhancement in recovery for method B did not necessarily translate to better MDLs, largely due to interference from contaminants within the sample matrix.
The two ICUs use different closed suction catheters (employed in method B), giving rise to varying levels of analytical contamination. These contaminants affected MDLs, either directly as exogenous sources of VAP-relevant species, or indirectly by affecting the chromatography of target compounds. Exploratory experiments whereby suction catheters were purged for over an hour showed a reduction in contamination (data not shown), however a significant proportion still remained. The catheters used in mechanically ventilated patients must be flexible in order to follow the anatomical routes of the upper respiratory system, and clinical ICU practice is restricted to CE marked equipment. Therefore, replacing the catheters with bespoke tubing made from lower emitting but more rigid materials (e.g., PTFE) is not an option. One of the most abundant VOCs emitted in catheter headspace was cyclohexanone, an organic solvent involved in the production process of certain medical devices containing polyvinyl chloride (e.g. endotracheal tubes, intravenous fluid bags, extracorporeal circulation tubing),39 and has been specifically related to patients receiving mechanical ventilation.40 Other high concentration species included several methylcyclosiloxanes, likely emanating from silicon elastomers built in to the catheters. Whilst compounds such as 2-ethyl-1-hexanol and isophorone may be considered exogenous species (the former is used to produce plasticizers whilst the latter is a polymer precursor), the extent to which they are produced in the body is unknown. To our knowledge, only one other study has described validation experiments in which air was sampled from a ventilator circuit in a laboratory.24 Low concentrations of 15 VOCs were found in medical air including cyclohexanone. The researchers did not consider this contamination to have substantially influenced the results of their clinical study investigating breath sampling in mechanically ventilated patients.
In many cases the LoDs in Table 4 were larger for method B than those associated with method A. This is due to the amount of contaminating signal on the chromatograms associated with the use of the suction catheters in method B. Relatively high LoDs for semi-volatile compounds in method A are likely due to interaction with surface materials along the longer sample flow path. Several of the compounds listed in Table 4 have shown to be of diagnostic interest: acetone, isoprene, octane, 3-carene, nonanal, dodecane and tetradecane have been linked to VA-LRTI,3,11–13,41–43 and isoprene has also been linked to muscle activity44 and may thus be altered in mechanically ventilated ICU patients. ARDS has been associated with higher concentrations of octane in exhaled breath (associated with approximately equal LoDs for both methods).45 For many of these compounds, method B showed lower LoDs (Table 4). However, the efficacy of both sampling methods is highly compound-dependent and thus will depend on the chemical species under study.
The non-invasive nature of method A is a significant advantage, allowing it to be implemented in large scale clinical studies (and potentially in clinical practice) more easily. However it should be noted that the creation of a single ‘ideal’ breath sampling method was not a goal of the present study, as different methodologies may require different sampling approaches.46 In our study, we were unable to quantify the potential differences between the ventilators, and as a result of different equipment being used by the different ICUs, the centres were associated with different MDLs. This is an important aspect to consider when conducting multi-centre studies and highlights the importance of a standardized approach to breath sampling both in the critical care unit and across the wider research community.
Footnotes |
† Iain White was involved in the execution of the experiments and in writing the initial manuscript, but he is unaware of the final manuscript as we are unable to contact him. Therefore he does not take responsibility for its contents. |
‡ Both authors contributed equally. |
This journal is © The Royal Society of Chemistry 2021 |