Electrokinetic technologies such as AC electro-osmosis (EO) and dielectrophoresis (DEP) have been used for effective manipulation of bacteria to enhance the sensitivity of an assay, and many previously reported electrokinetics-enhanced biosensors are based on stagnant fluids. An effective region for positive DEP for particle capture is usually too close to the electrode for the flowing particles to move toward the detection zone of a biosensor against the flow direction; this poses a technical challenge for electrokinetics-assisted biosensors implemented within pressure-driven flows, especially if the particles flow with high speed and if the detection zone is small. Here, we present a microfluidic single-walled carbon nanotube (SWCNT)-based field-effect transistor immunosensor with electrohydrodynamic (EHD) focusing and DEP concentration for continuous and label-free detection of flowing Staphylococcus aureus in a 0.01× phosphate buffered saline (PBS) solution. The EHD focusing involved AC EO and negative DEP to align the flowing particles along lines close to the bottom surface of a microfluidic channel for facilitating particle capture downstream at the detection zone. For feasibility, 380 nm-diameter fluorescent beads suspended in 0.001× PBS were tested, and 14.6 times more beads were observed to be concentrated in the detection area with EHD focusing. Moreover, label-free, continuous, and selective measurement of S. aureus in 0.01× PBS was demonstrated, showing good linearity between the relative changes in electrical conductance of the SWCNTs and logarithmic S. aureus concentrations, a capture/detection time of 35 min, and a limit of detection of 150 CFU mL−1, as well as high specificity through electrical manipulation and biological interaction.
Many previously reported electrokinetics-enhanced biosensors are based on stagnant liquids,2 where analytes can be manipulated more easily than under flowing liquids. If a biosensor is implemented within a pressure-driven flow rather than a stagnant liquid for higher throughput, automation, multiplexed detection, and/or continuous monitoring,11 stronger electrical forces need to be exerted on the analytes, such as pDEP, to overcome the hydrodynamic forces and to bind the particles to a detection zone, which usually constitutes a very small area for field-effect transistor (FET)-based sensors.12 This can be a challenge because an effective region for the pDEP is usually too close to the electrode of the channel surface to drag all the particles flowing from the top surface to the bottom toward the detection zone of a biosensor against the flow direction, especially when the particles flow with high speed and the detection zone is small. Hence, in most studies on electrokinetics-enhanced biosensors for flowing targets, capture efficiencies were either low or not measured, labeling was used, the limits of detection were relatively high, the electrical impedance technique was used for detection by using relatively large electrode areas, and/or sensor selectivity against non-targets was not tested (Table 1), although electrokinetic particle manipulation has been extensively explored for the last few decades.
Ref. | Target | Labeling | Buffer | Electrokinetics (stage 1) | Electrokinetics (stage 2) | Detection | Electrodes | Microchannel | Flow rate (averaged flow velocity) | Capture/detection time | Sensitivity | Selectivity test | Limit of detection | Measurement range | Remark |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
22 | S. cerevisiae | 0.1 M mannitol (2 μS cm−1) | AC pDEP capturing (5 Vpp & 100 kHz) | AC electropermeabilization (20 Vpp) | Electrical impedance | Planar IDEs | 500 μL min−1 (μm s−1) | 15 min (capture & detection) | 100 CFU mL−1 | 100–105 CFU mL−1 | |||||
38 | B. subtilis | DI water (1–2 μS cm−1) | DC iDEP capturing (1 kV) | Electrical impedance | Insulators & sensing electrodes | 1000 μm/90 μm | 40 μL min−1 (7407 μm s−1) | 2 min (capture & detection) | ∼2 [degree per (spores per mL)] | 103 spores per mL | 103–104 spores per mL | ||||
39 | E. coli K-12 | 0.1 M mannitol (1 μS cm−1) | AC nDEP pushing (20 Vpp & 1 kHz) | AC pDEP capturing (5 Vpp & 100 kHz) | Electrical impedance | Top-bottom electrodes | (270![]() |
15 min (capture & detection) | |||||||
40 | E. coli O157:H7 | DI water | AC pDEP focusing (5 Vpp & 5.6 MHz) | AC pDEP capturing (5 Vpp & 100 kHz) | Electrical impedance (Ab) | Planar IDEs | 300 μm/25 μm | 180 min (capture & detection) | 40 [k-ohm per (log CFU mL−1)] | E. coli O104:H4 | 300 CFU mL−1 | 300–3 × 105 CFU mL−1 | Blocking agent for non-specific binding was not used | ||
26 | S. choleraesuis/N. lactamica | Nanoprobe | 15 mM PBS-Tween20 (150 μS cm−1) | AC nDEP capturing (20 Vpp & 660 kHz) | SERS | Planar quadrupole electrodes | 800 μm/800 μm | 15 μL min−1 (391 μm s−1) | 10 min (capture & detection) | 70 CFU mL−1 | |||||
41 | S. epidermidis | 0.001× PBS | Resonant AC pDEP + ET collection (14 Vpp & 63 MHz) | Electrical impedance | IDEs | 1000 μm/300 μm | 5 μL min−1 (278 μm s−1) | 20 min (capture & detection) | 105 CFU mL−1 | ||||||
1 | E. coli | Drinking water (86 μS cm−1) | AC pDEP focusing (4 Vpp & 100 kHz) | AC pDEP collection (2.5 Vpp & 10 kHz) | Electrical impedance | Planar coplanar electrodes | 10![]() |
25 μL min−1 (1389 μm s−1) | 1 min/0.5 min | 0.01 [%/(CFU mL−1)] | 300 CFU mL−1 | 300–1500 CFU mL−1 | |||
42 | Prostate specific antigen | Ab-conjugated GNP | 0.1× serum in PBS (13![]() |
AC iDEP capturing (∼200 Vpp cm−1) | Voltammetry | Inlet–outlet reservoirs with nanoslit (30 nm) | Local nanochannel | ∼15 [(μA cm−2)/log (g mL−1)] | 1 pg mL−1 | 5–1000 pg mL−1 | |||||
43 | Avidin (inlet 1) | Biotin (inlet 2) | Diluted PBS (250 μS cm−1) | AC IET change (10 Vpp & 1–20 MHz) | Liquid interface monitoring | Inlet–outlet reservoirs | 100 μm/65 μm | 5 μL min−1 (12![]() |
626 fM | 50 fM–2.5 μM | |||||
44 | Dengue virus | Ab-conjugated silica beads | Diluted DMEM (1000 μS cm−1) | AC nDEP guiding (15 Vpp & 1 MHz) | AC nDEP capturing (15 Vpp; 1 MHz) | Fluorescence intensity | Top-down (ceiling-bottom) electrodes | 1000 μm/10 μm | 0.3 μL min−1 (500 μm s−1) | 5 min (capture & detection) | 104 PFU mL−1 | 104–106 PFU mL−1 | |||
8 | E. coli | DI water | AC EO + DEP collection (10 Vpp & 120 kHz) | Resonant frequency shift | Planar spiral IDEs | 400 μL min−1 (4444 μm s−1) | 10 min (capture & detection) | 326 [fg Hz−1] | 105 cells per mL | 105 cells per mL | Capture efficiency: 0.125% | ||||
7 | E. coli | Secondary Ab | DI water (1 μS cm−1) | AC pDEP collection (5 Vpp & 500 Hz) | SPR | Planar IDEs | 3000 μm/50 μm | 10 μL min−1 (1111 μm s−1) | 20 min/10 min | 0.7 [nm per (log CFU mL−1)] | S. epidermidis | 300 CFU mL−1 | 300–106 CFU mL−1 | ||
Present study | S. aureus | 0.01× PBS (200 μS cm−1) | AC EO + nDEP focusing (1.5 Vpp & 9 kHz) | AC pDEP collection (10 Vpp & 8 MHz) | FET conductance (SWCNT-Ab) | Coplanar electrodes | 200 μm/49 μm | 0.2 μL min−1 (340 μm s−1) | 10 min/25 min | 2.95 [%/(log CFU mL−1)] | B. subtilis, E. coli C3000, S. epidermidis | 150 CFU mL−1 | 200–106 CFU mL−1 | Capture efficiency: 23% |
One approach to address this challenge is to align all the particles along two-dimensional surfaces or one-dimensional lines during flow, which is commonly called “particle focusing.” Several particle focusing techniques have been developed in the literature. Inertial focusing is a passive technique that is generated by particle inertia under a flow field;13,14 however, complicated device structures are needed, and small particles are unlikely to be manipulated by this technique.15 Acoustic focusing involves surface acoustic waves, showing fast fluidic actuation and contact-free particle manipulation.16,17 DEP focusing incorporates an array of planar or vertical electrodes for applying the DEP forces; however, particle losses are unavoidable because many particles attach to the focusing electrodes in the case of the pDEP,1,18 and the DEP force is inherently insufficient in regions farther from the electrodes.19 These three focusing techniques are also strongly dependent on particle sizes; hence, small particles in the range of submicron sizes are difficult to focus.14,17 More importantly, none of these focusing technologies can align all the particles close to the detection area without losses to the channel surfaces, which is critical for development of a biosensor involving pDEP to capture flowing particles.
Herein, we present a microfluidic FET immunosensor equipped with electrohydrodynamic (EHD) focusing and DEP concentration for continuous and label-free electrical detection of Staphylococcus aureus in a 0.01× phosphate buffered saline (PBS) solution. Many flow-type DEP-mediated biosensors have involved impedance measurements with electron transfer through the media20,21 or pearl chain conductance of bacteria between the electrodes22 (Table 1). It has been suggested that FET-based sensors adopting silicon nanowires and CNTs are integrated with a microfluidic platform in order to enhance sensitivity and to achieve automated and multiplexed analytical systems;11 however, no studies in which particle focusing was implemented with FET-based biosensors under a pressure-driven flow have been reported.
EHD focusing uses a combination of AC EO and negative DEP (nDEP) rather than pDEP to align flowing particles along one-dimensional lines close to the bottom surface, and Ren et al.15 demonstrated EHD focusing for the first time using two electrical-bias-free electrodes at a fixed AC frequency (200 Hz) for 4 μm-diameter silica beads. The operation conditions in their study were very limited; both AC EO and nDEP were required to occur at the same frequency. In the present study, superimposed signals having two AC frequency components, lower (AC EO with weak pDEP) and higher (dominant nDEP), were implemented without electrical-bias-free electrodes to align incoming 380 nm-diameter polystyrene beads and label-free Staphylococcus aureus along one-dimensional lines close to the bottom surface to facilitate pDEP capture downstream. Polystyrene beads were used to demonstrate the enhanced electrokinetic characteristics of the present microfluidic device, as they are widely used in quantitative evaluation of DEP systems with their regular spherical shapes and homogeneities compared to actual biological particles.23
The bacteria aligned using EHD focusing were further concentrated downstream using pDEP forces and quantified with a carbon-nanotube-based FET immunosensor (Fig. 1a). This FET sensor consists of two embedded concentration electrodes, two detection (source and drain) electrodes connected to an array of antibody-functionalized and aligned single-walled carbon nanotubes (SWCNTs), and a dielectric layer (Si3N4) to separate these two pairs of electrodes vertically.6 The relative electrical conductance change (RCC) between the detection electrodes owing to the bacteria concentrated on the SWCNTs was continuously measured. A 0.01× PBS solution was used as the medium, with electrical properties similar to drinking or tap water. The selectivity of the sensor platform was tested against Bacillus subtilis, Escherichia coli C3000, and Staphylococcus epidermidis along with its sensitivity. The selectivity was also investigated by DEP-capturing S. aureus (target) and E. coli C3000 (non-target) on the S. aureus antibody-coated SWCNTs under flows without EHD focusing and then by flushing, which has not been clearly reported until now.
For the particle capture experiments with beads and bacteria, 380 nm-diameter red (excitation/emission: 542/612 nm) fluorescent polystyrene beads (Fluoro-Max R400; Thermo Scientific, USA) were suspended in 0.001× PBS buffer to obtain a number density of 5 × 105 particles per mL. The S. aureus and E. coli C3000 in 0.01× PBS solution were also labeled separately with SYTO™ 9 (S34854; Thermo Scientific, USA) (excitation/emission: 483/503 nm) (2.5 μM) at concentrations of 107 CFU mL−1.24 The electrical conductivity was measured using a conductivity meter (handylab pH/LF 12; SI Analytics GmbH, Germany), and these values were 200 μS cm−1 and 21 μS cm−1 for 0.01× PBS and 0.001× PBS, respectively, at room temperature.
The SWCNT channels were prepared between the detection electrodes and functionalized as described in previous work.6 Briefly, SWCNT powder (98% semiconducting) was suspended in N,N-dimethylformamide (DMF; 3057–4405; Daejung Chemicals, Korea) through ultrasonication and centrifugation (concentration: 10 μg mL−1). Then, 5 μL of the prepared SWCNT suspension was applied between the two detection electrodes, and aligned SWCNT bridges were formed by applying DEP at 5 Vpp and 200 kHz for 30–60 s (Fig. 1b). Polyclonal anti-S. aureus (ab20920, Abcam, UK) in 1× PBS (concentration: 10 μg mL−1) was then immobilized on the SWCNTs by incubation at 37 °C for 2 h via EDC(1-ethyl-3-(3-dimethylaminopropyl)carbodiimide)-NHS (N-hydroxysuccinimide) chemistry. SuperBlock T20 blocking buffer (0.05% Tween-20; 37536, Thermo Scientific, USA) was incubated for 30 min at room temperature to prevent non-specific attachment to the SWCNTs25 (Fig. 1d). All the surface modifications were performed through a polydimethylsiloxane (PDMS) well around the SWCNTs, and the sensors were rinsed with 1× PBS and dried after each modification step. The current–voltage (I–V) curves were measured after each modification step (Fig. S2†). All the bio-functionalized sensors were stored at 4 °C before further use to maintain stability.26
For the bead capture experiments, a lab-made voltage adder was used to superimpose signals from two function generators (AFG3022C; Tektronix, USA).3 The superimposed signals were applied to the focusing electrodes to activate EO vortices (1.25 Vpp, 5.4 kHz) and nDEP pushing (10 Vpp, 3 MHz) to align the particles along the stagnant lines (Fig. 1a). Another function generator (WW5064; Tabor, Israel) was used for the pDEP capture (10 Vpp, 200 kHz) of beads on the SWCNT channels. Here, each function generator offered sinusoidal signals that were 180° out of phase via the two-channel sources. A cooled interline transfer camera (ORCA-R2; Hamamatsu, Japan) connected to an inverted fluorescence microscope (Eclipse Ti-U; Nikon, Japan) was used to obtain images at an exposure time of 100 ms. The number of beads captured in a specified area (20 × 50 μm2) between the two detection electrodes was counted with ImageJ by dividing the total area occupied by the particles by the area occupied by a single bead in the fluorescence images.3,28
For the bacterial quantification experiments, EHD focusing (1.5 Vpp, 9 kHz) and concentration (10 Vpp, 8 MHz) electrodes were biased for 10 min, and the RCC of the SWCNTs was measured throughout the experiments via the detection electrodes using a source meter (2635B; Keithley Instruments, USA) at a fixed source–drain voltage of 100 mV. The RCC was computed as (G–G0)/G0, where G is the electrical conductance measured in real time and G0 is the equilibrated electrical conductance measured under steady flow before the actuation of AC signals. The stabilized RCC (SRC) was also computed as (GS–G0)/G0, where GS is the stabilized electrical conductance measured after the 10 min-bacterial attachment.29
The optimization process for the voltages and frequencies of EHD focusing in these two experiments is available in the ESI† (Fig. S3). The frequencies for DEP concentration were determined according to the Clausius–Mossotti (CM) factors under a maximal voltage output of 10 Vpp (peak-to-peak). Each experiment in this study was performed at least thrice, and the mean values are shown with their standard deviations indicated as error bars.
![]() | ||
Fig. 3 Clausius–Mossotti factors of selected bacteria and a bead. They are functions of electrical permittivity and conductivity of media and particles, where the wall, membrane, and cytoplasm were considered in the calculation in the case of bacteria.33–36 The dielectric property of B. subtilis is unknown. Red and blue vertical dashed lines indicate the AC frequencies for the bead and S. aureus experiments, respectively. |
It was observed that EHD focusing of the 380 nm-diameter beads was not easy to accomplish by the AC EO signal alone. Both AC EO vortices and weak pDEP attachment onto the focusing electrodes occurred at a frequency of 5.4 kHz. Hence, another nDEP signal (3 MHz) was superimposed on the AC EO signals, and the superimposed signals having both low (AC EO) and high (nDEP) frequency components were applied to the focusing electrodes. Successful particle focusing using the superimposed signals was then demonstrated experimentally (Fig. 4a and b) and numerically (Fig. 4c and d). The fluorescence images were obtained over the detection electrodes, and computer simulations were obtained with Comsol Multiphysics software (please see the ESI†). The beads were delivered to the DEP-effective region close to the concentration electrodes with minimal losses. The number of beads observed in the detection area was 14.6 times greater than those without EHD focusing, and no beads were captured without DEP concentration (Fig. 4e and f).
Both the number of captured beads and the corresponding capture efficiencies were analyzed (Fig. 4g) with respect to the flow rate, and the efficiency is defined as the number of beads captured in the detection area divided by the number of incoming beads. The beads were mostly captured at a flow rate of 200 nL min−1 (at an incoming number rate of 100 particles per min) and a flow velocity of 340 μm s−1. Under these conditions, the capture efficiency was approximately 23%. This is in contrast with an efficiency of ∼0.1% for electrode-deposited cantilever sensors in a microfluidic device,8 and efficiencies for electrokinetic capture in a narrow detection area have been rarely reported.
Fig. 5a shows real-time measurements of S. aureus with different bacterial concentrations (from 102 to 106 CFU mL−1) at the inlet. Before biasing the focusing and concentration electrodes, i.e. less than 0 min in Fig. 5a, the conductance variation was very small with a standard deviation (or noise) of 0.27% (Fig. S4†). The RCC owing to the captured bacteria was measured after turning off both the biases at 10 min for all bacterial concentrations, and the RCC was stabilized around 35 min. The SRC increased with increasing bacterial concentrations (Fig. 5b). That is, bacterial capture induced increases in the electrical conductance through the SWCNTs. In fact, as the surface of bacteria is negatively charged in the 0.01× PBS buffer, the binding of the negatively charged bacteria onto the bio-functionalized SWCNTs induces holes in the p-type semiconductor, resulting in the increase of electrical currents at the fixed source–drain voltages.31,32 The conductance signals measured during the first 10 min bias from 0 min, which was for focusing and bacterial capture, were caused by interference with the electric field of the concentration electrodes.
Fig. 5b shows the calibration plot of the microfluidic immunosensor with respect to the S. aureus concentration. The linear relationship (R2 = 95.6%) between the SRCs and the logarithmic S. aureus concentrations was maintained in a range of 200 to 106 CFU mL−1. This linearity did not continue when the bacterial concentration was around 107 CFU mL−1, possibly because of the attachment of too many bacteria beyond the capability of the bio-receptors on the SWCNTs. Interestingly, this characteristic was also reported in DEP-enhanced surface plasmon resonance sensors.7 The limit of detection (LOD) of the present immunosensor was calculated to be 150 CFU mL−1 based on the signal-to-noise (S/N) ratio of three (3-sigma). The SRCs were smaller than the 3-sigma when both the focusing and concentration were not turned on. Further, the measured relative standard deviation for 104 CFU mL−1 was 10.0% (Table S1†). The measurement of bacteria with a wide dynamic range (200 to 106 CFU mL−1) and improved LOD was rendered possible owing to the serial enrichment operation of EHD focusing and DEP concentration, and sensitive detection through the functionalized SWCNTs.
Fig. 5c shows continuous measurements of S. aureus at different bacterial concentrations (0, 200, and 1000 CFU mL−1) at the inlet during three repeated cycles; each cycle consists of turning on the EHD focusing and DEP concentration for 10 min and then turning off while measuring the RCC throughout the cycle. The measured values of SRC were −0.32%, 0.21%, and 0.07% at 0 CFU mL−1, 0.78%, 1.84%, and 2.82% at 200 CFU mL−1, and 3.05%, 4.05%, and 4.68% at 1000 CFU mL−1. The SRCs increased with the number of cycles at 200 and 1000 CFU mL−1. That is, S. aureus in the solution was continuously captured with the number of cycles until the bio-functionalized SWCNTs were spatially filled with the bacteria.
Fig. 5d shows the selectivity test against a mixture of S. aureus and S. epidermidis, E. coli C3000, and B. subtilis. A bacterial mixture (2 × 103 CFU mL−1, 1:
1 ratio) of S. aureus and S. epidermidis showed SRCs similar to that of pure S. aureus (103 CFU mL−1). Each of the B. subtilis (106 CFU mL−1) and E. coli C3000 (106 CFU mL−1) showed SRCs smaller than the 3-sigma. Most incoming E. coli and S. epidermidis were possibly attached to the focusing electrodes by pDEP with 9 kHz signals, whereas much less losses of S. aureus occurred by nDEP during focusing (Fig. 3).33–35
Both S. aureus and E. coli C3000 were captured on the SWCNTs coated with S. aureus antibodies when the concentration electrodes (10 Vpp, 8 MHz) alone were activated (Fig. 6), where an upright fluorescence microscope (Eclipse 80i; Nikon, Japan) with a CCD camera (CoolSNAP™ DYNO; Photometrics®, USA) was used to obtain images and videos with an exposure time of 200 ms. When the electrodes were deactivated, most of the captured S. aureus was still bound on the SWCNTs, whereas all the captured E. coli C3000 flowed away without non-specific binding (see ESI† Video). The pDEP can usually increase non-specific binding of particles on the surface, but this was prevented owing to the blocking buffer treatment (SuperBlock T20) and continuous channel flow. A few studies on avoiding electrokinetics-related non-specific binding have been reported; harsh electrothermal fluid motion was intentionally induced to remove non-specific binding in an assay,37 or blocking agent treatment and secondary antibody amplification were applied after the DEP capture of different bacteria.7 This FET immunosensor thus enabled an increase in sensitivity due to EHD focusing and pDEP concentration while maintaining a high level of selectivity against various non-target microorganisms, and high selectivity and low cross reactivity are highly required to address fundamental challenges in FET-based biosensors.11
This electrokinetics-enhanced microfluidic sensor can be used for detection of several other target analytes although there is a limitation on media conductivity and their DEP behavior needs to be known. Once it is revealed, the main parameters, AC frequency and media conductivity, can be adjusted for a particular target since both EO and nDEP should be generated for EHD focusing, and pDEP should be generated for concentration. For example, EO and nDEP occurred simultaneously at 9 kHz on the focusing electrodes for S. aureus detection in the present study; however, it may be difficult to avoid attachment to the electrodes during focusing in the case of S. epidermidis because pDEP can occur for this bacterium. In this case, an AC signal above 40 MHz can be superimposed with the 9 kHz (EO + pDEP) signal to generate stronger nDEP, resulting in EO and nDEP, which was also observed when collecting 380 nm-diameter beads in this study. Moreover, the temperature of a medium can increase significantly because of Joule heating in the case of high conductivity media. For example, the maximal temperature rise of the medium owing to Joule heating increased from 0.07 °C for 0.01× PBS (the present study) to 6.24 °C for 1× PBS (Fig. S6†).