Yongxiao
Zhou
a,
Erik M.
Werner
a,
Eugene
Lee
b,
Michael
Chu
a,
Thao
Nguyen
c,
Kevin D.
Costa
bd,
Elliot E.
Hui
a and
Michelle
Khine
*ab
aDepartment of Biomedical Engineering, University of California, Irvine, CA 92697, USA. E-mail: mkhine@uci.edu
bNovoheart, Vancouver, Canada
cDepartment of Chemical and Biomolecular Engineering, University of California, Irvine, CA, USA
dCardiovascular Research Center, Icahn School of Medicine at Mount Sinai, New York, NY, USA
First published on 10th December 2020
Microfluidic devices are traditionally monitored by bulky and expensive off-chip sensors. We have developed a soft piezoresistive sensor capable of measuring micron-level strains that can be easily integrated into devices via soft lithography. We apply this sensor to achieve fast and localized monitoring of pressure, flow, and valve actuation.
Despite tremendous advances of micro total analysis systems in recent years, widely accessible on-chip monitoring and closed loop control of fundamental parameters are still lacking. The dearth of on-chip monitoring solutions creates inherent limitations in the responsiveness and accuracy of the measurements that can be obtained. Off-chip hydraulic and pneumatic sensors are limited by the dead volume of the interface tubing connecting the sensors to the chip. This dead volume is typically large compared to the volume of the microfluidic device itself and can be the dominant factor in determining the response time and accuracy of a measurement.
Currently available options for local measurement of these parameters are difficult to integrate into microfluidic systems. While optical sensors can produce accurate, reliable, and robust flow and pressure measurements, they still require coupling to expensive and complicated imaging systems.10–13 Micro electromechanical systems (MEMS)-based sensors offer on-chip integration with high resolution but typically involve complex fabrication and contact-based measurements. For example, in-channel sensors that extend into the fluid channel affect the local flow profile and can suffer from confounding factors including fouling; increased drag force from fouling can cause inaccurate results.14 Also, unlike microfluidic devices, commercially available MEMS sensors are not intended for single-use applications; hence this mismatch in cost and complexity has prevented more pervasive integration.15
Soft, stretchable sensors have attracted research interest due to their ability to conform to different surfaces and their large dynamic range under deformation. These sensors convert mechanical displacement into electrical signals such as resistance or capacitance change. Liquid metal-based pressure sensors with a polydimethylsiloxane (PDMS) substrate can be easily integrated into microfluidic devices. However, channels that contain liquid metal require extra precautions during fabrication or are more prone to mechanical failure. Alternatively, thin metal film-based sensors are easier and safer to fabricate and handle while offering attractive performance and robustness characteristics.16–18 Due to their physical properties, these metal thin film based sensors are able to sense mechanical deformations in various planes; the resulting electrical signals can be correlated and calibrated to physical parameters-of-interest. However, these soft strain gauges, have been typically limited to macroscale applications.19–21 There are few reports of soft sensors capable of monitoring micro-scale strains. Even recent papers focused on micron-scale sensors still report monitoring deformations on the millimeter scale.22 The ability to monitor deformations from extremely small forces requires unique strategies. For instance, wearable sensors may not respond as linearly in this micro regime as in macro-level, and gauge factor has been reported to be different between low strain range and high strain range.23,24 Secondly, in micro-applications, the system may not be able to actuate the strain sensor due to limited force output (e.g. the small force generated from a monolayer of cardiomyocytes, or small pressure changes in a microfluidic channel). The stress generated by an isolated muscle strip ranges from 8 to 20.7 kPa,25 which is not strong enough to drive conventional rigid force gauges.
To date, there are a limited number of papers that have demonstrated the effective application of flexible sensors in micro-device monitoring.24,26 Parker and colleagues developed a high-sensitivity piezoresistive sensor using multi-material 3D printing to monitor stress induced by cardiac tissues, with a reported minimum tested strain of 0.0125%.26 Flexible sensors such as this have the potential to replace traditional optical methods to monitor tissue contractility.27 Wen and colleagues developed a silver powder doped-PDMS based piezoresistive pressure sensor that can be bonded to a microfluidic device.24 When the pressure in the channel increases, the flexible sensor is stretched.
In situations of pressure driven flow, the pressure is directly proportional to flow rate. Thus, the flow rate can be calculated from the pressure measured by a sensor in the fluid channel. While most reported noncontact flow meters have a resolution of tens to hundreds of μl min−1,28–30 some research groups have demonstrated nanoliter resolution temperature flow sensors31 and 0.5 μl min−1 resolution microwave flow sensors.32 However, temperature flow sensors31 could be disturbed by non-flow effects, such as environmental heat flux flowing into sensors during experiments. Unlike other parameters, pressure is still a flow indicator that is independent from surrounding noise such as electromagnetic waves and heat flux. In the flow sensor by Sanati-Nezhad and colleagues, pressure in a microfluidic channel deforms a membrane to modulate the permittivity of a microwave resonator, thus producing a flow measurement.32
From the current literature on available sensors for micron scale in situ monitoring, there remains the need to develop a universal sensor compatible with soft lithography that can be scaled, arrayed, and used to measure a range of critical microfluidic parameters.
In this paper, we present a completely encapsulated wrinkled metal thin film based flexible piezoresistive sensor with tunable elastic modulus that can measure micron-scale strain, microfluidic device pressure, and valve state. This soft strain sensor has a dynamic range of 50% and can detect linear displacements as small as 5 μm (0.025% strain). The displacement of the sensor can be used to calculate the force applied to the sensor. Due to its high strain sensitivity to linear stretching and ultra-soft substrate, small pressures applied on the surface will deform the sensor, causing it to expand orthogonally to serve as a highly sensitive pressure sensor for microfluidic applications. The pressure measured from the microfluidic device can be correlated to flow rate in the channel as well. Finally, the sensor can be integrated into a pneumatic valve to monitor valve actuation. To the best of our knowledge, there is no such sensor that can electrically monitor valve state in microfluidic devices.
The composition of the functional metal thin film was tuned to achieve a balance of brittleness and stability in the sensor to achieve a stretch resolution of 5 microns. The metal thin film is a bilayer of platinum and gold. Material brittleness will affect the number and size of cracks that form along with the energy required to form cracks. Platinum is a more brittle material while gold has good ductility.33 A thicker platinum layer results in more and larger cracks but leads to unstable resistance. As a more ductile material, a gold layer will lead to fewer cracks, but the change in resistance is significantly smaller. A balance can be achieved by controlling the thickness of platinum and gold, respectively. After testing various combinations, we chose a 40 nm platinum along with a 5 nm gold layer because it provided the highest signal detection while still maintaining stability. The sensor's substrate is 70 μm thick PDMS, with an encapsulation layer of 30 μm PDMS, with the wrinkled metal layer sandwiched in between the PDMS layers. A detailed cross-sectional dimension of the sensor is shown in ESI† Fig. S1.
To calculate the conversion between mechanical displacement and corresponding force, certain approximations and assumptions were made. As the sensor is stretched at the micron scale with negligible deformation, the deformation of the sensor is assumed to be a uniform beam that is undergoing uniaxial stress and has elastic like behavior. From eqn (1):
σ = E·ε | (1) |
We can further expand on this in eqn (2):
(2) |
Fig. 1b shows a representative sensor's behavior under different stretching frequencies. It demonstrates the sensor's repeatability. The sensor's pad area is fixed on a linear actuator (Zaber Technologies Inc) with the tip area clamped on a moving stage. The moving stage was cycled by 5 μm at 0.5, 1, and 2 Hz, respectively, while the sensor tracked the changes accordingly. A baseline shift of the stage movement was also captured at ∼265 s for 2 Hz. While it is not obvious in the figure, there is signal delay between the position and resistance. For 0.5, 1, and 2 Hz the response times are 169, 80, and 27 ms respectively.
To further understand the signal latency, one more experiment was performed. As shown in Fig. 1d, a typical sensor was stretched by 200 μm at a speed of 200 μm s−1, held for 10 seconds, and released back by 200 μm at a speed of 200 μm s−1. The position of the linear actuator and resistance of the sensor were both recorded. 34 tests (N = 10 sensors) were performed. On average, the actuator began to move at 5.03 ± 0.01 s while the sensor began to detect a resistance change at 5.08 ± 0.08 s. The stop time was defined as the time at which the sensor or actuator reach 90% of value of the maximum relative change. The actuator stopped at 5.90 ± 0.02 s and the sensor stopped at 5.95 ± 0.1 s. The data indicate that the sensors have an average response time of 50 ms. Computer processing and device communication time, however, also contribute to this response time.
To observe signal hysteresis, the sensor was cycled to 150 μm and stretched at 20 μm s−1 speed 20 times. The hysteresis for a representative sensor is shown in ESI† Fig. S2. From this figure, although reproducible, the sensor's resistance followed different trajectories when stretched and released at large deformations. With the loading and unloading behaviour displaying different sensitivities, it is important to know which trajectory the sensor is on when tested. Fig. 1c shows a representative sensitivity curve in terms of the change in resistance versus change in length. As the sensors were initially stretched, wrinkles in the metallic thin film unfolded, resulting in minimal changes in resistance. As strain increases and cracks form and propagate, the resistance increased nonlinearly.
When fluid is pushed through the microfluidic device, the pressure within the channel deformed the membrane of the piezoresistive sensor (ESI† Fig. S3) and changed the electrical resistance of the functional metal film. As shown in Fig. 2a, the channel (clear, blue) overlaps with the sensing area (black).
When pressure is applied normal to the sensor surface, the sensor substrate will expand in the transverse plane. Lateral expansion of the sensor elongates the metal film causing cracks to appear; when pressure is reduced from the surface, the substrate returns to its original shape, and the fractured metal will come back into contact with each other. Due to the design difference between the trace and pad area, the pad area had a larger metal area. However, from the simulation results in Fig. S4,† the pressure within the region that overlaps the sensor pad area was several-fold smaller than that of the region overlapping the trace area.
We assessed several aspects of the sensor performance, including working range, resolution, accuracy, and repeatability. For our microfluidic device, with a working flow rate range of 6 μl min−1 to 200 μl min−1, the measured pressure from the inline pressure sensor of the inlet fluid varied between 1 kPa to 74 kPa. Fig. S4† depicts a simulation of pressure within the channel under a 10 μl min−1 flow rate. The pressure map shows the gauge pressure which can be related to deformation on the channel wall and sensor. Gauge pressure inside the channel drops along the pathway and reaches 0 at the open-air outlet indicated in Fig. 2b. With a channel width of 250 μm and trace width of the sensor 300 μm, the overlap area is small in comparison to the entire sensor. The deformation of a single overlap area may be too small for the signal change of the sensor to be detected. Thus, multiple sensor-channel crosses are used to increase the overlap area to boost the signal. However, from the simulation (Fig. S4†), the pressure drops along with channel length, and the deformation of the cross area will become smaller with less pressure. As a result, more overlap will increase the total signal sensitivity, but with diminishing returns. With the variable pressure along the channel and the sensor having multiple crosses within channel to increase signal change, it is difficult to detect localized pressure. In this configuration, the sensor detects overall deformation caused by the pressure.
To confirm our results, a pressure sensitivity test was performed on the sensor. A 3 mm by 15 mm acrylic flat was placed over the sensor trace area. A force gauge (Mark 10 M5-025) was fixed on a test stand (Mark 10 ESM 303) and placed into contact until pressure was applied to the acrylic piece. As the pressure increased, the sensor's resistance increased as well (Fig. 2c). The blue line is the average resistance across 5 runs. Red markers indicate the standard error at every 2 kPa increment. The yellow line shows a linear fit to the 0–12 kPa pressure range, yielding an R2 value of 0.941. From the graph, although the resistance value varies across sensors, they all follow the same trend and are relatively linear, especially at low pressures.
Fig. 2d represents a variable flow rate test showing pressure and sensor data versus time. Flow rate increases from 0 to 50 μl min−1 in 10 μl min−1 increments, and the entire test is repeated three times consecutively.
The working range for our device is 6 μl min−1 to 200 μl min−1. The criterium for minimum resolution is that the signal change between two different flow rates is at least 3-fold larger than root mean squared noise. In a flow rate test ranging from 0 to 30 μl min−1 with 2 μl min−1 increments, data show that the minimum detectable flow rate is 6 μl min−1, and resolution is 2 μl min−1 (Fig. S5†). Although the flow rate working range for our device is tested to an upper limit of 200 μl min−1 (as shown in Fig. 2e), the device has been tested up to 300 μl min−1 without failure (data not shown).
Although sensor data shows good correlation to changes in pressure and flow rate, the baseline signal decays when strain is removed and the sensor returns to an unstretched state (as shown in Fig. 1c). The flow rate drops from 20 μl min−1 to 0 μl min−1 as shown in Fig. 2f. Signal decay is noticeable when the sensor reading drops even when the linear actuator or syringe pump is idle. The decaying tails observed in Fig. 2d and ESI† Fig. S2a illustrate the baseline decay with a time duration greater than ∼85 min. Similarly, signals at zero flow rate decrease in value as well (Fig. 2d, four separate zero flow rate points are ∼35 min, 55 min, 75 min, and 90 min). This decay complicates the data analysis and limits the duration that the resistance to pressure relationship is accurate but can be accounted for with subsequent data processing. Because baseline decay occurs in all sensor data, every test data has a different baseline value. In order to compare inter-trial data with different starting baselines, all data are subtracted by the beginning baseline resistance so that it starts at 0. Additionally, the decaying trend is compensated by data post-processing. As shown in Fig. 2f, the value of each valley was extracted, and set to 0 Ohm. Each valley point is used to form a linear interpolated line. The data points between the valleys were adjusted by subtracting the linearly interpolated lines, between the valley, from the signal so that the sensor signal at each zero flow rate is set to 0 Ohm.
The system elasticity is one minor issue that contributes to the signal decay; another possible contribution to the signal decay is the polymer relaxation. Relaxation is an intrinsic property of the polymer substrate. As the channel wall and sensor floor undergo mechanical hysteresis and relaxation, the formation and contact points of cracks in the embedded metal thin film are affected, resulting in an electrical hysteresis as well. Other groups have demonstrated that the hysteresis in piezo-resistive based elastomeric strain sensors can be potentially accounted for using machine learning.34
To ensure repeatability of the sensor, conditioning tests were performed on the chip device. The fluid was flowed through the pre-primed device at 20 μl min−1 for 2 minutes and then paused for 2 minutes; this cycle is repeated 10 times. The sensor resistance difference between 0 and 20 μl min−1 was compared for 10 cycles (Fig. 2f). Although some decay remained, the difference in resistance decrease was greatly reduced after 3 cycles.
We first investigated the ability of the integrated piezoresistive sensor to measure the state of a valve configured to switch fluid flow on or off (Fig. 3a). An external hot wire anemometer (Zephyr HAF, Honeywell) was configured to measure the flow rate of air through the valve as it was opened and closed. Upon valve actuation, the integrated sensor produced a sharp spike in signal followed by an increase in baseline resistance when opened and a decrease when closed. These results show the membrane stretching before the valve opens followed by the membrane remaining in a partially stretched state while the valve remains open (Fig. 3c). Upon closing, the sensor signal spikes again as the vacuum is released and the membrane contacts the valve seat sealing the valve closed. We observed that the spike that occurred during valve state changes was dependent on the orientation of the sensor and was most pronounced when the sensor was placed directly over the seat of the valve. Data from the external air flow sensor showed the valve completely opened and closed, and the sensor did not interfere with normal operation of the valve. These results indicate the piezoresistive sensor is suitable for monitoring the state change of the valve.
Normally closed elastomeric membrane valves can also be used to create digital logic gates that are well suited for building integrated microfluidic control circuitry.9 Therefore, we next investigated the ability of the integrated piezoresistive sensor to measure the state of a valve configured as a microfluidic inverter gate. This circuit adds a pull-up resistor before the vacuum connection to the valve and an output connection upstream of the resistor to produce a digital pressure output signal that is the inverse of the input signal. The sensor reported an increase in resistance of approximately 6 ohms when the valve was opened and returned to baseline when the valve was closed, providing a clear electronic signal that corresponded to changes in the pneumatic output of the inverter gate.
Finally, to create a simple integrated microfluidic control circuit, an oscillator pump9 was constructed consisting of three identical inverter gates connected in a ring and three liquid handling valves, each connected to the output of an inverter gate (Fig. 3g). When a constant vacuum pressure was applied to the oscillator, the pressure sequence generated by each inverter opened and closed the pump valves to create a peristaltic pumping action. The piezoresistive sensor was placed under the final valve in the pump while high speed video imaging was used to monitor the incident light reflected when the valve membrane was pulled open (Fig. 3g inset). The oscillation frequency measured by the sensor agreed well with the measurements acquired using high speed video imaging, and the sensor was able to accurately measure oscillation frequencies as high as 24.9 Hz, approximately the Nyquist frequency of our acquisition device (Fig. 3i and ESI† Fig. S8). Previous work has shown that the frequency of a ring oscillator can be tuned by changing the input pressure9 and the average flow rate of an oscillator pump is dependent on the frequency.35 Using this information, the oscillation frequency can be used to calculate the average flow rate from the pump.
When placed under the final valve in the peristaltic pump (Fig. 3h), the sensor readings also aligned well with the pulsatile flow rate measurements acquired from a hot wire anemometer (Zephyr HAF, Honeywell) connected to the output of the pump (Fig. 3j). A small backflow was detected when the final pump valve opened that was mostly negated by the closing of the middle valve in the pump. Finally, a strong forward pulse occurred when the final valve closed. Monitoring the state of the final valve in the pump, the sensor can be used to indicate the instantaneous flow rate produced by the pump. A detailed explanation of the working principle of the oscillator is provided in ESI† Fig. S7.
As an indirect flow meter, our sensor also can detect flow rate in situ as low as 6 μl min−1 with a resolution of 2 μl min−1 in our device. However, for our current set up (with only a single sensor), failure will occur if the device is clogged. This causes the pressure to build up and the sensor readings to increase, but nothing will be flowing. In future work, multiple sensors can be used such that separate measurements at each intersection of the channel can be acquired. This allows for more precise local pressure and flow monitoring. Moreover, it will allow for the detection of clogging in the channel. As the pressure increases prior to the clogged point and decreases after the clogged point, the sensors will show relatively high or low readings at different intersections.
For monitoring microfluidic valve state, our integrated sensor provides a more direct method to monitor valve actuation than existing optical monitoring methods. Currently the sensor can monitor the binary status of a single valve precisely. Due to the analog output of the sensor, it could potentially detect partially opened valve rather than binary open and closed status; however, this may require individual calibration of each sensor.
The sensor still has some drawbacks. As mentioned in the results section, hysteresis and decay of the signal affect repeatability of the sensor. With hysteresis present in our system, only the loading signal path is used for analysis. For our strain and liquid flow tests and experiments, we focused on the loading trajectory rather than unloading trajectory for consistency, particularly as the decay is less severe. For the valve experiment, we are qualitatively checking for the opening and closing of the valve along with other features of the actuation (such as spikes shown in Fig. 3). Although hysteresis still exists, it is not as critical here and can be compensated for with machine learning algorithms34 for this specific application.
Due to decay, the signal baseline varies during experiments so the starting resistance must be subtracted to zero the baseline for different experiments. Several attempts have been made to minimize hysteresis and decay. In one case, we observed that stiffer substrates demonstrate less decay; however, stiffer substrates require larger loads to deform which may decrease the detection resolution of the sensor. For some physiological applications, it may not be possible to apply larger forces. Thus, adjusting stiffness according to different applications is a potential solution to minimize hysteresis and decay. Additionally, use of other substrate materials with less intrinsic hysteresis than PDMS is possible, too.
We demonstrated the ability to integrate our soft and extremely sensitive strain sensor into microfluidic devices to provide contactless detection of pressure that can be correlated with flow rate. The sensor can also be embedded into PDMS based valves to detect the extent of valve opening in microfluidic devices. Moreover, being PDMS-based, the sensor can be easily trimmed and bonded to any other silicone-based devices via plasma treatment. The measurement results show good linear correlation between sensor reading and flow rate and pressure in the device. The sensor has a flow rate detection range from 6 microliters per minute (μL min−1) to 200 μL min−1 and a resolution of 2 μL min−1. The sensor can confirm partial or complete valve actuation under different pressures.
Because the sensor is made of PDMS, it is compatible with soft lithography and easily integrated into microfluidic chips. The stiffness of the substrate along with the sensitivity and dimensions of the sensor can be adapted to different applications. The soft and flexible substrate also makes it possible to integrate the sensor into biological applications and monitor micron-scale tissue movement. The sensors can also be readily arrayed; for example, it can be extended from one valve to multiple valves to measure several valves' status, important for large-scale microfluidic systems that require real-time feedback to control each valve.
The thin film based piezoresistive sensor consists of two layers of PDMS with customized stiffness (Young's modulus ∼250 KPa) and one layer of wrinkled bimetallic thin film (platinum and gold). The total thickness of the layer is ∼100 μm. The metal film is sandwiched and firmly bonded in between PDMS layers to stay insulated and prevent from wearing and scratching. The polymer layers and wrinkled metal film will deform under stretching or compression; due to the brittle wrinkled structure of the metal film, micro-cracks will form. As more and larger cracks form on the metal film, the electrical resistance increases.
The sensor is directly embedded at the bottom of the chip and serves as the base of the channel. The pressure required to drive fluid flow deforms the channel. As the upper and side walls of the channel are about 10-fold stiffer than the bottom sensor wall, most of the deformation will occur on the sensor surface. The electrical resistance of the sensor increases due to the deformation described above.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/d0lc01046d |
This journal is © The Royal Society of Chemistry 2021 |