Beatrice W. Soh*a,
Aniket Chitre
b,
Shu Zheng Tanc,
Yuhan Wanga,
Yinqi Yi
a,
Wendy Sohc,
Kedar Hippalgaonkar
ac and
D. Ian Wilson
b
aInstitute of Materials Research and Engineering, Agency for ScienceTechnology and Research (A∗STAR), Singapore 138634, Singapore. E-mail: beatrice_soh@imre.a-star.edu.sg
bDepartment of Chemical Engineering and Biotechnology, University of Cambridge, Philippa Fawcett Drive, Cambridge CB3 0AS, UK
cDepartment of Materials Science and Engineering, Nanyang Technological University, Singapore 117575, Singapore
First published on 5th February 2025
We present an improved high-throughput proxy viscometer based on the Opentrons (OT-2) automated liquid handler. The working principle of the viscometer lies in the differing rates at which air-displacement pipettes dispense liquids of different viscosities. The operating protocol involves measuring the amount of liquid dispensed over a set time for given dispense conditions. Data collected at different set dispense flow rates was used to train an ensemble machine learning regressor to predict Newtonian liquid viscosity in the range of 20–20000 cP, with ∼450 cP error (∼8% relative to sample mean). A phenomenological model predicting the observed trends is presented and used to extend the applicability of the proxy viscometer to simple non-Newtonian liquids. As proof-of-concept, we demonstrate the ability of the proxy viscometer to characterize the rheological behavior of two types of power-law fluids.
Viscosity is a key physical property of many products, hence the measurement of viscosity plays a critical role in a wide range of industries, from pharmaceuticals3,4 to cosmetics5,6 and food production.7 The traditional methods of measuring viscosity are typically time-consuming and labor-intensive. For example, a single run on a rotational rheometer requires careful sample loading and instrument calibration, followed by thorough cleaning after the measurement. High-throughput is becoming a priority for rapid screening purposes, particularly with many industries looking to re-formulate their portfolio of products to meet sustainability and regulatory pressures.8–11 Alternate methods for measuring viscosity have been developed, such as microfluidic-based rheometers.12–14 However, such techniques cannot attain the degree of throughput and automation required for screening large numbers of samples. Automated rheometers are available commercially, but these are expensive and difficult to integrate into larger workflows. There remains a need for an easily accessible solution to automated, high-throughput viscometry.
In this paper, we extend prior work15 on an automated and high-throughput method of measuring viscosity based on the Opentrons (OT-2) pipetting robot. The key extensions are an increase in applicable viscosity range for Newtonian liquids through implementation of real-time mass measurements and the integration of an analytical model to extend the proxy viscometer to non-Newtonian liquids. These improve the versatility of the platform, for example by enabling the screening of low-viscosity fluids, and represent a significant stride towards a more general-use proxy viscometer. The working principle of the proxy viscometer leverages the differing ability of air-displacement pipettes in dispensing liquids with varying viscosities. Under given dispense conditions, the actual amount of liquid dispensed depends on the liquid viscosity. We measure the amount of liquid dispensed over time for a range of dispense flow rates and liquid viscosities (for Newtonian fluids) and train a regression model to predict liquid viscosity within the range of 20 to 20000 cP. We develop an analytical model to describe the experimental results and extend the model to a simple non-Newtonian (power-law) liquid, demonstrating proof-of-concept cases on molasses and a surface cleaner. A single viscosity measurement on our proxy viscometer takes ∼1.5 minutes and requires minimal human intervention. Moreover, the Opentrons is cost-effective, readily integrated into workflows and uses open-source software, all of which have resulted in its widespread use in automated workflows.16–18
![]() | ||
Fig. 1 Experimental setup for the high-throughput viscometer: automated liquid handling robot Opentrons (OT-2) with a precision balance fitted beneath the dispense plate. |
Stage | Step | Description |
---|---|---|
Aspiration | Pick up pipette tip | Begin a run |
Aspirate fluid at 100 μL s−1 | Withdraw fluid from reservoir | |
Delay for 30 s | Allow for equilibration of air pressure | |
Touch pipette tip to four sides of the well, performed at three different heights | Remove droplets on various locations of pipette tip exterior | |
Move pipette tip to dispense plate | Aspiration stage complete | |
Dispense | Dispense fluid at set flow rate for 10 s | Weigh fluid dispensed from pipette tip |
Delay for 30 s | Allow remaining fluid to drip | |
Dispose pipette tip | Dispense stage complete |
For a given set dispense flow rate, the liquid is dispensed for 10 s and the amount of liquid collected on the plate is measured at intervals of 0.2 s. After 10 s, there is a delay of 30 s to allow any remaining droplets to fall onto the dispense plate. The pipette tip is then disposed of and a single measurement is concluded within 1.5 minutes. Five replicates were carried out for each fluid tested. The pipette tip was changed between each replicate. The key difference in measurement protocols between this work and previous work lies in the measurement of the mass of fluid. Previously, the mass of fluid dispensed was measured at a single time point, at the end of the dispensing step;15 in this work, the mass of fluid dispensed is measured in real time, providing higher-fidelity information about the dispensing process and thus enabling us to extend the viscosity range of the proxy viscometer.
Two non-Newtonian liquids tested for proof-of-concept purposes were molasses (Brer Rabbit) and Cif cream surface cleaner (Unilever). Cif is a non-colloidal suspension of non-spherical calcium carbonate particles (average size around 50 μm) in a viscous binder. The rheological behavior of each fluid was measured at 25 °C using steady state shear rate sweeps on a rotational rheometer with parallel plates. Three replicates were performed for each fluid and the average values are reported. The densities were measured as described for the Newtonian standards.
For each dispense profile, we fit the following sigmoid function to the data:
![]() | (1) |
![]() | ||
Fig. 3 Sigmoid fits to the dispense curves. (a) Five replicates of the measurement protocol for a Newtonian standard with viscosity of 6033 cP at a set dispense flow rate of 80 μL s−1. The black lines are fits of eqn (1) to the red experimental data points. The fit parameters to eqn (1) for all measured viscosities and set dispense flow rates are plotted: (b) Vplat, (c) m and (d) s. |
The fitted parameters Vplat, m and s for the Newtonian liquids are plotted as a function of viscosity in Fig. 3(b)–(d). We observe strong relationships between the fitted parameters and viscosity, especially for Vplat, suggesting that regression models can be trained to predict liquid viscosity. Generally, with an increase in viscosity, Vplat decreases and m increases. This corresponds to smaller final dispense volume for liquids with higher viscosities and the liquids being dispensed at a slower average rate.
Since the viscosity of a Newtonian liquid is independent of the applied shear rate, we expect the regression model trained on different set dispense flow rates to predict similar viscosities. As such, we developed an ensemble model by taking a weighted average of the viscosity predictions from the models at different flow rates, where the weighting was based on the train RMSE values, i.e., individual model accuracies. As seen from Fig. 4(f), the weighted ensemble model outperforms the model trained on the individual flow rate datasets, with train and test RMSE values of ∼450 cP (∼8% relative to sample mean). We note that the ensemble model presents only a marginal improvement in accuracy over the model trained on the individual flow rate datasets (for dispense flow rates larger than 10 μL s−1). Hence, the ensemble model can be employed if a user prioritizes accuracy over speed. Otherwise, the user can run the measurement protocol at only one set dispense flow rate (≥20 μL s−1) and use the model trained on a single flow rate to obtain a less accurate prediction. This might be desirable for the rapid screening of many samples.
To circumvent this, we develop a phenomenological model that can be applied to non-Newtonian liquids. Consider the pipette tip in Fig. 5: the pipette operates by firstly immersing the tip below the surface of the sample liquid. The plunger motion initially expels air through the tip and then draws liquid up as the plunger withdraws and creates suction in the tip cavity. The pipette tip is then moved out of the liquid. Fig. 5(a) shows the configuration: Vair is the volume of air in the cavity at internal pressure Pin. During the dispense stage, the plunger moves downward, its motion occupying volume at a rate Qp. The air in the cavity is compressed, increasing Pin and this causes the liquid to leave through the tip at volumetric flow rate QL:
QL is determined by the geometry of the pipette tip, the rheology of the liquid and the instantaneous pressure difference Pin − Patm, where Patm is the pressure of the surroundings.
Assuming that the tip walls are rigid and the liquid is both involatile and incompressible, a volume balance on region ADD′A′ yields
![]() | (2) |
![]() | (3) |
Contributions from gravity and capillary action are neglected in the flow model. The pressure difference associated with a drop at the pipette tip is of order 2Γ/RAA′, where Γ is the surface tension and RAA′ is the radius of the duct at the tip exit (of order 1 mm). For water at 20 °C, this gives a pressure difference of order 150 Pa for a hemispherical drop, which would halve if the liquid leaves as a cylindrical thread. These differences are small when compared to the pressure required to expel a viscous liquid (see below), but capillary action could impose an initial lag before the first drop is collected.
Inertia and transients in flow are neglected and QL is modelled assuming instantaneous steady state, with the pressure drop dominated by that required to pass through the converging conical duct ABB′A′. The Cogswell result21 for converging flow of a power-law fluid with shear stress σ related to shear rate by
σ=K![]() | (4) |
![]() | (5) |
![]() | (6) |
The first term on the RHS is the apparent shear rate, app, and the flow rates studied by Soh et al.,15 of 5–100 μL s−1, in the pipette tip in Fig. 5 correspond to shear rates of 9–175 s−1. Eqn (6) can be written as
Pin − Patm = BQLn | (7) |
yielding the governing equation
![]() | (8) |
![]() | (9) |
Rearranging gives
![]() | (10) |
Since p(t = 0) ∼ 0, this predicts that p (and Pin) will initially increase and approach a limit asymptotically, where plim = BQpn, i.e. QL = Qp: the air in the cavity undergoes compression until the pressure is sufficiently high to drive the liquid out at the rate set by the piston.
The instantaneous liquid flow rate is set by p and the volume dispensed after experiment duration tE is
![]() | (11) |
![]() | (12) |
For the pipette tip in Fig. 5, β = 15 × 109 Pa s m−3. Taking P0 = 101325 Pa, V0 = 2000 μL and a set dispense rate, Qp, of 80 μL s−1, tC is given by 2083/μ and c′ = 1 + 83.3/μ. The characteristic timescale in eqn (12) is then tC/c′2. For a 6 Pa s liquid, representative of those considered by Soh et al.,15 c′ = 14.9 and tC is 347 s, giving a characteristic timescale of 1.6 s, which is comparable with the length of the dispense times (5–10 s): the flow rate will change noticeably over the duration of a test with a viscous liquid due to air compression. For less viscous solutions, assuming P0 ∼ Patm, the characteristic time scale will be approximately βμV0/P0: for aqueous solutions, with μ ∼ 1 mPa s, this is of order milliseconds and the pipette will dispense accurately.
For the 6 Pa s liquid, the ratio of the pressure drop across the pipette tip to that across Section BC is >40, justifying neglecting the contribution from above BB’ for this and other liquids tested by Soh et al.15 Fig. 6 presents a series of model predictions for Newtonian liquids with viscosity 6.08 Pa s at a set dispense flow rate of 80 μL s−1. The numerical integration of eqn (10) matches eqn (12). The flow rate approaches Qp after 6 s, so the volume collected after 5 s or 10 s (as used by Soh et al.15) will be noticeably less than QptE in both cases for this viscous liquid.
![]() | ||
Fig. 6 Predicted evolution of scaled flow rate and volume dispensed from the pipette tip in Fig. 5 for a 6.08 Pa s Newtonian liquid with Qp = 80 μL s−1, P0 = 101325 Pa and V0 = 2000 μL. |
Soh et al.15 tested 33 Newtonian viscosity standards and reported the amount of liquid collected over time tE as the averaged flow rate, = V(tE)/tE. They found that
varied systematically with viscosity as
∝ μ−1/3, thereby offering a simple method of estimating the liquid viscosity. We demonstrate here that this result can be predicted by the model. For a Newtonian liquid, for large values of c′ (as in their tests), an approximate form of eqn (12) is (see ESI†)
![]() | (13) |
This can be integrated analytically to obtain the volume dispensed, giving the profile in Fig. 6 which follows the exact result closely. Writing the reciprocal of the characteristic time as φ, the analytical result for is
![]() | (14) |
Fig. 7 shows eqn (14) predicts the form of the experimental results reported by Soh et al.15 and gives good agreement with the ∝ μ−1/3 relationship they obtained by a regression fit to their data set. Better agreement could be obtained by adjusting the model parameters P0 and V0 for each case, but the uncertainty regarding the effect of capillarity and flow onset would remain. Soh et al.15 presented a scaling argument explaining the
∝ μ−1/3 relationship which did not consider the role of air compression in the flow mechanism. This model corrects that oversight and demonstrates the importance of air compression by analyzing V(t) data series collected in related experiments.
![]() | ||
Fig. 7 Comparison of predicted effect of viscosity on average flow rate for Newtonian viscosity standards reported by Soh et al. (2023) for tests with dispense period of 5 s and Qp = 50 μL s−1. Error bars indicated 95% confidence interval from 9 repeats. Loci show predictions of simplified model, eqn (14), with parameters P0 = 101325 Pa, V0 values indicated. Inset shows the data in the linearized form reported by Soh et al. |
Soh et al.15 also demonstrated that the pipette tip design affected the volume of liquid collected (and hence ). Their data, reproduced in Fig. 8, showed that a standard pipette tip (see ESI†), with smaller exit orifice, was more sensitive to liquid viscosity than the wide-bore tip in Fig. 5. A custom pipette geometry gave an intermediate response which was almost linearly sensitive to viscosity. Fig. 8 shows that these behaviors can be predicted by eqn (14). A single set of model parameters was used in the simulations in this case; better agreement could be obtained by matching these to the individual pipette tips as they have different filling characteristics.
![]() | ||
Fig. 8 Effect of Newtonian viscosity on predicted average flow rate (eqn (14)) for different tip designs, Qp = 50 μL s−1, 5 s dispense period. Points – experimental data, error bars indicated 95% confidence interval from 9 repeats. Loci – model predictions, with P0 = 101325 Pa and V0 = 2000 μL for both nozzles. β values – standard, 13.6 × 1010 m−3; custom, 6.2 × 1010 m−3; wide bore 1.5 × 1010 m−3. |
The dispense profiles were obtained by running the measurement protocol on power-law fluids at different set dispense flow rates. We used differential evolution to minimize the mean squared error between the numerical model and experimental data, with ln(K) and n as the fitted parameters. Due to the disparity in scales of K and n, we optimized over ln(K) instead of K to ensure better convergence. Furthermore, we applied L2 regularization terms to K and n, which penalize excessively large parameter values and help to balance the optimization between the fitted parameters. We imposed the following bounds for the fit: 0.001 Pa sn ≤ K ≤ 10 Pa sn and 0.1 ≤ n ≤ 1 for molasses; 1 Pa sn ≤ K ≤ 30 Pa sn and 0.1 ≤ n ≤ 1 for Cif. We note that the regularization terms and bounds should be adjusted for optimal fits depending on the fluids of interest. For the model, we selected values of P0 = 95 kPa and V0 = 8000 μL. Sensitivity analysis was conducted to calculate the 95% confidence intervals using a perturbation method, i.e. by systematically perturbing the best-fit parameters in small increments and recalculating the loss function until it exceeded a given threshold value.
Fig. 11(b) and (d) show the experimental dispense profiles and fitted solutions for several set dispense flow rates, displaying good fits to the data. We consider only the dispense profiles at high set dispense flow rates (≥50 μL s−1) because the stepped profiles at low flow rates result in unreliable fits. Despite the underlying assumptions of the analytical model, the extracted values of K and n are usefully close to the actual values. For molasses, the errors for the fitted K values are between 2% and 33% and for the fitted n values are between 2% and 10% across the three set flow rates; for Cif, the errors for the fitted K values are between 7% and 92% and for the fitted n values are between 8% and 192% across the three set flow rates. We acknowledge that the errors arising from the fits can be large and there is room for improvement through refinement of the model. Nevertheless, considering the wide, general bounds for K and n implemented for the fits, these results show promise in the proxy viscometer being used to screen rheological behavior in simple non-Newtonian liquids.
The following workflow would be used to characterize a fluid anticipated to exhibit power-law behavior:
(1) Run the experimental protocol at a high, low and intermediate flow rate.
(2) Compare the apparent viscosity obtained for the three flow rates. If these differ within the bounds of measurement uncertainty, and show a systematic change with increasing flow rate, run the protocol for further flow rates to collect a larger data set (in this work, five were used).
(3) Select bounds for n and K. Prior approximation of the fluid flow behavior would be useful, for example, shear-thinning or shear-thickening, and expected broad range for the consistency coefficient.
(4) Select regularization terms for n and K. The terms can be tuned algorithmically for best agreement across fits from the different flow rate data and to avoid being stuck at the imposed bounds.
(5) Perform the fits and average extracted parameters across different flow rate data.
It should be noted that many non-Newtonian fluids exhibit constant viscosity at low shear rates, and the ability to capture this behavior depends on the ability to access a wide window of shear rates.
The principal advantage of the proxy viscometer is that it is based on an affordable and easily accessible platform. The OT-2 is inexpensive compared to other automated liquid handlers and uses open-source Python API, hence it has gained traction in laboratories globally.23–25 As a pipetting robot, the OT-2 facilitates a variety of sample preparation tasks. This enables the proxy viscometer to be seamlessly integrated into fully automated workflows for high-throughput screening purposes and closed-loop experimentation.16–18 Additionally, the proxy viscometer can be incorporated into more complex workflows involving external operations beyond the OT-2 platform, for example via a robotic arm that transfers samples to and from the OT-2.
Several other high-throughput viscosity screening systems have been described in the literature. Walker et al.26 trained a convolutional neural network on video data of fluid motion and used the model to identify solvents and estimate sample viscosity. The methodology is fast and non-invasive, but is more suited for the coarse classification of viscosity within predetermined ranges. Furthermore, the approach is not extendable to non-Newtonian liquids. Deshmukh et al.20 proposed a viscometer based on the Hamilton Microlab Star liquid handling workstation, using calibration curves constructed from dispensed liquid mass and measured pressure to determine the viscosity of Newtonian liquids. The pressure profile in the pipette tip during transient flow was analyzed numerically and used to calculate viscosity as a function of shear rate for non-Newtonian liquids. While fast and reasonably accurate, the viscometer requires a costly liquid handling system with pressure measurements.
The proxy viscometer presented here exhibits several key improvements over our previous work.15 The real-time measurement of fluid mass dispensed provides crucial information about the dispense process, which we quantify via the parameters Vplat, m and s. The previous work relied only on the mass of fluid collected at the end of the dispense process, equivalent to Vplat. Here, the additional parameters m and s allow us to discern between fluids of lower viscosity, hence extending the lower viscosity limit of the proxy viscometer from 1500 cP to 20 cP. Furthermore, we introduce an analytical model to describe the dispense process. With the incorporation of this model, the proxy viscometer can be applied to characterize non-Newtonian liquids. Given the ubiquity of non-Newtonian liquids in industrial applications, this presents an exciting development towards practical implementation.
There is room for further development of the proxy viscometer. Firstly, the amount of liquid dispensed depends significantly on its viscosity, but is also influenced by other material properties, such as surface tension and wetting behavior.27 To improve the generality of the prediction model, we can run the measurement protocol on Newtonian liquids with a wider range of material properties and train the model with the properties as inputs alongside viscosity. Secondly, the analytical model approach has demonstrated promising results with power-law fluids. The model can be tested more rigorously with varying types of time-independent non-Newtonian liquids and extended beyond power-law fluids. Visualization of the charged pipette would help to eliminate variation between replicates and estimation of V0. Thirdly, temperature is a key determinant of viscosity and its effect has yet to be accounted for. Given the flexibility of the Opentrons platform and availability of temperature control solutions, it is feasible to extend the proxy viscometer to operate across a temperature range.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d4dd00368c |
This journal is © The Royal Society of Chemistry 2025 |