Sean Cross,
Christopher O'Rourke
and
Andrew Mills
*
School of Chemistry and Chemical Engineering, Queens University Belfast, Stranmillis Road, Belfast, BT9 5AG, UK. E-mail: andrewmills@qub.ac.uk
First published on 3rd July 2025
At present, micro-respirometry for measuring total viable count, O2 μR-TVC, is based on the time taken, TT, for an inoculum to significantly reduce the dissolved O2 level (typically from 21% to ≤ 10.5%). Here, a simple kinetic model relevant to μR-TVC is presented which describes the growth of the bacteria from an initial inoculum, No, to a maximum level, Nmax, and concomitant consumption of O2 and generation of CO2, in which the half-way time point, , corresponds to Nmax/No = 0.5, at which point %O2 = %CO2 = 10.5%. The model shows that it is not possible to reduce the TT in O2 μR-TVC below
, as TT increases above
with increasing sensitivity of the O2 sensor. In contrast, the same model shows that if a CO2 sensor is used instead, TT can be reduced significantly below
and consequently CO2 μR-TVC could be made much faster than conventional O2 μR-TVC. To test this model prediction, a range of colourimetric CO2 sensors of varying sensitivity, α, were prepared and used to make CO2 μR-TVC measurements. The results confirm that the greater the sensitivity of the sensor, the shorter the TT, as predicted by the kinetic model. Two CO2 indicators, one of moderate sensitivity and one of high sensitivity were used to generate straight-line log(CFU mL−1) vs. TT calibration plots, which can then be used to determine the unknown TVCs of subsequent samples. The future of CO2 μR-TVC as a possible new, faster alternative to conventional O2 μR-TVC is discussed briefly.
In μR for measuring TVC, i.e., O2 μR-TVC, the level of dissolved O2 is monitored as a function of incubation time, t. The measurement of O2 is usually carried out using an O2 sensor which employs a phosphorescent dye, D, the excited-state lifetime of which is quenched by molecular oxygen. The relationship between the measured excited-state lifetime, τ, and the ambient oxygen concentration, %O2, is described by the Stern–Volmer equation,
τ/τo = 1 + Ksv%O2 | (1) |
Typically, the dye used in O2 μR-TVC is a Pt porphyrin, such as Pt(II) tetraphenyl tetrabenzoporphyrin (PtBP), encapsulated in a polymer such as polyvinyl butyral (PVB), and has a luminescent lifetime of ca. 50 and 21 μs in the absence and presence of air, respectively, giving the sensor a Ksv value of 0.071% O2−1.5 In O2 μR-TVC, τ is monitored as a function of incubation time, and the threshold time, TT, is recorded when the value of τ reaches its midpoint, i.e., when τ = (τo + τair)/2, as the bacteria respire and deplete dissolved O2 in the growth medium. In O2 μR-TVC, it is found that log(TVC) of the bacterium under test is proportional to TT, and this feature is used to construct a log(TVC) vs. TT calibration graph which can then be used to determine the TVC of any subsequent test sample used to inoculate the growth medium, via its measured value of TT.5 Note that in O2 μR-TVC, eqn (1) is not usually used to calculate the %O2, instead it is sufficient to use the midpoint lifetime value, τ = (τo + τair)/2, to identify the value of TT.
When compared to the traditional APC method for measuring TVC, O2 μR-TVC is relatively rapid (usually < 12 h). Additionally, unlike APC, it is amenable to automation and thus commercial TVC-measuring systems based on O2 μR-TVC have been developed.6 However, although relatively fast, the speed of measurement in O2 μR-TVC is limited by the fact that TT corresponds to the time taken for the dissolved O2 concentration to drop significantly, from its initial value of 21% (as the initial inoculum/growth medium is air saturated) to typically ≤10.5%. This marked drop in %O2 only occurs when the bacterial load in the growth medium has increased substantially, typically to ≥5 × 107 CFU mL−1. It follows that for any aerobe used in O2 μR-TVC, the measured value of TT associated with any inoculum (like 104 CFU mL−1) is limited by the bacterium's growth kinetics and so cannot be reduced significantly by increasing the sensitivity of the O2 sensor.5
The above feature can be simply demonstrated using a kinetic model based on the often-cited continuous logistic equation for bacterial growth kinetics,7 which assumes that the rate of growth, dN(t)/dt, at any time, t, is proportional to both the size of the population, N(t), and the remaining material resources in the growth medium. The integrated form of the continuous logistic rate equation is,
N(t) = Nmax/[1 + A*exp(−kt)] | (2) |
%O2 = 21(1 − 1/[1 + A*exp(−t*)]) | (3) |
Using the kinetic model, it can be shown that in O2 μR-TVC, when the lifetime, τ, of the O2 sensor reaches its half-way value, (τo + τair)/2, at the associated threshold time, , the corresponding %O2 is given by the following expression,
%O2(TT) = 21/(2 + 21Ksv) | (4) |
In a simple model calculation, it was assumed that No and Nmax were 104 and 108 CFU mL−1, respectively, and eqn (3) was used to calculate the variation of %O2 vs. t* illustrated in Fig. 1. As expected, the profile shows that after the usual lag phase associated with bacterial growth, there is a rapid drop in the level of dissolved O2 as the bacteria enter their exponential growth phase. The %O2 vs. t* data in Fig. 1 and eqn (4) were then used to predict the variation in TT in O2 μR-TVC, , for sensors with sensitivities (Ksv) ranging from 0.01 to 0.49% O2−1. This range corresponds to using O2-sensitive dyes entrapped in the same polymeric medium, with τo values spanning the experimentally relevant range of 7 to 350 μs. The resulting plot of the Ksv vs.
is illustrated in Fig. 1 by the broken red line.
![]() | ||
Fig. 1 Plot of %O2 vs. unitless time parameter t*, calculated using eqn (3), assuming Nmax and No are 108 and 104 CFU mL−1, respectively. The broken red line is a plot of the variation in ![]() ![]() |
The Ksv vs. plot illustrated in Fig. 1 (broken red line) shows that as the sensitivity of the O2 sensor increases, the value of
rises above a minimum value of
, at which No/Nmax = 0.5 and %O2 = 10.5%. This plot illustrates the general feature of O2 μR-TVC, that for any given inoculum, the measured TT value is always greater than
and increases with increasing O2 sensor sensitivity.5
In contrast to the above, when the O2 sensor is replaced with a CO2 sensor, as in CO2 μR-TVC, the same kinetic model predicts that TT values can fall below . This suggests that, by employing highly sensitive CO2 sensors, CO2 μR-TVC could enable significantly faster TVC measurements than O2 μR-TVC. It is worth noting that CO2 μR-TVC is a relatively new technology, and most recent reviews of CO2 sensors focus on their applications in food packaging.9,10 In contrast, fluorescence-based O2 sensors are very well established and widely used in fields such as environmental monitoring, biomedical science and food packaging, as highlighted in two recent reviews.11,12 Consequently, unlike the emerging CO2 μR-TVC approach, O2 μR-TVC is a well-established technique that continues to attract academic research.8,13
The very short TT values predicted for CO2 μR-TVC represent a significant advantage over its established, commercial O2 μR-TVC counterpart, as it implies that, for the same bacterial sample under test, CO2 μR-TVC has the potential to measure TVC much more rapidly. O2 μR-TVC is already attracting significant interest, as it has been shown to be much faster (typically 3–4×) than the gold-standard APC method that is currently employed by most laboratories to measure TVC. Like O2 μR-TVC, and in contrast to APC, CO2 μR-TVC can be readily automated, is inexpensive and does not require significant technical support nor a large amount of plasticware. As such, CO2 μR-TVC has strong potential as a high throughput, ultra-rapid methodology for TVC determination. In this paper, a simple kinetic model is used to demonstrate that the TT in CO2 μR-TVC can be reduced significantly by increasing the sensitivity (α) of the CO2 sensor, and this prediction is validated using a range of easily fabricated CO2 sensors spanning a wide range of sensitivities. The results indicate that CO2 μR-TVC has the potential to become a leading technique for rapid TVC measurement.
Before use, each type of ink taken from the fridge was allowed to warm up to room temperature, 20 °C, (15 min) before being spread evenly on Tyvek (which provided a white, gas-permeable background support substrate) using a K-Bar No. 3 ink spreader.17
For each ink, the corresponding CO2 sensor was prepared by cutting a small square, ca. 5 mm, of the ink-coated Tyvek film and placing it between two 1 cm square layers of the 25 μm LDPE film. This sensor ‘sandwich’ was then heat-sealed using a heat press (VEVOR, London, UK) operated at 110 °C for 1 min. A wide range of such ‘sandwich’ LDPE/CO2 sensors were prepared in this way using both the solvent- and water-based inks, each incorporating one of the following pH-sensitive dyes – cresol red (CR), m-cresol purple (MCP), α-naphtholphthalein (NP), xylenol blue (XB), thymol blue (TB), o-cresol purple (OCP) and thymolphthalein (TP). The chemical structures and key properties (pKa value, absorption maxima of the protonated, HD, and deprotonated, D−, forms) of these dyes are given in S2 in the ESI.† Each heat-sealed LDPE layer acts as a gas-permeable but ion-impermeable barrier which prevents interference from any non-gaseous, usually ionic, species in the growth medium. Photographs of a typical 1 cm square, heat-pressed (and so LDPE covered), CO2 sensor are shown in S3 of Fig. S4 of the ESI,† showing a XB-silicone sensor in the presence and absence of CO2, where its respective colour is blue and yellow.
The stability exhibited by these CO2 sensors is demonstrated in Fig. S5 in the ESI,† which shows a plot of the unchanging measured apparent absorbance, A′, of a XB-silicone sensor held in the LB growth medium at 30 °C over a period of 14 days. Other work showed that its sensitivity also remained unchanged over this period. Although volatile organics, such as methane, are not expected to be generated by the E. coli bacterium under test, other work showed that such a gas, which is not acidic nor very reactive, had no effect on the sensor's performance, see Fig. S6 in the ESI.†
As illustrated in Fig. S7(a),† the CO2 sensor is placed with the sensor film facing the wall of the clear Falcon® tube so that it can be easily photographed and to prevent interference from any scattering or highly coloured species, which are obscured by the white Tyvek supporting substrate. The high gas permeability of the latter ensures the CO2 sensor responds promptly to the any change in the %CO2 dissolved in the growth medium on the non-indicator side of the Tyvek.
D− + CO2 + H2O ⇌ HD + HCO3− | (5) |
R = [HD]/[D−] = (Ao − A)/(A − A∞) = α·x%CO2 | (6) |
In CO2 μR-TVC, the threshold time, TT, is defined as the time at which the %CO2 reaches 1/α, the half-way colour change point. As in O2 μR-TVC, the value of log(CFU mL−1) is found to be directly proportional to TT for a given inoculum.21 Since α is primarily determined by the reciprocal of the dye's acid dissociation constant, Ka, CO2 indicator films with a wide range of sensitivities can be made by using pH indicator dyes with varying pKa values (pKa = −log(Ka)).
In a modified version of the kinetic bacterial growth model outlined earlier, it is assumed that the %CO2 generated at any time t mirrors the %O2 consumed, and is described by the simple expression, %CO2 = 21 − %O2. It follows that the model predicted variation in %CO2 generated as a function of incubation time is described by,
%CO2 = 21/[1 + A*exp(−t*)] | (7) |
![]() | ||
Fig. 2 Plot of %CO2 vs. unitless time parameter t*, calculated using eqn (7), assuming Nmax and No are 108 and 104 CFU mL−1, respectively. The broken red line is a plot of the calculated variation in ![]() |
A comparison of the model-predicted Ksv vs. and α vs.
profiles in Fig. 1 and 2, respectively, for the O2 μR-TVC and CO2 μR-TVC systems, highlights a key advantage of the CO2 μR-TVC approach. Specifically, by employing highly sensitive CO2 indicators, the value of
can be reduced well below the point at which N(t)/Nmax = 0.5, %CO2 = %O2 = 10.5% and t =
. Crucially, this suggests that CO2 μR-TVC, when implemented with high-sensitivity sensors, has the potential to achieve significantly faster TVC measurements than have been possible so far with conventional O2 μR-TVC systems.
![]() | ||
Fig. 3 (a) Photographs of the XB/silicone indicator in air at room temperature (20 °C), when exposed to different levels of %CO2 and (b) subsequent plot of the RGB-based data in (a) in the form of R vs. %CO2, where R was calculated using the apparent absorbance (vide infra), A′, values and eqn (6), and where the A′ values were determined via DCA of the photos in (a), with the ![]() ![]() |
Previous work by this group has demonstrated that digital colour analysis (DCA) of a CO2 colourimetric sensor can be used to generate apparent absorbance, A′, values which are directly proportional to the actual absorbance, A. As such, A′ can be used as a direct replacement for A in eqn (6) to calculate a R value for the %CO2 value under test, based on apparent (colour-based) values rather than actual (spectrophotometric-based) absorbance values.19 In this work, for each %CO2 gas mixture tested, the red component, RGB (red), was extracted from the Red–Green–Blue (RGB) profile of the sensor's image. The corresponding value of A′ was then calculated using the following expression,
A′ = log(255/RGB(red)) | (8) |
The reproducibility and accuracy of a typical CO2 sensor was tested using the same method as described in Fig. 3, using fixed mixture calibration gases, comprising, 0 (argon) and 0.1, 1, 5, 25 and 100% CO2 in air, to evaluate ten XB/silicone sensors in terms of their respective, , A∞ and α values. The results of this work are illustrated in Fig. S8 in the ESI† and reveal little variation in all three of these key parameters, with α = 0.74 ± 0.04 (5.4%) %CO2−1. All the CO2 sensors tested exhibited similar, small variances in these parameters, indicating the sensors have a good degree of accuracy and reproducibility for measuring CO2.
Finally, the 90% response time, t(90)↓, and recovery time, t(90)↑, of the XB/silicone CO2 sensor at 20 °C were determined by monitoring A′ as the sensor was cycled between air and 5% CO2. The results of this work are illustrated in Fig. S9 in the ESI,† from which t(90)↓ and t(90)↑ values of 2.8 and 7.1 min were calculated, respectively. These values are typical of all the CO2 sensors tested, and appropriate for use in %CO2 μR-TVC, since the colour transition due to the exponential growth phase typically occurs over 1–2 h.
Although it is possible to calculate the dissolved level of %CO2 in the growth medium at any time t from the calculated value of A′ using eqn (6), this is an additional and unnecessary step, just as it is unnecessary to calculate the %O2 from the recorded lifetime values in O2 μR-TVC. Instead, in CO2 μR-TVC, the value of TT is taken as the midpoint of the sensor's colour change, i.e., when A′ = ( +
)/2. This point is indicated by the broken red lines in Fig. 4(b). As will be shown later, repeating this experiment across a range of known bacterial loads yields a linear calibration curve of log(CFU mL−1) vs. TT, which can then be used to determine the TVC of unknown samples. As noted earlier, all microbiological assays were carried out in triplicate and the average value taken. In all cases, the standard deviation was ≤10%, which is significantly lower than that typically reported for the APC method (18% for counts above 30).22,23 Previous work using 3D-printed CO2 sensors has demonstrated that CO2 μR-TVC is statistically equivalent to O2 μR-TVC,21 which itself has been shown to be equivalent to the gold-standard APC method.24
The repeatability of the CO2 sensors in CO2 μR-TVC was tested by making ten XB/silicone sensors and running the same micro-respirometry experiment as outlined above, see Fig. 4. The results, illustrated in Fig. S10 in the ESI,† show a series of near superimposable A′ vs. t curves, with an average TT of 7.9 ± 0.2 h (ca. 2.5%). This variance is comparable to that reported for 3D-printed CO2 sensors (2.4%) and commercial O2 sensors (3.0%) in μR-TVC applications.21,24
Sensor durability and stability were also assessed by testing the performance of a single XB/silicone sensor held in the same (un-inoculated) growth medium on five consecutive days at 30 °C. As shown in Fig. S11 in the ESI,† the values of ,
and α remain unchanged over five consecutive days, indicating no loss in sensitivity. These findings are consistent with the results presented in Fig. S5,† which showed stable
values over a 14-day period.
Dye | pKa | α (%CO2−1) | TT (h) | Hourly photographic images of the sensor when inoculated with 104 E. coli |
---|---|---|---|---|
Solvent-based indicators | ||||
a An extra 1 mL of base was added to standard ink formulation.b An extra 2 mL of base was added to standard ink formulation. Photographs recorded every 30 minutes for TP* indicator. TB† refers to the TB/Ethyl cellulose (EC)/TOAH ink-based sensor.16 | ||||
CR | 8.2 | 0.09 | 9.0 | ![]() |
MCP | 8.3 | 0.22 | 7.9 | ![]() |
NP | 8.5 | 0.66 | 7.6 | ![]() |
XB | 8.9 | 0.76 | 7.5 | ![]() |
TB | 8.9 | 0.98 | 6.9 | ![]() |
TB† | 8.9 | 1.16 | 6.0 | ![]() |
As predicted by the kinetic model (see Fig. 2), for a fixed bacterial load, such as 104 CFU mL−1 used in this work, TT should decrease with increasing CO2 sensor sensitivity. This trend is supported by the plot of the data in Table 1 in the form of TT vs. α, illustrated in Fig. 5, which shows a clear linear relationship with a negative gradient. A t-test analysis comparing the slopes of the TT vs. α regression lines for the solvent- and water-based sensors found no statistically significant difference between them (P = 0.46).
![]() | ||
Fig. 5 Plot of TT vs. CO2 sensor sensitivity (α) using data taken from Table 1. The open and closed circles refer to water and solvent based CO2 sensors, respectively. The gradient and intercept of the line of best fit to all the data are −2.54 ± 0.16 and 9.13 ± 0.12, respectively. |
The plot shows that, as predicted by the kinetic model, the value of TT can be markedly reduced by increasing the sensitivity (α) of the CO2 sensor, thereby allowing much faster TVC measurements in CO2 μR-TVC. This is in direct contrast to O2 μR-TVC, where both model (Fig. 1) and experimental data5 show that TT increases with O2 sensor sensitivity (KSV) and is always greater than . The results in Fig. 5 suggest that using a very sensitive CO2 sensor could enable TVC to be measured much faster (possibly in minutes rather than hours) than is currently possible with conventional O2 μR-TVC, which itself is much faster than the standard APC method. This is very promising since CO2 μR-TVC, like O2 μR-TVC, is amenable to automation and so capable of measuring the TVC of many samples at the same time. Finally, the x-intercept value of the line of best fit to the data in Fig. 5 suggests that a sensor with an α value ≥3.6% CO2−1 would be capable of detecting CO2 production during the lag phase of bacterial growth, when the bacteria are respiring but not yet multiplying rapidly.
For simplicity, the kinetic model described in section 3 assumes that all the CO2 generated by the bacteria appears as dissolved CO2 in the growth medium, which, in practice is unlikely. However, the model also shows that the shape of the profile would be identical to that predicted by the model (see Fig. 2) if the pH of the growth medium remained constant during a run, as might be achieved using a pH buffer. For example, if the pH of the growth medium stayed at pH 7 through a kinetic run, although only 18% of the model predicted %CO2 would appear as dissolved CO2, the shape of the actual %CO2 vs. t run would be the same as that illustrated in Fig. 2, but with all the model predicted %CO2 values multiplied by 0.18.
In this work, no additional pH buffer was added, and the pH was found to remain at pH 7 for ca. 6 h before dropping to pH 6 after 10 h, as illustrated in Fig. S12 in the ESI.† The effects of this pH change on the model-predicted %CO2 vs. t* and α vs. curves are illustrated in Fig. S13 in the ESI,† and show that, despite this change in pH, the shapes of both curves remain similar to those illustrated in Fig. 2. Also unchanged is the key predicted feature of the model, namely that increasing sensor sensitivity reduces TT, thereby enabling faster analysis, as established experimentally by the TT vs. α plot illustrated in Fig. 5.
As noted earlier, this plot suggests a sensor with a sensitivity ≥3.6% CO2−1 would be ideal for rapid CO2 μR-TVC measurements. However, given the ambient level of CO2 in air is ca. 0.04%, it is likely that sensors with α ≥ 25% CO2−1 would likely be too sensitive for practical use in CO2 μR-TVC. For this reason, model predicted variations of α vs. were limited to the experimentally realistic range of 0.14 to 7.0% CO2−1.
![]() | ||
Fig. 6 (a) Photographic images of the XB/silicone indicator in a CO2 μR-TVC study, in which the bacterial load of E. coli was varied over the range 108 to 101 CFU mL−1, (b) plots of A′ vs. t for each different inoculum, calculated using DCA analysis of the images in (a) and eqn (8) (the red broken line highlights the mid colour-change point which was used to derive values for TT for each inoculum), (c) plot of the log(CFU mL−1) vs. TT data derived from (b). |
![]() | ||
Fig. 7 As in Fig. 6, (a) photographic images of the TP*/HPC indicator in a CO2 μR-TVC study, in which the bacterial load of E. coli was varied over the range 108 to 101 CFU mL−1, (b) plots of A′ vs. t for each different inoculum, calculated using DCA analysis of the images in (a) and eqn (8) (the red broken line highlights the mid colour-change point which was used to derive values for TT for each inoculum), (c) plot of the log(CFU mL−1) vs. TT data derived from (b). |
As seen in Fig. 6(b) and 7(b), both CO2 sensors give linear log(CFU mL−1) vs. TT calibration plots, which can then be used to determine the TVC of unknown samples of E. coli. The gradients of the lines of best fit in the calibration curves, Fig. 6(c) and 7(c), are −0.64 and −1.09 log(CFU mL−1) h−1, respectively, demonstrating that the more sensitive TP*/HPC sensor enables TVC measurements to be conducted approximately 70% faster than the XB/silicone sensor. For comparison, the same experiment conducted using an O2 sensor yields a gradient of −0.74 log(CFU mL−1) h−1, which is like that of the XB/silicone CO2 sensor, but, as expected, much lower (and so slower) than the TP*/HPC sensor. However, as noted earlier, the TP*/HPC sensor exhibits poor stability in growth media, for reasons that are unclear at present. This is made apparent from the near superimposable A′ vs. t plots illustrated in Fig. 7(b) for the 103–101 and control (no bacteria) runs, and the vertical drop in the otherwise straight-line trend observed in the log(CFU mL−1) vs. TT plot in Fig. 7(c). Thus, although this work shows that it is possible to create a very rapid method for the measurement of the TVC of aerobes (or anaerobes for that matter) based on CO2 μR-TVC using a highly sensitive CO2 sensor, the most sensitive CO2 sensors currently available, such as the TP*/HPC sensor, lack stability for their routine use. Encouragingly, all other CO2 sensors tested in this work were found to be stable in growth medium for several weeks, suggesting that with further work, a stable, highly sensitive alternative to TP*/HPC may be developed for reliable, rapid TVC measurements.
In contrast, the new CO2-based micro-respirometry method described here, CO2 μR-TVC, does not require the bacterial load to reach such high levels and so can be used to make much faster TVC measurements. This work demonstrates a clear, direct inverse relationship between TT and the sensitivity of the CO2 sensor, α, and that very short analysis times (<1 h) should be possible using very sensitive (α ≥ 3.3% CO2−1) CO2 sensors. However, to realise the very rapid measurement potential of CO2 μR-TVC, a stable, high sensitivity CO2 sensor needs to be developed, and work is currently in progress to achieve this goal.
The advantages of CO2 μR-TVC over the traditional APC method include all those associated with O2 μR-TVC, namely a simple, non-subjective, low-cost method for measuring TVC, which is amenable for automation and so the simultaneous, rapid analysis of many samples. Unlike APC, it does not require the use of a large amount of plasticware, nor significant technical support, and does not have a subjective element. APC has a subjective element in that the plate number returned is a skilled technician's interpretation of the number of colonies on the plate, which often varies from technician to technician and laboratory to laboratory.25 Furthermore, CO2 μR-TVC has been shown to be statistically equivalent to O2 μR-TVC,21 which in turn is equivalent to APC.24 Most importantly, CO2 μR-TVC is capable of much faster analysis times than O2 μR-TVC, which in turn are much faster than APC. Consequently, once an appropriate, high sensitivity CO2 sensor has been developed, it is likely commercial CO2 μR-TVC instrumentation will emerge that will make CO2 μR-TVC the go to alternative to the APC method that is currently employed in most microbiology laboratories.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d5sd00078e |
This journal is © The Royal Society of Chemistry 2025 |