In this study, a method has been used to prepare and evaluate the content of some metals in samples of eye shadow. The samples were manufactured in different countries (Brazil, China and USA). The sample preparation was performed using dilute nitric acid and a hot block. The Pb content was measured using graphite furnace atomic absorption spectrometry (GFAAS), and the quantification of Cd, Co, Cr, Cu and Ni was performed using inductively coupled plasma optical emission spectrometry (ICP OES). Chemometric tools were used for multivariate optimisation and to perform an exploratory analysis according to the metal concentration levels. Some samples presented metal concentrations above the values indicated by the FDA (USA) and ANVISA (Brazil).
The U.S. and Brazilian legislative branches do not specify limits for the metal concentrations in makeup; however, the FDA and ANVISA have established a threshold limit for the metal content in raw materials. According to the FDA regulations, the metal concentration limits depend on each additive and its color.7 According to the Brazilian regulations, the limits of some metals in organic artificial colourants are 3 mg kg−1 for As, 20 mg kg−1 for Pb and 100 mg kg−1 for other elements.8
Canada has a guide on heavy metals found as impurities in cosmetic products. The country follows the European legislation, however, by testing its own samples, cosmetic limits for some metals were established in the final product. The maximum tolerable limits for Pb, As, Cd and Hg are 10, 3, 3 and 5 mg kg−1, respectively.9
Metals such as Ni, Co and Cr as mentioned before are considered major causes of allergies, but this risk can be reduced if the level of these metals remains below 5 μg g−1.10
The sample preparation, in the scope of the analytical sequence, is the process most likely to introduce errors; it is time consuming and involves high costs.11 Usually, during the sample preparation, concentrated HNO3 and HF are used to completely digest the cosmetic samples (e.g., lipstick and eye shadow).6,12,13 In this context, sample preparation using dilute HNO3 is an interesting alternative and environmentally friendly. This sample preparation is easier, safer and in accordance with green chemistry due to the low consumption of reagents and the consequent reduction of laboratory residues.14 Furthermore there is a reduction of the blank signal and dilution is avoided before determination.15 Although the use of diluted nitric acid was extremely evaluative for different types of samples, there are no studies employing diluted acids for metal extraction in eye shadows.16,17
The efficiency of using diluted acids is a result of the temperature gradient inside the reaction vessel during the initial steps of sample digestion and due to the presence of a rich oxygen atmosphere inside the closed vessel. The nitric acid is regenerated by nitrogen oxide species with hydrogen peroxide as shown below:18
2NO + 3H2O2 → 2HNO3 + 2H2O |
In some cases strong treatments (concentrated acids) may not be feasible; following this thought some authors have proposed to employ a partial digestion of cosmetics.10,19 Some parameters can be studied to optimize the process of sample preparation such as acid concentration, sample mass, heating time and the use of other reagents employed for digestion of the samples.20
One of the strategies to optimise the process of sample preparation is multivariate analysis. This approach enables the collection of information about the effects of each evaluated variable and their interactions to find the best possible or most favourable conditions.21 One of the chemometric tools used for system optimisation is factorial design. Using this approach, a reduced number of experiments are performed, and the information is maximised.21 When factorial design and multielement analysis are combined, several responses can be obtained. In this case, the desirability function becomes an alternative method to normalise these responses and combine them in a single value named the overall desirability (D) and then achieve better working conditions for all responses evaluated.21 Thus, the goal of this study was the development of a simple method of cosmetic sample preparation and determination of Cd, Co, Cr, Cu and Ni using ICP OES and Pb using GFAAS in eye shadow. Several chemometric tools were used for multivariate optimisation and to perform an exploratory analysis by categorising the samples according to the metal concentration levels.
The multi-element standard solutions were prepared daily from 1000 mg L−1 Cd, Co, Cr, Cu, Ni and Pb stock solutions (Qhemis, Jundiaí, SP, Brazil) via dilution in 0.67 mol L−1 HNO3. This acid solution was also used as the blank.
A mixture of magnesium nitrate solution (0.03% w/v) and ammonium dihydrogen phosphate (0.5% w/v) solution was used as a chemical modifier in Pb determination using GFAAS.
GFAAS (iCE 3000 Series, Thermo Scientific, Waltham, MA, USA) was used for Pb determination. The heating program (Table 1) for Pb determination was used under the recommended conditions provided by the manufacturer. For each measurement the autosampler of the GFAAS collected 15 μL of the sample or reference solution along with 5 μL of the modifier solution (a mixture of magnesium nitrate and ammonium dihydrogen phosphate), so this mixture was introduced into the graphite tube. Preliminary tests in ICP OES demonstrate high values of recuperation indicating possible spectral interferences arising due to the sample complexity. Graphite AAS and the mixture of chemical modifiers such as magnesium nitrate and ammonium dihydrogen phosphate were applied to overcome these problems. All measurements of integrated absorbance were made in triplicate.
Step | Temp (°C) | Time (s) | Ramp (°C s−1) | Gas flow (L min−1) |
---|---|---|---|---|
1 | 100 | 30 | 10 | 0.2 |
2 Pyrolysis | 800 | 20 | 150 | 0.2 |
3 Atomisation | 1200 | 3 | 0 | Off |
4 | 2500 | 3 | 0 | 0.2 |
The samples were also digested using a microwave oven (Speedwave four, Berghof) furnished with 12 digestion vessels (TFM™-PTFE) with an internal volume of 100 mL (DAP-100+). An analytical balance (model AY 220, max. 220 g, 0.1 mg resolution, Shimadzu, Kyoto, Japan) was used for sample preparation.
The samples were accurately weighed and were mineralised in a block digester. In the mineralization, 100 mg of the eye shadow sample was weighed in PFA tubes, followed by the addition of 5 mL of HNO3 (2 mol L−1), 2 mL of H2O2 (30% w/w) and 1 mL of Triton X-100 (5% w/v). The tubes were closed, the mixture was heated at 100 °C for 180 min and the solutions were quantitatively transferred to polypropylene flasks and diluted with water to 15 mL. The mineralization was made in triplicate with and without the addition of a standard to verify the accuracy of the analytical method and to detect possible losses of analytes during sample preparation. The final concentrations added were 40, 40, 700, 700, 200 and 80 μg L−1 for Cd, Co, Cr, Cu, Ni and Pb, respectively.
A sample of black color was used for microwave digestion with HF. Two hundred mg of each eye shadow sample was weighed into a Teflon (DAP 100) vessel and 6 mL of concentrated HNO3, 2 mL of H2O2 (30%) and 2 mL of concentrated HF were added to it.6 After cooling, the samples were transferred to 50 mL volumetric flasks, 24 mL of H3BO3 (4%) was added and the flasks were filled up to the required volume with deionized water. The microwave oven heating program performed was composed of two steps: (1) 5 minutes ramp (800 W), 5 minutes holding time (800 W); (2) 5 minutes ramp (1800 W), 30 minutes holding time (1800 W). The maximum temperature and pressure for steps 1 and 2 were 180 °C and 70 bar and 210 °C and 70 bar, respectively.
Triton X-100 was used as a surfactant because it was observed that the eye shadows remained on the solution surfaces due to the small particle size and a high value of surface water tension, and the surfactant helps with homogenization.22
The analyses were performed in axial and radial modes and a total of 12 responses were obtained for each experiment.
As several responses were analysed simultaneously, an important aspect is how to combine these data. In this case, the desirability function23 was used, and the signals were normalised between 0 (the lowest signals) and 1 (the highest signals). Eqn (1) shows how this transformation was performed:
(1) |
After the calculation of the individual desirability values (di), it is possible to combine the results and obtain the global desirability (D), calculated using the geometric mean (eqn (2)):
(2) |
The global desirability (D) values can be viewed in Table 2. As can be observed, the best working conditions (values near 1) for axial mode (D1) were obtained in experiment 5 (0.95), while for radial mode (D2), the best result was obtained in experiment 8 (0.96). However, the goal was to identify a commitment condition for both modes. Thus, a new calculation of the geometric mean was made using the global desirability values of each mode .
Experiments | Desirability values | ||
---|---|---|---|
Axial (D1) | Radial (D2) | Axial and radial (D) | |
1 | 0.62 | 0.64 | 0.63 |
2 | 0.44 | 0.47 | 0.46 |
3 | 0.88 | 0.83 | 0.85 |
4 | 0.71 | 0.77 | 0.74 |
5 | 0.95 | 0.89 | 0.92 |
6 | 0.85 | 0.88 | 0.86 |
7 | 0.93 | 0.93 | 0.93 |
8 | 0.90 | 0.96 | 0.93 |
After performing the new global desirability calculation, it was observed that the experiments 5, 7 and 8 showed the best conditions with very similar desirability values (ranging from 0.92 to 0.93). The acid concentration used in the experiments 7 and 8 was 7 mol L−1, while in experiment 5 it was 2 mol L−1. Thus, the acid concentration was a decisive parameter, and the experimental condition presented in experiment 5 (2 mol L-1) was selected as the ideal. In addition the advantages of the use of dilute acid, which require lower dilution for analysis by ICP OES, have been reported. Table 3 shows the final conditions of the sample preparation.
Parameters | Conditions |
---|---|
Heating time | 3 hours |
Temperature | 100 °C |
Volume of H2O2 (30% m/v) | 2 mL |
Sample mass | 250 mg |
Volume and concentration of HNO3 | 5 mL (2 mol L−1) |
Volume of Triton X-100 | 1 mL (5% m/v) |
Variables | Conditions |
---|---|
V1: Integration time for low emission line (s) | 5 |
V2: Integration time for high emission line (s) | 5 |
V3: Sample introduction flow rate (mL min−1) | 4.2 |
V4: Sample flow rate during the analyses (mL min−1) | 2.1 |
V5: Pump stabilisation time (s) | 25 |
V6: Radio frequency applied power (W) | 1200 |
V7: Auxiliary gas flow rate (L min−1) | 0.25 |
V8: Nebulisation gas flow rate (L min−1) | 0.83 |
V9: Cooling gas flow rate (L min−1) | 16 |
After optimising the conditions of sample preparation and instrumental parameters of the ICP OES, the analytical performance parameters were evaluated. Table 5 shows the limits of detection and quantification, linearity, sensitivity and precision to axial and radial views in ICP OES and GFAAS (Pb determination).
Parameters | Cd 228.802 nm | Co 228.616 nm | Cr 357.869 nm | Cu 224.700 nm | Ni 341.476 nm | Pba 283.3 nm |
---|---|---|---|---|---|---|
a GFAAS. b Radial view. | ||||||
Calibration curve | y = 7.3x + 10.4 | y = 5.03x + 6.25 | y = 14.6x + 5.9 | y = 2.7x + 16.8 | y = 3.7x + 2.6 | y = 0.004x + 0.008 |
y = 1.03x + 1.28b | y = 0.7x + 0.6b | y = 0.9x + 0.6b | y = 0.3x + 1.8b | y = 0.3x + 0.1b | ||
Linearity (μg L−1) | 2.5–80 | 5–1500 | 5–1500 | 5–1500 | 5–1500 | 5–120 |
Regression coefficient (r) | 0.99 | 1.00 | 1.00 | 1.00 | 1.00 | 1.00 |
LOD (mg kg−1) | 0.03 | 0.06 | 0.06 | 0.08 | 0.3 | 0.02 |
0.2b | 0.2b | 0.8b | 0.8b | 0.9b | ||
LOQ (mg kg−1) | 0.1 | 0.2 | 0.2 | 0.3 | 0.9 | 0.06 |
0.7b | 0.7b | 3b | 3b | 3b | ||
Precision (%) n = 10 | 0.2 | 0.2 | 0.3 | 0.2 | 0.1 | 0.9 |
0.9b | 0.7b | 0.7b | 0.4b | 0.8b |
Particularly for the calculation of limits of detection in ICP OES, the concept of background equivalent concentration (BEC), defined as the concentration of the analyte that produces a signal equivalent to the emission intensity of the background at the spectral line measured, was used. The incorporation of BEC values in the calculation of LOD allows the evaluation of any change in operational conditions.25 The LOD and LOQ for the GFAAS were calculated from 10 independent blank samples measured once each in accordance with IUPAC recommendations.26
The precision was evaluated in terms of repeatability and the relative standard deviations (RSDs) of 10 measurements of a multielement solution were determined with a concentration of 40 μg L−1 for Cd, 50 μg L−1 for Pb and 250 μg L−1 for Co, Cr, Cu and Ni. The precision for each element is adequate according to the standards of the acceptability criterion set by INMETRO (Instituto Nacional de Metrologia) which establishes an acceptable relative standard deviation of up to 10% for concentrations above 100 ng g−1.27 The linearity was confirmed in the working range of each element, and the correlation coefficients found were equal to or exceeded 0.99.
The methods with analytical performance parameters are adequate to determine these elements in eye shadow samples and meet the limits required by legislations.7,9
Certified reference materials similar to eye shadows were not available; thus, the validity of the proposed method was verified by addition–recovery studies and by comparing the proposed method with the conventional acid digestion procedure performed with a microwave oven (closed vessel system) and HF.
The results, obtained from the proposed method and the microwave-assisted digestion procedure (Table 6), were compared using the paired t test. The sample used was the same employed in all optimisation studies. The results obtained by both procedures for all elements were in concordance at the 95% confidence level. The values of Cd were lower than the LOQ (see details in Table 5).
Cd | Co | Cr | Cu | Ni | Pb | |
---|---|---|---|---|---|---|
Concentration by the proposed method (mg kg−1) | <LOQ | 3.8 ± 0.1 | 30.4 ± 0.3 | 39 ± 1 | 19.5 ± 0.6 | 10 ± 1 |
Concentration by the comparative method (mg kg−1) | <LOQ | 3.7 ± 0.2 | 38 ± 3 | 42 ± 1 | 16.8 ± 0.9 | 8.3 ± 0.7 |
Recovery rates between 80% and 120% were obtained (Table 7); these findings are considered according to FDA7 (which requires recoveries between 80% and 120% for concentrations above 1 μg g−1). Thus, the results of the performance parameters studied confirmed quality assurance when using the proposed method for the determination of Cd, Co, Cr, Cu, Ni and Pb in eye shadow samples.
Concentration (mg kg−1) | Recovery (%) | |||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Samples | Origin country | Cd | Co | Cr | Cu | Ni | Pb* | Cd | Co | Cr | Cu | Ni | Pb* | |
Children | 1-Purple | China | 0.13 ± 0.01 | 0.29 ± 0.02 | 1.07 ± 0.01 | <LOQ | <LOQ | 1.34 ± 0.08 | 94 | 81 | 85 | — | — | 115 |
2-Green | China | <LOQ | 0.32 ± 0.02 | 0.78 ± 0.06 | 0.74 ± 0.01 | <LOQ | 0.94 ± 0.05 | — | 83 | 85 | 90 | — | 119 | |
3-Blue | China | <LOQ | 0.34 ± 0.04 | 1.39 ± 0.09 | <LOQ | <LOQ | 0.99 ± 0.04 | — | 85 | 83 | 67 | — | 114 | |
4-Blue | China | <LOQ | 0.37 ± 0.02 | 1.06 ± 0.01 | 0.9 ± 0.3 | <LOQ | 1.00 ± 0.03 | — | 83 | 91 | 85 | — | 115 | |
5-Red | China | <LOQ | <LOQ | 2.5 ± 0.2 | 1.1 ± 0.1 | <LOQ | 7.7 ± 0.2 | — | — | 90 | 86 | — | 100 | |
6-Orange | China | 4.9 ± 0.2 | <LOQ | 3.5 ± 0.1 | 1.1 ± 0.2 | <LOQ | 44 ± 2 | 89 | — | 89 | 83 | — | 120 | |
7-Pink | China | <LOQ | 0.29 ± 0.01 | 0.65 ± 0.03 | 0.4 ± 0.2 | <LOQ | 0.46 ± 0.04 | — | 80 | 88 | 83 | — | 86 | |
8-Yellow | China | <LOQ | 0.47 ± 0.02 | 1.8 ± 0.1 | 0.9 ± 0.3 | <LOQ | 0.50 ± 0.06 | — | 100 | 98 | 103 | — | 103 | |
9-Blue | China | 6.0 ± 0.1 | <LOQ | 0.89 ± 0.03 | 48.2 ± 0.9 | <LOQ | 34 ± 4 | 89 | — | 101 | 105 | — | 105 | |
10-Pink | China | 4.6 ± 0.1 | <LOQ | 1.05 ± 0.05 | 0.48 ± 0.03 | <LOQ | 26 ± 2 | 87 | — | 102 | 90 | — | 97 | |
11-Black | China | <LOQ | 3.88 ± 0.09 | 44 ± 3 | 16.7 ± 0.6 | 23.1 ± 0.5 | 1.36 ± 0.08 | — | 92 | 91 | 90 | 103 | 81 | |
12-Blue | China | <LOQ | 2.4 ± 0.4 | 2.3 ± 0.3 | 4.6 ± 0.4 | <LOQ | 1.54 ± 0.06 | — | 80 | 94 | 98 | — | 94 | |
13-Pink | China | <LOQ | 1.47 ± 0.04 | 1.75 ± 0.06 | 3.2 ± 0.5 | <LOQ | 1.59 ± 0.07 | — | 91 | 92 | 98 | — | 95 | |
14-Gold | China | <LOQ | 1.22 ± 0.05 | 11.5 ± 0.7 | 15.3 ± 0.8 | 3.4 ± 0.1 | 2.25 ± 0.05 | — | 86 | 92 | 87 | 95 | 95 | |
Adults | 15-Black | China | <LOQ | 4.9 ± 0.2 | 62.7 ± 0.9 | 21.1 ± 0.7 | 10.4 ± 0.3 | 1.05 ± 0.09 | — | 90 | 97 | 102 | 98 | 86 |
16-Black | China | <LOQ | 1.8 ± 0.2 | 2.4 ± 0.2 | <LOQ | 2.8 ± 0.9 | 2.0 ± 0.3 | — | 91 | 91 | 116 | 101 | 93 | |
17-Black | Brazil | <LOQ | 3.8 ± 0.5 | 28 ± 4 | 36 ± 1 | 19.5 ± 0.5 | 0.09 ± 0.01 | — | 89 | 91 | 94 | 92 | 91 | |
18-Black | Brazil | <LOQ | 3.5 ± 0.7 | 29 ± 1 | 37 ± 2 | 19.6 ± 0.8 | 0.58 ± 0.05 | — | 93 | 96 | 96 | 100 | 117 | |
19-Black | USA | <LOQ | 3.45 ± 0.09 | 12.8 ± 0.4 | 39.1 ± 0.7 | 4.9 ± 0.2 | <LOQ | — | 80 | 89 | 81 | 95 | — | |
20-Black | USA | <LOQ | 3.4 ± 0.1 | 91 ± 1 | 26.5 ± 0.6 | 16.0 ± 0.4 | 0.78 ± 0.06 | — | 92 | 90 | 94 | 93 | 86 | |
21-Blue | China | 0.39 ± 0.01 | 1.02 ± 0.05 | 1.0 ± 0.1 | 61 ± 8 | <LOQ | 1.74 ± 0.07 | 103 | 103 | 101 | 101 | — | 96 | |
22-Blue | China | 0.41 ± 0.01 | 0.80 ± 0.07 | 0.5 ± 0.2 | <LOQ | <LOQ | 1.22 ± 0.04 | 106 | 110 | 96 | — | — | 82 | |
23-Blue | Brazil | 0.9 ± 0.1 | <LOQ | <LOQ | <LOQ | <LOQ | 1.8 ± 0.2 | 93 | — | — | — | — | 85 | |
24-Blue | Brazil | <LOQ | 1.85 ± 0.07 | 31 ± 1 | 10.0 ± 0.4 | 3.2 ± 0.3 | 1.09 ± 0.09 | — | 91 | 93 | 104 | 92 | 94 | |
25-Blue | USA | <LOQ | 2.2 ± 0.09 | 21.8 ± 0.9 | 9.7 ± 0.4 | 10.0 ± 0.4 | 1.51 ± 0.08 | — | 87 | 88 | 82 | 87 | 84 | |
26-Orange | China | <LOQ | 0.8 ± 0.1 | 17.2 ± 0.7 | 9.2 ± 0.5 | 3.3 ± 0.3 | 1.84 ± 0.07 | — | 100 | 95 | 98 | 96 | 90 | |
27-Orange | China | 0.36 ± 0.02 | 0.58 ± 0.03 | 1.73 ± 0.09 | 0.90 ± 0.08 | 0.73 ± 0.06 | 3.3 ± 0.2 | 83 | 103 | 98 | 108 | 99 | 93 | |
28-Orange | Brazil | 0.41 ± 0.02 | 0.37 ± 0.04 | 3.0 ± 0.2 | 7.5 ± 0.5 | 2.9 ± 0.2 | 0.59 ± 0.06 | 80 | 80 | 85 | 98 | 80 | 86 | |
29-Orange | Brazil | <LOQ | 0.75 ± 0.01 | 2.4 ± 0.1 | <LOQ | 1.23 ± 0.07 | 0.10 ± 0.02 | — | 84 | 87 | — | 88 | 89 | |
30-Orange | USA | <LOQ | 0.88 ± 0.04 | 7.4 ± 0.1 | <LOQ | 16.9 ± 0.3 | 1.3 ± 0.2 | — | 91 | 89 | — | 89 | 108 | |
31-Pink | China | <LOQ | 0.36 ± 0.02 | 1.7 ± 0.1 | 1.14 ± 0.06 | 1.8 ± 0.5 | 3.21 ± 0.09 | — | 83 | 94 | 99 | 98 | 104 | |
32-Pink | China | <LOQ | 0.10 ± 0.02 | 0.81 ± 0.07 | <LOQ | <LOQ | 3.40 ± 0.09 | — | 87 | 89 | — | — | 90 | |
33-Pink | Brazil | <LOQ | 0.15 ± 0.03 | 3.0 ± 0.6 | 21 ± 3 | <LOQ | 0.33 ± 0.09 | — | 97 | 82 | 107 | — | 95 | |
34-Pink | Brazil | <LOQ | <LOQ | 2.1 ± 0.2 | <LOQ | <LOQ | 2.1 ± 0.2 | — | — | 94 | — | — | 86 | |
35-Pink | USA | <LOQ | 0.44 ± 0.02 | 1.00 ± 0.07 | <LOQ | <LOQ | 1.4 ± 0.1 | — | 86 | 89 | — | — | 107 |
To analyse the behaviour of the samples in relation to the levels of the elements, PCA (Principal Component Analysis) was performed on a 104 × 11 data matrix that contained samples with three authentic replicates in the rows (except for the sample 16, which contains two authentic replicates) and the analytes in the columns, which were determined in the axial and radial modes (except for Pb). The data were auto-scaled and Pirouette version 4.5 was used in the calculation (Infometrix, Bothell, USA).
A model with 4 principal components (PCs) was selected to evaluate the behaviour of the samples and explained 96% of the total variance. The results are shown in Fig. 1 as graphical representations of the scores (related to samples, Fig. 1a) and loadings (related to the analyte, Fig. 1b).
Fig. 1 Graphical representation of the (a) scores and (b) loadings of PC1 × PC2 in relation to element concentrations presented in Table 7. |
The authentic replicates and data obtained from the radial and axial views are very close, showing the low standard deviation in the measurements, but it was not possible to see the separation between the child and adult samples in the figure.
If the scores (Fig. 1a) are analysed with the graph of loadings (Fig. 1b), it is possible to see that some child samples have high values of Cd and Pb. These samples have values which are above those recommended by the legislation.
It is also possible to observe the separation of the samples with respect to the black color. The black color samples (see dotted line ellipses) have higher concentrations of Co, Cr, Cu and Ni.
Under the optimised conditions of sample preparation and instrumental parameters of the ICP OES and GFAAS, it was possible to determine the Cd, Co, Cr, Cu, Ni and Pb concentrations in the eye shadow samples from different brands, consumers, colors and countries and, by employing chemometric tools, observing separation of the samples according to the concentration levels of the elements.
The analyte concentrations found in some samples were above the values recommended for both adult and child samples. These results confirm the importance of quality control in the production of cosmetics as well as the applicability of the proposed method.