Pablo H. S. Martinsa,
Maria A. Barrosa,
Caroline L. Silvaa,
Poliana Riccia,
Laís M. B. Castilhoa,
Allyson L. R. Santosa,
Hugo S. Rodriguesb,
Rosana M. N. Assunçãoab and
Anizio M. Faria*ab
aInstituto de Química, Universidade Federal de Uberlândia, 38304-402 Uberlândia, MG, Brazil. E-mail: anizio@ufu.br
bInstituto de Ciências Exatas e Naturais do Pontal, Universidade Federal de Uberlândia, 38304-402 Ituiutaba, MG, Brazil
First published on 29th May 2024
This work presents an eco-friendly approach for determining free glycerol in biodiesel samples, using a cellulose monolith stir bar in the sorptive extraction method with analysis by high-performance liquid chromatography and a refractive index detector. The cellulose monolith was produced from cellulose acetate by non-solvent-induced phase separation and subsequent alkaline deacetylation. The cellulose monolith presented a hierarchically porous structure, with 68% porosity and almost total deacetylation, with morphological and polarity characteristics that favor an efficient extraction of free glycerol from biodiesel. The sorptive extraction method using a cellulose monolith stir bar was optimized, obtaining a total extraction time of 30 min at 70 °C, using ultrapure water as the desorption solvent, and extraction of free glycerol of 93.6 ± 2.3%. The proposed method showed selectivity in free glycerol extraction, with limits of detection and quantification of 6.60 × 10−5% w/w and 2.18 × 10−4% w/w, respectively. Compared with the official reference method, the proposed one presented similar precision and accuracy, with few manipulations and any reagent/solvents. Furthermore, it is compatible with the principles of green chemistry and can be considered an eco-friendly method for determining free glycerol in biodiesel.
The official American Oil Chemists' Society (AOCS) method Ca is recommended to determine free glycerol in biodiesel. 4-56,11 and American Society for Testing and Materials (ASTM) D6584-21,12 and European Norm (EN) 14105:2020 (ref. 13) standard test methods. However, the AOCS 4-56 method is a volumetric one that uses several reagents, some with considerable toxicity, for example, sodium arsenite. In addition, this method is highly laborious and dependent on the analyst. ASTM D6584 and EN 14105 methods are gas chromatographic (GC) methods that require expensive derivatization agents and internal standards. In addition, the GC methods use high temperatures for the column oven and detector, close to the temperature limits for these components, and require thermally stable GC columns.14–17 Although the chromatographic methods are recognized for their accuracy in glycerol quantification, these conditions make their application difficult in most analytical laboratories since they make the process more expensive and less reproducible with the derivatization of glycerol before the chromatographic analysis.5,18
In this context, several methods for determining free glycerol in biodiesel have been proposed, aiming at a more viable, simple, and environmentally friendly.14–31 Among these methods, approaches employing electroanalytical,15,19–21 chromatographic,14,22–24 electrophoretic,17,25 spectrophotometric16,26–30 or volumetric31 techniques have been evaluated. However, most of these methods require the use of chemical reagents and organic solvents, such as derivatization reagents, dissolution solvents, and adsorbents, or specific devices (electrodes, apparatus, glassware, cartridges), which are not accessible and very expensive, maintaining necessary the search for a simple method for the determination of free glycerol in biodiesel.
A strategy that has been consolidated is the previous step of extracting glycerol from biodiesel since the joint determination can be limiting for samples of biodiesel produced from some types of raw materials (vegetable oils and animal fat).14,16,29 Solid phase extraction has been studied for the extraction of glycerol with subsequent quantification of the extract by high-performance liquid chromatography, in which the analysis time is shorter than GC one and sample derivatization is unnecessary.14 Stir-bar sorptive extraction (SBSE) is a promising alternative for free glycerol extraction from biodiesel.32,33 SBSE is an extraction method that uses a stir bar coated with an analyte-selective adsorbent in a complex matrix.34 As main advantages, this method is efficient, requires few steps and little analyst manipulation in the extraction process, and can concentrate the analyte during extraction, providing a gain in detectability to modern instrumentation.32–34 The stir bar coating can ensure analyte selectivity with the proper choice of adsorbent material. For glycerol, cellulose can be highly attractive for its extraction from biodiesel.35–39 Several studies show cellulose has a high adsorption capacity for glycerol due to its similar polarity and a strong tendency to form hydrogen bonds with the glycerol molecule. Cellulose has already been analyzed as a filtration membrane,35 pulp,36,37 and residual biomass4,38 as a biodiesel purification method. The use of cellulose as a stir bar coating can result in a process that is fully compatible with the principles of green chemistry, as in addition to using a biopolymer, which is widely available, it may require only water as a desorption solvent for glycerol extraction of biodiesel.
Due to the lack of a simple and eco-friendly method for the determination of glycerol in biodiesel, in this work, we propose a strategy based on the SBSE extraction of glycerol from biodiesel samples followed by HPLC analyses with a refractive index detector (RID) without the use of any chemical reagents, organic solvents, and sophisticated apparatus or devices. The monolithic cellulose-coated stirring bar will be prepared from cellulose acetate (CA) solutions using non-solvent-induced phase separation (NIPS),39 followed by deacetylation of the CA monolith. Magnets will be embedded in the cellulose monoliths during phase separation, and the bars will be employed for glycerol extraction.
Cellulose monolith stir bars (CMSB) were obtained through the alkaline hydrolysis of CAMSB. The CAMSB were immersed in 10 mL of a 2.0 mol L−1 solution of NaOH in ethanol and degassed for 5 min under a gentle inert flow. The mixtures were left at room temperature (∼27 °C) for 24 h to deacetylate the monoliths. Afterward, the CMSB was carefully removed from the solution and washed with ultrapure water until neutral (pH ∼7). Fig. S2 in ESI† presents a scheme for preparing CAMSB and CMSB.
(1) |
The infrared spectroscopic measurements were obtained using a Cary 630 FTIR spectrometer (Agilent, Santa Clara, CA, USA) with an attenuated total reflectance (ATR) accessory to evaluate spectral differences between CAMSB and CMSB. A small piece of the monoliths, ∼10 mg, was placed directly on the diamond crystal for the measurement. The FTIR spectrum background was corrected using a freshly prepared pure KBr pellet. The IR spectral range was evaluated from 650 to 4000 cm−1 using a resolution of 4 cm−1 and a scan rate of 32 scans min−1.
The thermal stability of CAMSB and CMSB was studied using samples of ∼5 mg, heating from 25 °C to 600 °C at 10 °C min−1 in a nitrogen gas atmosphere, using a TA instruments TGA-55 model (New Castle, DE, USA). The DSC experiments were performed using DSC-25 equipment (TA Instruments) at heating and cooling rates of 10 °C min−1 from 25 °C to 300 °C in a nitrogen gas atmosphere.
Images of the particles were used to evaluate CMSB morphologically. Samples were sputter-coated with gold and examined with a Tescan model Vega3 scanning electron microscope (Brno, Czech Republic) at 20 kV.
The SBSE method, using the CMSB to extract glycerol from the biodiesel sample, was optimized to evaluate the adsorption and desorption time, desorption solvent, and temperatures of sorption and desorption. The extraction procedure initiates by contacting 5.0 mL of the spiked biodiesel sample with 0.02% w/w glycerol under magnetic stirring at 200 rpm with a CMSB for different times (15 min, 30 min, 60 min, and 2880 min) at various temperatures (30, 40, 60 and 70 °C). After that, excess methyl oleate from the CMSB was removed with paper towels. Then, the CMSB was placed in contact with 5.0 mL of desorption solvent (water, methanol, 0.02 mol L−1 NaOH sol, 0.02 mol L−1 HNO3 sol.), maintaining magnetic stirring at 200 rpm for different times (15 min, 30 min, 60 min, and 2880 min) under different temperatures (30, 40, 60 and 70 °C). Subsequently, HPLC-RID analyzed a 1.0 mL aliquot of the extract.
The reuse of CMSB was evaluated in the extraction of glycerol from biodiesel samples. A volume of biodiesel fortified with 0.020% w/w glycerol was prepared and divided into 5 mL portions. Triplicates of the portions were extracted using CMSB in the optimized SBSE method. After extraction, the CMSB was placed in 1 L of ultrapure water for 24 h, under gentle stirring. To evaluate total removal, the CMSB was placed in 5 mL of water at 70 °C for 15 min and analyzed by HPLC-RID. After that, the CMSB was again subjected to extraction of new portions of biodiesel fortified with glycerol, analyzing the extract by HPLC-RID. This procedure was repeated four times.
Standard glycerol solutions were prepared in ultrapure water at concentrations of 0.00022, 0.00044, 0.00087, 0.00435, 0.00870, 0.01740, and 0.03480% w/w for the analytical curve. The limits of detection and quantification of glycerol were obtained by sequentially injecting aqueous glycerol solutions in decreasing concentrations until a signal of 3× and 10× the noise level of a blank sample in the glycerol retention time, respectively.
(2) |
Application of the SBSE method for glycerol in biodiesel (B100) samples.
The proposed method for determining glycerol in biodiesel was applied to two biodiesel (b100) samples obtained from the transesterification of soybean oil with methanol and ethanol,41 resulting in methylic and ethylic soybean biodiesel samples. The total fatty acid esters and composition of biodiesel samples were obtained by gas chromatography-flame ionization detection using the EN: 14103 method.42 Samples without and with 0.02% w/w glycerol fortification were analyzed by optimized SBSE-HPLC-RID and standard AOCS 14-56 methods.
Differential Scanning Calorimetry (DSC) was used to characterize the cellulosic monoliths through the heat flux in the monoliths, analyzing their phase transitions, changes in their properties, or their degradation. Fig. 1 shows the DSC curves for the cellulose acetate and cellulose monoliths.
The two DSC curves show a primary thermal event related to the water outflow from the monoliths at 94 °C for cellulose acetate and 108 °C for cellulose as a high-intensity endothermic peak. A second endothermic peak on the DSC curve for the cellulose acetate monolith (CAM) was recorded at 234 °C and is associated with the melting of cellulose acetate.44 This same endothermic peak is not observed in the DSC curve for the cellulose monolith (CM), as expected, since the melting temperature of cellulose is above 300 °C.45
Thermogravimetric analysis of the monoliths was also performed, observing different TG curves for cellulose acetate and cellulose (Fig. 3S, ESI†). The TG curve for the CAM showed a thermal event in the range of 240–370 °C, corresponding to the decomposition of the cellulose acetate chains, representing a loss of 76% of the mass. For CM, the event occurs in the range of 230 to 320 °C and corresponds to the dehydration and decomposition of cellulose. The cellulose decomposition temperature was recorded in Tonset at 245 °C. This value is lower than the temperature at which cellulose begins to decompose and may be associated with the expanded structure of the monolith, which facilitates and accelerates the internal degradation process of cellulose.39,44
Fig. 2 presents the FTIR spectra of the cellulose acetate monoliths before and after the deacetylation reaction.
As seen in Fig. 2, the FTIR spectra for CAM and CM showed marked differences. At 3350 cm−1, high-intensity broadband is observed for the cellulose monolith, which is associated with the O–H stretching of adsorbed water. The first evidence of CAM deacetylation is that the signal for cellulose acetate is less intense. At 1742 cm−1, a peak is registered in the CAM spectrum, which is attributed to the CO stretching of the acetate groups, while in the CM spectrum, this signal is absent, indicating an effective removal of these groups with the alkaline deacetylation reaction. Peaks at 1368 and 1214 cm−1 in the CAM spectrum are attributed to the symmetric deformation of CH3 groups and C–O stretching of acetates, respectively. These signals are only present in the CAM FTIR spectra, confirming the efficiency of the deacetylation process and the production of cellulose monoliths.
The degree of swelling (Sw) was measured for the cellulose monoliths by mass measures before and after water immersion. The Sw shows the polymer's ability to absorb solvent molecules and indicates the polymeric chain mobility and the conformational changes in the polymeric material.45 The results, shown in Fig. 3a, demonstrate that immersion of CM in water resulted in a mass increase of more than seven times on average compared to the initial CM dry mass, Sw of 619 ± 30%. These findings are significantly higher than those reported for particulate cellulosic materials (150–300%), suggesting that the expanded structure of CM allows for better water absorption.45,46 Cellulose monoliths quickly reach equilibrium when in contact with water and do not conform to the viscoelastic model.
Fig. 3 (a) Degree of swelling of cellulose monoliths. (b) Total porosity of cellulose monoliths used as sorptive stir bars. |
As seen, water causes an accentuated swelling of the monolithic structure of cellulose. For this reason, the total porosity of the CM was determined by the fluid saturation method, using n-butanol as solvent. Fig. 3b shows the variation of cellulose monolith porosity as a function of time. According to Fig. 3b, the cellulose monoliths' total porosity (ε) was 68 ± 2% and decreased over time. This result indicates that practically two-thirds of the monolith volume is empty.
The macroporous structure of the cellulose monoliths can be visualized in Fig. 4, which shows the SEM micrographs of the material. The CM is highly composed of cavities with 5–15 μm diameter along its entire surface and structure resembling a beehive. The solid CM structure presents a porous globular material that is sponge-like. Thus, the morphological and structural characteristics of the cellulose monolith are adequate for the extraction of glycerol in biodiesel samples.
Fig. 4 Electron micrographs of cellulose monolith used as a sorptive bar for glycerol extraction in biodiesel samples—cross-section at (a) 5k× and (b) 10k×. |
Initially, the best solvent for desorbing glycerol from CMSB was tested at room temperature (∼27 °C) for 15 minutes for glycerol adsorption and 15 minutes for desorption from CMSB. Ultrapure water, acidic and alkaline aqueous solutions, and methanol were evaluated as extraction solvents. Based on Fig. 5a, water was the most effective solvent for desorbing glycerol from the CMSB, resulting in a recovery rate of approximately 43%. Since glycerol does not dissociate within the pH range of 0–14, extraction with acidified or alkalized aqueous solutions was ineffective. Therefore, ultrapure water was chosen as the extraction solvent.
Fig. 5 Effect of (a) desorption solvent, (b) desorption time, (c) adsorption and desorption temperature for SBSE method using cellulose monolith stir bar for free glycerol extraction from biodiesel. |
To further optimize the extraction process, the contact times of CMSB with the sample and the extraction solvent were varied simultaneously. For example, combinations of 30 min of adsorption and 30 min for desorption or 60 min of adsorption and 60 min for desorption were tested. Fig. 5b shows the effect of CMSB adsorption/desorption time on glycerol extraction from biodiesel samples using water as the extraction solvent.
According to Fig. 5b, increasing the contact time between the CMSB and the sample/extractor improves the glycerol extraction rate. However, the extraction increment was more significant after the adsorption and desorption times were extended up to 48 h, reaching a maximum extraction of 69.5%. Thus, we chose 15 min for adsorption and 15 min for glycerol desorption from CMSB. The high viscosity of biodiesel reduces the mass transfer rate and glycerol diffusion to the CMSB. Therefore, for a significant increase in the glycerol extraction rate, the decrease in biodiesel viscosity was performed with increases in the temperature of the solutions in the adsorption and desorption steps. Fig. 5c shows the glycerol extraction rates by the SBSE method with the CMSB as a function of temperature. The temperatures refer to the bottles containing the biodiesel sample and the ultrapure water. From 40 °C, a significant increase in glycerol extraction rate by CMSB was observed, reaching 93.6% at 70 °C. So, 70 °C was defined as the temperature at which glycerol adsorbs on CMSB and desorbs from it. The effect of sample and eluent volumes was not evaluated in the SBSE extraction process because volumes lower than 5 mL left the CMSB partially uncovered by the liquid. All measurements were performed, at least in triplicates, and the dispersion of the results in each experiment was always less than 3%, indicating an excellent precision of the SBSE method with the CMSB. This also shows the CMSB's good batch-to-batch reproducibility since at least five different preparations were carried out for these studies.
The reuse of CMSB in the extraction of glycerol from biodiesel samples is viable, as seen in Fig. S4 (ESI).† The analyses of the ultrapure water samples, interspersed between extractions of the fortified biodiesel samples, did not record the presence of glycerol, indicating the total desorption of glycerol from CMSB. So, the CMSB was prompted for new use after the regeneration procedure. Furthermore, the glycerol extraction efficiency remained at levels close to 100% extraction after three reuses.
The method's limits of detection and quantification were obtained from the injection of decreasing concentrations of glycerol solutions in the HPLC-RID system. LOD and LOQ values were 6.6 × 10−5% w/w and 2.2 × 10−4% w/w, respectively. These values are below the reference concentrations for glycerol in biodiesel and the recommended methods. Furthermore, the detectability of the proposed method is lower than those other published methods using solvent extraction and capillary electrophoresis,25 gas chromatography-flame ionization detection,47 solid phase extraction (SPE) coupled with enzymatic-spectrophotometric technique,48 and HPLC with diode array detection.24 All used glycerol derivatization for detection. The proposed method shows lower detectability of other extraction methods with HPLC-RID analysis.14,49,50
Analytical parameters of linearity, range, and sensitivity were obtained from the construction of analytical curves for glycerol. Fig. S5 (ESI)† shows a typical analytical curve for glycerol using the SBSE-HPLC-RID method with a CMSB, in which excellent linearity is observed, with a linear regression coefficient (r) of 0.999, a range of 0.00022% to 0.03500% w/w and a sensitivity of 1.595 × 107% w/w.
The accuracy and precision of the method were determined by addition and recovery assays of free glycerol in a biodiesel sample and compared with the result obtained by the official AOCS Ca. 14-56 method. The results obtained are presented in Table 1. It is observed that the glycerol recovery rate was 94 ± 2% by the SBSE-HPLC-RID method using the CMSB, and the result did not show a statistical difference for the AOCS method at the 95% confidence level by the t-test.
The limit concentration of free glycerol in biodiesel is 0.0200% w/w according to several regulatory agencies for different countries,8–10 the AOCS Ca. 14-56 method is recognized as an official standard protocol since it is reliable for this determination. The method developed in this work presents similar results for determining free glycerol in biodiesel samples. However, the SBSE-HPLC-RID method with CMSB needs a few steps, low handling, low cost, and no chemical reagent. It is fully compatible with sustainability and can be considered an eco-friendly method.
The method proposed for determining free glycerol in biodiesel, using SBSE with monolithic cellulose stir bars, is the simplest method available until now. Official methods either use various reagents, which can be toxic (e.g., AOCS Ca. 14-56 method)11 or require experimental conditions that are inaccessible to most analytical laboratories, such as the chromatographic conditions specified in EN: 14103 method.42 In contrast, our method requires only two steps to determine glycerol in B100 biodiesel, utilizing available experimental conditions such as a magnetic stirrer, water, and monolithic cellulose stir bars. Due to its simplicity, this method can be easily scaled and applied for industrial purposes. Thus, it presents a viable alternative for the efficient and safe determination of free glycerol in the biodiesel quality control process. Looking ahead, the automation of this determination by inline SBSE extraction with HPLC-RID analysis can facilitate the process even further.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d4ra02985b |
This journal is © The Royal Society of Chemistry 2024 |