L.
Vázquez-Palomo‡
a,
P.
Montoro-Leal‡
a,
J. C.
García-Mesa‡
a,
M. M.
López Guerrero‡
*ab and
E.
Vereda Alonso‡
*ab
aDepartment of Analytical Chemistry, Faculty of Sciences, University of Malaga, Campus de Teatinos, 29071, Malaga, Spain. E-mail: mmlopez@uma.es; eivereda@uma.es
bInstituto Universitario de Materiales y Nanotecnología, IMANA, University of Malaga, Campus de Teatinos, 29071, Málaga, Spain
First published on 14th November 2024
This work presents a straightforward, highly sensitive, and cost-effective method for the simultaneous determination of V, Ti, Ni and Ga by high resolution-continuum source electrothermal atomic absorption spectrometer (HR-CS ETAAS) in aqueous environmental samples (tap and seawater samples). The system is based on retention of the analyte onto a novel magnetic nanomaterial (M@GO magnetic graphene oxide) functionalised with methylthiosalicilate (MTS). The formed complexes between the M@GO-MTS and the target analytes were broken, adding 1 mL of nitric acid (6%) and sonication for 5 min. The optimized method achieved detection limits of 0.71 μg L−1 for Ti, 0.20 μg L−1 for V, 0.04 μg L−1 for Ga, 0.66 μg L−1 for Ni. The accuracy of the proposed method was demonstrated by analysing two certified reference materials and by determining the analyte content in spiked environmental water samples. The results obtained using this method were in good agreement with the certified values of the standard reference materials, and the recoveries for the spiked tap water and seawater samples ranged from 94% to 120%.
Vanadium, nickel, titanium, and gallium are elements widely found in the environment and used in various industrial and medical applications. Vanadium (V), found in products as catalysts and fertilizers,5–9 has been listed as a highly toxic substance according to the World Health Organization (WHO) classification and as the 13 priority pollutants mentioned in the American Environmental Protection Agency's Clean Water Act.10–12 The concentration of dissolved V in water should not exceed the long-term ecotoxicology limit of 1.2 μg L−1.13 Nickel (Ni), essential in trace amounts for biological processes,14 enters ecosystems through natural processes15,16 and human activities as mining.17,18 It is regulated due to its potential toxicity, common levels of nickel in uncontaminated freshwater sources can range from a few μg L−1 to around 20 μg L−1.18,19 However, higher levels may be found in areas with industrial discharges or pollution sources. Titanium (Ti), is typically found in low concentrations, generally below 10 μg L−1, depending on the geological and environmental characteristics of the area. Higher concentrations of Ti may be found in waters near industrial areas or due to anthropogenic contamination. The determination of Ti in environmental water can be significant, not so much due to its toxicity, but as an indicator of industrial pollution or for studying water treatment processes.20,21 Gallium (Ga), an emerging contaminant, is used in high-tech industries and medical procedures.22–25 Prolonged exposure can cause severe health issues include tachycardia and vertigo,26 but there are no established water quality guidelines for this metal in marine environments.22–24
Biologically active V, Ni, Ga and Ti compounds could pose a potential threat for marine biota when released into the environment. Marine invertebrates are known to accumulate high levels of metals in their tissues. Therefore, accurate risk assessment tools adapted to marine ecosystems based on relevant techniques are needed to identify the pollutants of interest. Such tools can assist in predict potential risks and the hazards associated with marine pollution. The monitoring and control of these trace elements in the environment sources require processing large numbers of samples to accurately characterize their abundance and to reach reliable conclusions.
Electrothermal atomic absorption spectrometry (ETAAS) has been extensively used for trace element analyses. The use of more efficient modifiers, advancements in controlling atomization temperatures, new designs of atomizers, and improvements in methods for correct background spectral interferences have allowed an enhancement in sensitivity, an increase in detection power, reduced sample handling and improved reproducibility of results. Furthermore, the current introduction of high resolution continuum source electrothermal atomic absorption spectrometry (HR-CS ETAAS) has facilitated direct solid sampling, reducing background noise and opening up the possibility of even faster quantification of certain elements and even the simultaneous determination of some elements.27 The combination with the higher radiation intensity of the Xe (only one lamp source is needed) makes it feasible to obtain better defined signals at lower analyte levels, resulting in improved limits of detections. The improved analytical performance of this technique can be seen in elsewhere.28–30 This technique shows a relatively cost-effective approach with low sample consumption, providing figures of merit that can be equivalent to those of inductively coupled plasma mass spectrometry (ICP-MS).30
To increase the sensitivity of the technique, solid phase extraction (SPE) is widely used as pretreatment method for the extraction and preconcentration of analytes. This approach is characterized by its easy to use, fast, low solvent consumption, and high enrichment factors. Thus, the selectivity, sensitivity and precision of the analytical method can be significantly enhanced.31,32 Graphene oxide (GO) and magnetic nanoparticles (MNPs) are materials which offer several advantages that make them attractive for analytical applications, including a large surface area and the presence of active sites on their surfaces. GO is an excellent sorbent; however, it is difficult to separate from the sample. The degradability, biocompatibility, chemical stability, low toxicity, and high magnetic response of iron oxide MNPs (Fe3O4) have been exploited as solid phase sorbents termed magnetic solid phase extraction (MSPE). To exploit the excellent characteristics of both sorbents and mitigate their drawbacks, e.g., to prevent the MNPs aggregates, and to facilitate the separation of GO from the sample, MNPs–GO coupling has been established as an excellent analytical tool for the development of MSPE pre-treatment methods.33
In this work, silica coated magnetic nanoparticles joined to graphene oxide (M@GO)34 modified with a chelating group methylthiosalicilate (MTS) were used as MSPE sorbent to increase the selectivity. The re-extraction of the analytes was reached with only 1 mL of eluent (HNO3 6%).
Considering the diverse sources and characteristics affecting water quality, it is imperative to analyze and understand the concentrations of specific elements, such as vanadium (V), nickel (Ni), gallium (Ga), and titanium (Ti), in aquatic ecosystems. The development of reliable analytical methods capable of the simultaneous determination of several elements (Ti, V, Ni and Ga) is challenging from an analytical point of view because considerably increasing sample throughput (by eliminating sequential determination) and reducing the risk of contamination and/or losses. Moreover, it is worth noting that this approach is particularly intriguing as it offers a cost-effective solution, given that simultaneous analysis of multiple elements is typically not feasible with this conventional equipment.
In this manuscript, the developed method has shown to be a highly useful tool for simultaneous quantitation of V, Ni, Ti and Ga in environmental samples. To the best of the authors' knowledge, this method represents the first instance allowing the simultaneous determination of the four target analytes (V, Ti, Ni and Ga) by ETAAS in water samples. For the quality control of the analytical performance and the validation of the developed method, the analysis of two certified samples, SPS-SW2 Batch 125 and TMDA 64.3, Fortified Lake Waters were addressed. The results obtained from the study exhibited an acceptable level of agreement with the certified values. The method has been applied to tap and seawater samples with goods results.
The lines selected for the simultaneous absorbance measurements and their sensitivities were 294.318 nm (V), sensitivity 2.3%, 294.364 nm (Ga), sensitivity 100%, 294.199 nm (Ti), sensitivity 42%, and 294.391 nm (Ni), sensitivity 1.4%, respectively, and the type of background correction used was IBC (Iterative Baseline Correction). A similar spectral area was previously used by us for the direct analysis of Fe, V, Ga and Ni in solid fuel fly ash.35
The peak height selected absorbance was equivalent to three pixels (central pixel ± adjacent ones). The electrothermal atomizer heating program for the simultaneous determination of V, Ni, Ti and Ga is provided in Table 1.
Stage | Temperature (°C) | Ramp (°C s−1) | Hold time (s) | Ar flow rate |
---|---|---|---|---|
Drying | 110 | 6 | 60 | Max |
Pyrolysis | 1400 | 70 | 3 | Max |
Gas adaptation | 1400 | 0 | 5 | Stop |
Atomization | 2650 | 3000 | 5 | Stop |
Cleaning | 2650 | 0 | 5 | Max |
Argon with a purity of 99.99% was used as purging and protective gas for the graphite atomizer.
A conductometer SensION™ + EC7 (HACH, Colorado, EEUU) was used for salinity measurements of the water samples.
For the synthesis of M@GO MTS graphite powder, sodium chloroacetate, ferrous chloride tetrahydrate (FeCl2·4H2O), ferric chloride hexahydrate (FeCl3·6H2O), 37% HCl (wt/wt), ammonium hydroxide 30% (wt/wt), 97% 3-aminopropyltriethoxysilane, ≥99% tetraethoxysilane (TEOS), methanol, sodium chloride, ≥99.5% ethylenediamine (EN) and 99% N,N′-dicyclohexylcarbodiimide (DCC) were used. 97% Methyl thiosalicylate (MTS) was used for functionalization. All reagents were supplied by Merck (Merck, Darmstadt, Germany).
V(V), Ti(IV), Ni(II), and Ga(III) stock standard solutions 1000 mg L−1 (Merck, Darmstadt, Germany) were used for preparing the calibration curve. Standards of 1000 mg L−1 Pd(II)/Mg(II), Ir(III), Nb(V), and W(VI) solutions (PerkinElmer pure, Darmstadt, Germany) were used to coat the graphite furnace platform.
A pH 5.5 buffer was prepared by mixing acetic acid (0.2 M, 14.8 mL) and sodium acetate (0.2 M, 35.2 mL) (Merck, Darmstadt, Germany) and diluting to 100 mL with de-ionized water.
The accuracy of the proposed procedure was determined by analysing the certified reference materials (CRMs) from National Research Council of Canada (NRCC) TMDA 64.3, Fortified Lake Waters and Standard reference material by National Institute of Standards and Technology (USA) SPS-SW2 Batch 125 Surface Water. To study the applicability of the proposed method, seawater and drinking water samples were also analysed.
The pre-treatment of pyrolytically coated graphite tubes involved a specific procedure to ensure a permanent and high-quality Nb (niobium) coating. This process included the injection of a total of 20 aliquots of 20 μL of a 10 mg L−1 Nb modifier standard solution into the graphite tube, followed by running the device temperature program. The injections were dried slowly by heating the atomizer at 100 °C with a ramp rate and a hold time of 2 °C s−1 and 70 s, respectively. Subsequently, they were thermally treated at 1400 °C with a ramp rate and a hold time of 45 °C s−1 and 130 s, respectively. At the end of these 20 treatments, a reduction step at 1600 °C was applied for 15 s. A total mass of 0.004 mg of the modifier were deposited on the tube graphite wall.35–37 This coating enhances the performance of the tubes during subsequent analysis and extends its useful lifetime to around 350 firings.
On the other hand, the improvement of the vanadium atomization peak required further attention. Therefore, a study on non-permanent matrix modifiers was also conducted, including Pd/Mg (150 ng Pd/100 ng Mg), Ir/W (150 ng Ir/150 ng W), 50 ng and 100 ng Nb (5 μL Nb 5 and 10 mg L−1), as described in the literature.35–37
For the extraction, aliquots of standard or samples were placed in 250 mL volumetric flasks, 25 mL buffer (pH = 5.5) and de-ionized water were added up to the mark. Then 20 mg M@GO-MTS was added. The M@GO-MTS and the sample were kept in contact for 5 min into an ultrasonic bath. After that, a strong Nd/Fe/B magnet was placed at the bottom of the tube and when the solution became clear, the supernatant was able to be easily decanted. The formed complexes between the M@GO-MTS and the target analytes were broken adding 1 mL of nitric acid (6%) and sonication for 5 min. Finally, the sorbent was separated using the Nd/Fe/B magnet and a volume of 20 μL of the supernatant was injected into the Nb coated graphite tube plus 5 μL Nb 10 mg L−1 and analysed by ETAAS.
The experimental conditions and the heating temperature program used were carefully optimized by a univariate form with standards of 15 μg L−1 V(V) and Ti(IV), 30 μg L−1 Ni(II) and 1.5 μg L−1 Ga(III). The optimized furnace temperature program is summarized in Table 1.
The absorbance was measured in quadruplicate (n = 4). All the experiments were carried out in graphite furnace with platforms, preliminary treated with Nb as permanent modifier. The temperature program for the graphite furnace treatment with Nb was described in Section 2.3.
To analyse the CRMs, aliquots of 2.5 mL of sample were placed in 25 mL volumetric flasks, 2.5 mL of buffer pH 5.5 and de-ionized water were added up to the mark. For tap water analysis, 225 mL of tap water were place in 250 mL volumetric flasks, 25 mL of buffer pH 5.5, and for seawater, aliquots of 100 mL were placed in 250 mL volumetric flasks, 25 mL of buffer pH 5.5, and de-ionized water were added up to the mark.
Then, 2 mg of M@GO-MTS for CRMs and 20 mg for tap and sea water was added, following the procedure described in Section 2.4.
Four aliquots of each sample extract were injected for determination of target analytes under the optimal conditions of the proposed method. For seawater samples, the salinity of samples and standards were measured by conductimetry and adjusted at the same value with NaCl before the extraction. The salinity of each sample was analysed and found to be 35.5 g L−1 for seawater.
Fig. 1 Atomic lines in the same spectral window for V, Ti, Ni and Ga. Atomization peaks obtained with a solution of 15 μg L−1 V, Ti, 30 μg L−1 Ni and 1.5 μg L−1 Ga. |
Additionally, the thermo-chemical behaviour of the target analytes in a graphite furnace requires to be considered. Thus, the furnace temperature program and modifiers were studied, as well.
The sensitivity was also studied varying the number of pixels for the determinations. In this case, the number of pixels was set to 3 (the central pixel plus the adjacent ones, CP ± 1). In this work, a further extension of the peak volume detection was not necessary because only noise would be integrated. The signals were measured in terms of both peak area and peak heights. The signals measured as peak height provided more reproducible results due to the tail of the vanadium and titanium atomization signals, which were reduced by the employ of modifiers, as will be discussed in the following sections. Peak height absorbance is probably less susceptible to interferences that can affect peak shape and, thus the accuracy of area integration.41
The best signal-to-noise (S/N) ratios between a blank and a standard solution of 15 μg L−1 V, Ti, 30 μg L−1 Ni and 1.5 μg L−1 Ga were chosen as the optimization criteria. In the study, the analysis of each sample was repeated four times to ensure repeatability and reliability of the results.
Several parameters were optimized in the study to enhance the performance of the analytical method: (a) furnace program; (b) pH; (c) selection of the eluent; (d) extraction and elution times and amount of M@GO-MTS; and (e) volume of sample (preconcentration factor).
Pyrolysis and atomization steps are the most decisive steps to remove most of the matrix components without analyte loss, and to achieve well defined atomization peaks. Commonly, the pyrolysis temperature is established by considering the most volatile element present in the sample, and the atomization temperature is determined by considering the most refractory (least volatile) element present in the sample.42 The pyrolysis temperature was varied over the range 1200–1450 °C. When the pyrolysis temperature increased up to 1400 °C, the signal increased for three of all four, Fig. 2A. Although Ga showed a decrease in the signal, the sensitivity to the line is enough for the determinations. This method has shown the capability to determine concentrations of Ga up to ten times lower than those for the other elements. In comparison to nickel, gallium concentrations are twenty times lower. Consequently, temperature selection is based on maximizing the signal for nickel. The optimum pyrolysis temperature was determined to be 1400 °C. Slower pyrolysis temperature ramps help to eliminate matrix effect. Slower pyrolysis plays a crucial role in minimizing matrix effects during elemental analysis, allowing a better separation of analytes from interfering matrix components. This process enhances the accuracy and precision of measurements. Gradual heating helps the effective volatilization of matrix components without prematurely releasing the analytes, reducing the interference during the measurement phase.43
Fig. 2 (A) Optimization of pyrolysis temperature between 1200–1450 °C. (B) Optimization of atomization temperature between 2500–2650 °C. |
The controlled temperature increase can lead to a more uniform particle size distribution, which aids in the consistent behaviour of the sample matrix, further mitigating matrix effects.43
Slower pyrolysis ramps can reduce the likelihood of signal suppression caused by matrix elements. This is particularly important when are used matrix modifiers, which can sometimes introduce additional variability in the results.44
The pyrolysis ramp was investigated between 30–110 °C s−1 and only a little influence on the analytical signal was found and was considered the optimal values 70 °C s−1. The pyrolysis hold time was varied between 3 and 10 s, only the signal of Ti was increased until 7 s, the signals of V, Ni and Ga started to decrease at times greater than 3 s. Thus, 3 s was selected as the optimum hold time.
Since most of the analytes are refractory, the atomization temperature was varied over the range 2500–2650 °C. When the atomization temperature increased, the signal also increased. The optimum atomization temperature was determined to be 2650 °C because at lower temperature, broader peaks were observed, Fig. 2B. The atomization ramp should be high enough to produce instantaneous atomization, thus a ramp time of 3000 °C s−1 was selected. The atomization hold time was studied between 3 to 10 s. The signals for all analytes increased until 5 s, after that the signals were constant or decreased. Thus, 5 s was selected as optimum hold time.
The high-atomization temperature obtained could potentially affect the thermal stability of the permanent chemical modifiers and lifetime of the graphite tube. However, the number of firings was also studied, and at least, 350 firings could be made without significant decrease in the target analyte signals. The optimal established furnace operation conditions for V, Ni, Ti and Ga based on permanent modifiers, have been listed in Table 1.
The analysis of each sample was repeated four times, employing a one-at-a-time method to optimize the conditions. This optimization led to the selection of Nb as the most effective permanent modifier. In the case of the graphite surface modified Nb, the integrated absorbance for V is higher than for the graphite surface modified with other modifiers.
These results are supported by the study reported elsewhere36,37 where the niobium carbide formation on the graphite surface prevented the formation of vanadium and titanium carbides. Despite using a modifier and increased atomization temperatures, minor tailing in the atomization V peak still appeared on the spectra. To address this issue, several chemical matrix modifiers added along with the sample aliquot were also tested, including Ir/W, Nb, and Pd/Mg, Fig. 3. These modifiers have proven to be effective in thermally stabilizing metals, removing matrix constituents during vaporization and atomization, and extending the lifespan of the furnace.47 For analytical measurements, peak heights were used with better reproducibility to be independent of the peak tails.
Thus, the quantities used for matrix modifiers were determined based on their effectiveness in facilitating the efficient thermal separation of the target analytes and concomitants during the pyrolysis step. Based on these studies, the optimal amount of Nb required was 0.004 mg as a permanent modifier plus 50 ng with each sample aliquot as a non-permanent modifier.
Thus, the pre-concentration of traces of the target analytes from buffered solutions at different pH values was studied. In this study, 25 mL of the standards buffered at different pHs were mixed with 2 mg M@GO-MTS during 5 min in the ultrasonic bath, after the magnetic decantation, 1 mL of 6% HNO3 was added and ultrasonicated 2 min. Finally, 20 μL of the decanted eluent was injected into the Nb coated graphite tube plus 5 μL Nb 10 mg L−1 and analysed by ETAAS. The pH of the buffered standards varied between 1.4 and 10.3. The pH from 1.4 to 5.0 was adjusted using glycine–HCl or sodium acetate–acetic acid buffer solution and from 5.0 to 9.9 using sodium hydroxide–boric acid buffer. The Fig. SM3† shows the absorbances versus pHs.
The optimum pre-concentration pH for Ti was from 4.0 to 6.0; for V was from 4.0 to 6.0, Ni showed optima pH interval, between pH 5.5–8.0, with a maximum to pH 5.5, and Ga showed a range between 4.0–7.0, being maximum at 5.0. Thus, to accomplish the simultaneous determination of the target elements, a pH value of 5.5 was chosen as overall optimum.
The sorption mechanism on the nanomaterial has been studied for other M@GO derivatives and analytes,48 revealing a complex competition among three mechanisms: chelation, hydrogen bonding, and π–electron interactions. In the case of Ni2+ and Ga3+, the predominant extraction mechanism probably involves the formation of chelates via thiol or diamine groups in M@GO-MTS.49,50 For Ti4+, the [TiO]2+ oxo-cation form in aqueous solutions, and its structure in a slightly acidic medium facilitates adsorption through chelation and hydrogen bonding, Fig. SM4A.†,51 As demonstrated in elsewhere,46 the orthovanadate and metavanadate anions possess double bonds, leading to a competition between chelation, hydrogen bonding, and π–electron interactions, similar to the sorption behaviour of arsenic species, Fig. SM4B.†48
The studied eluents were phosphoric acid at 1% plus EDTA 0.5%, nitric acid at 3% plus EDTA 0.5%, nitric acid 3%, nitric acid 6% and nitric acid 7%. As shown in Fig. 4, the best signals were obtained at 6% HNO3, being concluded that effective elution of target analytes was obtained using at eluent.
The volume of eluent required for the quantitative desorption of target analytes was also investigated. The optimum eluent volume was chosen as the smallest possible volume to achieve maximum preconcentration. Quantitative elution was achieved using 1.0 mL of 6% HNO3.
With the optimized extraction parameters, a recovery evaluation study of target analytes was carried out by varying the amount of the sorbent across the range of 1–20 mg, maintaining the volume of sample in 250 mL.
The influence of the amount of M@GO-MTS on the extraction of V, Ni, Ti and Ga (60 μg L−1, 30 μg L−1, 30 μg L−1, 60 μg L−1 respectively) was assessed, Fig. SM6.† The extraction recovery increase with the increase of M@GO-MTS mass. On the other hand, the preconcentration was carried out with the least amount of material possible, allowing quantitative simultaneous recovery, and therefore, 20 mg was selected as optimal amount.
The preconcentration factor can be calculated as the sample to eluent volume ratio. A higher preconcentration factor can be provided by either increasing the sample volume or decreasing the eluent volume.
The recoveries were found to be stable until 250 mL and close to 100%; and the adsorbed V, Ti, Ni and Ga were able to elute with 1.0 mL of eluent. Thus, a preconcentration or enrichment factor (EF) of 250 can be obtained for the target analytes.
Analyte | Calibration graph | Blank ± SD | LOD (pg) | LOQ (pg) | RSD intra-day (%) n = 7a | RSD intra-day (%) n = 7b | RSD Inter-day (%) n = 7c |
---|---|---|---|---|---|---|---|
a 5 μg L−1 Ti and V, 0.5 μg L−1 Ga, 10 μg L−1 Ni. b 10 μg L−1 Ti, and V, 1 μg L−1 Ga, 20 μg L−1 Ni. c 5 μg L−1 Ti, 2 μg L−1 V, 0.5 μg L−1 Ga, 10 μg L−1 Ni. | |||||||
Ti(IV) | y = 0.0028x + 0.0032 | 0.0016 ± 0.0012 | 14.3 | 74.3 | 1.4 | 1.7 | 1.1 |
V(V) | y = 0.0005x + 0.0003 | 0.0006 ± 0.0012 | 4.0 | 32.0 | 2.0 | 2.3 | 2.9 |
Ni(II) | y = 1.22 × 10−5x + 0.00011 | 0.0001 ± 6 × 10−6 | 13.1 | 82.0 | 5.6 | 4.6 | 3.1 |
Ga(III) | y = 0.025x + 0.00048 | 0.0007 ± 0.0003 | 0.9 | 2.6 | 4.6 | 4.3 | 5.6 |
Under the optima conditions described before, the linear ranges were evaluated using the optimised experimental and instrumental conditions. Four linear calibration graphs were obtained from 74.3 pg–0.28 ng for Ti; 32 pg–0.28 ng for V; 82 pg–0.6 ng for Ni, and 2.6 pg–0.1 ng for Ga, with correlation always better than 0.997.
The equations for calibration curves, blank signals, limits of detection and quantification (LODs and LOQs) and other analytical characteristics are presented in Table 2. The limits of detection (LOD) and quantification (LOQ) were calculated as the concentration of the target analyte giving signals equivalent to three and ten times (n = 11), respectively, the standard deviation of the blank plus the net blank intensity.
As can be seen in Table 2, LOD values, based on an EF of 250, ranged from 0.9 to 14.3 pg for the four elements. Considering 20 μL were injected, these LODs correspond with 0.71; 0.20; 0.66; and 0.04 μg L−1 for Ti, V, Ni, and Ga, respectively. The LODs obtained can be considered adequate for the pursued analytical purpose. Using the optimized conditions, this method enables the simultaneous determination of V, Ni, Ti and Ga using the external calibration technique. The precision, evaluated as the intra-day relative standard deviation (RSD) and inter-day values, for seven replicate measurements of solutions, are shown in Table 2. “Intraday RSD” assesses variability within a single day, and “Interday RSD” measures variability between three different days. A low RSD value in both categories indicates precise and consistent measurements. RSDs between 1.1 and 5.6% were obtained, indicating that the method is repeatable and reproducible, showing low variability among analyses performed on the same day and between days. Thus, the results indicated that the method for the simultaneous determination of V, Ti, Ga and Ni is sensible and precise.
For comparative purposes, the analytical performance data of analogous methods reported in the literature are registered in Table 3 (LODs has been shown with amount and concentration units for a better comparison). The direct comparison of the figures of merit is difficult due to the different experimental conditions. Nonetheless, the detection limits for the target elements and the precisions of the developed method were similar or superior to those reported in the bibliography.
Sample | Certificate value (μg L−1) | Found value (μg L−1) | Experimental t < 4.303 = tabulated t for 95% confidencea | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Ti | V | Ni | Ga | Ti | V | Ni | Ga | Ti | V | Ni | Ga | |
a t values obtained comparing certified and found value for 2 d.f. | ||||||||||||
TMDA-64.3 | 124 ± 7 | 280 ± 18 | 251 ± 17 | 50.3 ± 3.4 | 125 ± 1 | 278 ± 2 | 267 ± 11 | 47 ± 2 | 1.73 | 1.73 | 2.52 | 2.85 |
SPS-SW2 | — | 50.0 ± 0.3 | 50.0 ± 0.3 | — | — | 51.3 ± 0.8 | 52.5 ± 1.8 | — | — | 2.81 | 2.40 | — |
Sample | Added (μg L−1) | Found (μg L−1) ± SDa | Recovery (%) ± SDa | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Ti | V | Ni | Ga | Ti | V | Ni | Ga | Ti | V | Ni | Ga | |
a The results are expressed as the average ± standard deviation for four separated determinations. | ||||||||||||
Tap water | — | — | — | — | <LOD (0.71) | <LOD (0.20) | <LOD (0.66) | <LOD (0.04) | — | — | — | — |
5 | 5 | 20 | 3 | 5.9 ± 0.2 | 5.4 ± 0.7 | 22.4 ± 0.6 | 2.9 ± 0.3 | 118 ± 4 | 108 ± 14 | 112 ± 3 | 97 ± 10 | |
15 | 15 | 40 | 5 | 14.4 ± 0.2 | 15.5 ± 0.2 | 43 ± 5 | 4.8 ± 0.3 | 96 ± 1 | 103 ± 1 | 107.5 ± 12 | 96 ± 6 | |
Seawater | — | — | — | — | <LOD (0.71) | <LOD (0.20) | 1.0 ± 0.1 | 0.44 ± 0.02 | — | — | — | — |
15 | 5 | 20 | 1 | 14 ± 1 | 6.0 ± 0.2 | 20 ± 1 | 1.4 ± 0.1 | 94 ± 6 | 120 ± 4 | 95 ± 5 | 96 ± 10 | |
30 | 15 | 40 | 3 | 29.9 ± 0.4 | 16.0 ± 0.2 | 41 ± 4 | 3.44 ± 0.02 | 100 ± 1 | 107 ± 1 | 100 ± 10 | 100 ± 1 |
The obtained results highlight the good harmony between the certified and found values. The t-test applied to each individual result determined there is no significant difference between the experimental average and the population mean (μ) or “true value” = certified value for each result. No significant differences at 95% confidence level were found between the concentrations of V, Ni, Ti and Ga after M@GO-MTS enrichment and the certified concentrations, indicating the accuracy of the proposed method.
The t-test for paired samples was also used in the statistical analysis of data to demonstrate that there are no significant differences between certified and found concentrations of all results between the two samples. The tcalculated (1.85) value was lower than the critical value (tcritical = 2.57) at 95% confidence level, also indicating a good accuracy.
The validated method was applied to the analysis of two real samples: tap and sea water samples, Table 5. The recoveries for the spiked samples were not significantly different to 100%, indicating that the proposed method is useful to the simultaneous determination of the four analytes in samples with very complex matrix such as sea water with salinity about 35.5 g L−1.
A total score exceeding 0.5 is deemed indicative of a green analytical approach. The AGREEprep pictogram, Fig. 6, illustrates a favourable outcome, affirming that the established methodology aligns with contemporary green chemistry principles by eliminating toxic substances, facilitating a sustainability and renewability sorbent material, and minimizing energy usage, and also guaranteeing safe operational practices for the technician.
Nonetheless, this metric suggests that there exists a potential for enhancement, several parameters as sample preparation placement, waste generation, or the automation of the associated procedures.
The proposed method offers important advantages such as simplicity, reproducibility and economy. The sensitivity of the analytical technique is high, and no problems could be determined regarding the minimum content of the target elements. The proposed method was carried out using an external calibration since the standard addition method is time consuming demonstrating the absence of interferences. The sensitivity of different atomic lines was evaluated, and the LOQ and detection were shown to be adequate for the samples analysed. The accuracy of the method was checked by analysing certified reference waters and the results obtained confirmed a good concordance with the certified values.
Additionally, the proposed analytical protocol can be applied to real samples using a single extraction and a single heating program, enabling simultaneous determination, and providing precise and accurate data at low analysis costs. Furthermore, the proposed method adheres to green chemistry principles by avoiding the use of contaminating materials and requiring only a minimal amount of acid, rendering this method environmentally friendly. The greenness of the method has been quantitatively evaluated employing the AGREEprep metric, revealing that proposed method shows an acceptable green analytical method.
To conclude, magnetic dispersive solid phase extraction appears to be a promising approach for various applications, including the extraction of trace metals from environmental samples, the purification of pharmaceuticals, and the recovery of valuable elements from waste materials. Its versatility and efficiency in selectively isolating target analytes suggest that magnetic dispersive solid phase extraction could play a significant role in enhancing analytical methodologies across diverse fields, such as environmental monitoring, food safety, and materials recycling.